  "format": "devicerts.device.v1",
  "cite_as": "Ysio, Siemens Medical Solutions USA, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/ysio-kpr (data as of 2026-11-16).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/ysio-kpr",
    "json_url": "https://devicerts.com/devices/ysio-kpr.json",
    "markdown_url": "https://devicerts.com/devices/ysio-kpr.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-16",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/ysio-kpr",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Ysio",
    "registry_name": "YSIO",
    "url": "https://devicerts.com/devices/ysio-kpr",
    "manufacturer": {
      "name": "Siemens Medical Solutions USA, Inc.",
      "url": "https://devicerts.com/manufacturers/siemens-medical-solutions-usa-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K081722",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPR"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K081722",
      "number_label": null,
      "device_udi_di": null,
      "date": "2008-08-22",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0841-2022",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-01-31",
      "reason": "Siemens Healthineers has identified a software error in previous software version VB10 for the Ysio fully automated system. The unit may process incorrect internal values for the position of its ceiling tube stand. If the user activates a longitudinal or transversal movement at the time of the error, the system controls may initiate a correction of the tube stand position, resulting in an unintended movement.  If the described error occurs, a potential for collision risk for objects or persons in the vicinity of the moving stand exists. This may result in minor to serious injuries to persons.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-1501-2020",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2020-02-10",
      "reason": "The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-0299-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-08-06",
      "reason": "In rare cases of high clinical workload, the steel cables inside the lifting column of the overhead tube, which are designed to take the load, may rapture without triggering the safety lock. This could lead to overloading the safety rope, which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-2680-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-08-27",
      "reason": "Siemens discovered that an unlikely error may occur on the Ysio system with fixed detector in the wall stand. This error may result in line artifacts in the image.  If these artifacts appear in the region of interest, the examination may need to be repeated.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-2292-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-07-07",
      "reason": "It was discovered that during a RAD examination using Siemens Luminos dRF, Ysio or Uroskop Omnia systems with software version VB10C to VB10F and automatic exposure control, a highly unlikely malfunction can result in longer radiation exposure than required. This may also result in an overexposed image that is not of diagnostic quality. As a result the taken examination needs to be repeated. \nThi",
      "applies": "shared submission",
      "via": "K081722"
    },
    {
      "number": "Z-1717-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-05-09",
      "reason": "Excessive usage of cleaning liquid on Siemens Ysio Systems may cause infiltration of the generator console, which may lead to a short circuit causing an unintended release of radiation.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-0374-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-09-03",
      "reason": "A potential malfunction and hazard to patients exists when using the Ysio, Luminos Agile, Luminos dRF, and Uroskop Omnia systems with software version VC10, which may result in radiation continuing longer than required and overexposed images that are not of diagnostic quality.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-1845-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-08-28",
      "reason": "When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in the image appearing very dark, which may prompt unnecessary repetition of the examination.",
      "applies": "this device",
      "via": "K081722"
    },
    {
      "number": "Z-0016-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-05-06",
      "reason": "Siemens has discovered through product monitoring a potential malfunction when using system operating with software versions VB10D, VB10F, VB10C and VB10.  During a RAD examination the possibility exists that radiation may continue longer than expected and may result in unnecessary dose applied during an acquisition.  Additionally, it is not possible for the user to stop the radiation manually; ho",
      "applies": "shared submission",
      "via": "K081722"
    },
    {
      "number": "Z-1159-2012",
      "classification": "",
      "status": "terminated",
      "initiated": "2012-02-15",
      "reason": "Firm has become aware of unintended behavior when using Ysio, AXIOM dRF, AXIOM Luminos Agile/TF and Uroskop Omnia with RAD Fluoro Uro function using software version VC10A.  Due to an error in the image pipeline, images may display dark and not suitable for diagnostic.  Due to this software error customers have to repeat x-ray exposure to receive diagnostic images.",
      "applies": "shared submission",
      "via": "K081722"
    },
    {
      "number": "Z-2787-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-28",
      "reason": "There is a potential hazard to staff or patients when lowering the table height on the Ysio system.  If staff or patients lower extremities are placed underneath the table and the user lowers the table top, it is possible that parts of the table may collide with extremities and cause injuries.",
      "applies": "shared submission",
      "via": "K081722"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K081722",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPR",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1680",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}