# Xhibit Central

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/xhibit-040-1410-02>
- Manufacturer: [Spacelabs Healthcare (Washington), Inc](https://devicerts.com/manufacturers/spacelabs-healthcare-washington-inc) (FDA registration: United States)
- Device class: Class II (US FDA)
- FDA product code: MHX
- GMDN: Electrocardiography monitoring system receiver (36365)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 10841522127038 · listed 2019-02-25 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K141156 · decision 2014-09-09 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0809-2025 · Class II · ongoing · initiated 2024-11-25: Telemetry receiver systems may experience a Windows proxy Issue causing the system to shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts; necessary clinical intervention could be delayed.
- Z-0810-2025 · Class II · ongoing · initiated 2024-11-25: Telemetry receivers with affected software versions may experience unsolicited system shut down and restart (Offline displayed), lasting approximately 1 to 1.5 minutes, which will cause temporary loss of monitoring. If a clinical event occurred during this period, staff may not detect until the system restarts and necessary clinical intervention could be delayed.
- Z-1513-2022 · Class II · ongoing · initiated 2022-05-31: Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
- Z-0532-2018 · Class II · terminated · initiated 2018-01-17 · matched through shared submission K141156, may concern other models: Spacelabs has received multiple complaints reporting telemetry beds dropping off the Spacelabs Central Station resulting in a loss of monitoring.
- Z-2886-2016 · Class II · terminated · initiated 2016-08-25: The firm received multiple reports of Xhibit Telemetry System going offline or locking up. In addition, short-duration asystole alarms may not display an audio and visual alarm as reported from a single Canadian hospital.
- Z-1502-2016 · Class II · terminated · initiated 2016-03-28: Reports of an abnormally high frequency of squelch events on Xhibit Telemetry Receivers (XTR), which is indicative of the loss of the telemetry data signal. In some cases, the Xhibit Central Monitor was also reported to display an Xhibit Offline error message for all patients being monitored.

## Identity

- Trade name: Xhibit [FDA GUDID (AccessGUDID)]
- Generic name: Electrocardiography monitoring system receiver [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: Xhibit Central [FDA GUDID (AccessGUDID)]; 040-1410-02 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 10841522127038 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K141156 [FDA GUDID (AccessGUDID)]
- FDA product code: MHX [FDA GUDID (AccessGUDID)]
- GMDN term: Electrocardiography monitoring system receiver (36365) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 870.1025 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Xhibit Central, Spacelabs Healthcare (Washington), Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/xhibit-040-1410-02 (data as of 2026-11-17).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-17
- JSON: <https://devicerts.com/devices/xhibit-040-1410-02.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/xhibit-040-1410-02
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.