Winstation Retinal Imager

by Ophthalmic Imaging Systems

510(k) K982689

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

1 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK982689 Date23 Oct 1998ClearedStatusActive

Recalls

1 recall · FDA

    • Terminated Z-0939-2019 FDA Class II initiated 20 Dec 2018

    Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
HKI
Device Class
Class II
Regulation
21 CFR Part 886.1120
Advisory Committee
Ophthalmic
FDA Decision
SESE Substantially Equivalent