Winstation Retinal Imager
510(k) K982689
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK982689 | Date23 Oct 1998 | StatusActive |
Recalls
-
- Terminated Z-0939-2019
Certain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HKI
- Device Class
- Class II
- Regulation
- 21 CFR Part 886.1120
- Advisory Committee
- Ophthalmic
- FDA Decision
- SESE Substantially Equivalent