Xhibit Central
by Spacelabs Healthcare (Washington), Inc 9 variants
Xhibit Central (040-1040-03) is an electrocardiography monitoring system receiver (FDA Class II) made by Spacelabs Healthcare (Washington), Inc, cleared under FDA 510(k) K141156 on 9 Sep 2014 and listed in FDA GUDID since 6 Jan 2017 (UDI-DI 10841522120244), active as of 23 Oct 2026. 6 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing | Number10841522120244 | Date6 Jan 2017 | StatusListed |
| United States | FDA 510(k)SESE | NumberK141156 | Date9 Sep 2014 | StatusActive |
Official records
- AccessGUDID record 10841522120244
- FDA 510(k) database K141156
From the FDA record
- Device description
- KIT,DUAL ANT,ACT,W/OPS, 614,W/2AMP
- MRI safety
- Labeling does not contain MRI Safety Information
- Natural rubber latex
- No latex statement on the label
- Sterility
- Supplied non-sterile
Across registries
- FDA 510(k) K141156 decided 9 Sep 2014 Active
- FDA GUDID UDI-DI 10841522120244 listed 6 Jan 2017 Listed
Recalls
-
- Ongoing Z-0809-2025
Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
-
- Ongoing Z-0810-2025
Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
-
- Ongoing Z-1513-2022
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
-
- Terminated Z-0532-2018
Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.
-
- Terminated Z-2886-2016
The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
-
- Terminated Z-1502-2016
Reports of an abnormally high frequency of squelch events on Xhibit Telemetry Receivers (XTR), which is indicative of the loss of the telemetry data signal. In some cases, the Xhibit Central Monitor was also reported to display an Xhibit Offline error message for all patients being monitored.
Timeline
- Cleared FDA 510(k) K141156
- Recall initiated FDA recalls Z-1502-2016
- Recall initiated FDA recalls Z-2886-2016
- Listed FDA GUDID 10841522120244
- Recall initiated FDA recalls Z-0532-2018
- Recall initiated FDA recalls Z-1513-2022
- Recall initiated FDA recalls Z-0809-2025
- Recall initiated FDA recalls Z-0810-2025
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Xhibit
- Generic name
- Electrocardiography monitoring system receiver
- Model / Ref / Catalog
- Xhibit Central, 040-1040-03
- Primary DI / UDI-DI
- 10841522120244
- 510(k) Number
- FDA product code
- MSX, MHX, MLD, DSI
- GMDN term
- Electrocardiography monitoring system receiver
- Device class
- Class II
- Regulation number
- 870.2300, 870.1025
- GUDID distribution status
- In Commercial Distribution
Similar devices
Other devices filed under FDA product code MSX (SYSTEM,NETWORK AND COMMUNICATION,PHYSIOLOGICAL MONITORS).
- Wearable Device, Home Package in Carry Case with Homehub, Home Package in Carry Case without Homehub by Current Health, Ltd.
- BeneVision Central Monitoring System by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- Current Health System by Current Health, Ltd.
- AirStrip RPM by AirStrip Technologies, Inc.
- UNITY NETWORK PATIENT DATA SERVER by GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES, INC.
- CARDIO-PAGER SYSTEM by Data Critical Corp.
- WIRELESS LAN (LOCAL AREA NETWORK) - WIRELESS ETHERNET by Marquette Electronics, Inc.
- PHILIPS TELEMEDICINE SERVER/SYSTEM, MODEL 861441 by Philips Medical Systems
- DATEX-OHMEDA S/5 WEB VIEWER by Datex-Ohmeda Inc.
- Hypervisor VI Central Monitoring System (Including TMS) by Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Other devices from Spacelabs Healthcare (Washington), Inc
See all devices from Spacelabs Healthcare (Washington), Inc →