Skip to content

YSIO X.pree 11107464

by Siemens Healthcare GmbH

Ref 11107464Model MDR_16_II b_ERLN H 127_Rad_digital

YSIO X.pree 11107464 is a stationary basic diagnostic x-ray system (EU MDR Class IIb; FDA Class II; FDA product code KPR, System, x-ray, stationary) made by Siemens Healthcare GmbH, registered in EUDAMED on 26 May 2026 (Basic UDI-DI 0405686901981WH), cleared under FDA 510(k) K233543 on 21 May 2024 (1 earlier submission) and listed in FDA GUDID since 1 Nov 2020 (UDI-DI 04056869232904), active as of 12 Nov 2026. 2 FDA recalls are linked to it.

Active European Union and United States. Also listed in GUDID (US).

CE MDR 510(k)

Valid until

No date given

CE: no date in EUDAMED. FDA: not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

None linked in EUDAMED

A group certificate may exist

Trust

Unverified

Source EUDAMED, GUDID, synced 12 Nov 2026.

Regulatory status

as of registry sync 12 Nov 2026

RegionAuthorizationNumberDateStatus
European UnionCE/MDR EUDAMED Number0405686901981WH Basic UDI-DI 04056869232904UDI-DIDate26 May 2026RegisteredStatusActive
United StatesFDA 510(k)SESE GUDID from GUDID Substantially Equivalent · Traditional NumberK233543 Date21 May 2024DecidedStatusActive
United StatesGUDID listing GUDID Number04056869232904 UDI-DI Date1 Nov 2020ListedStatusListed GUDID: in commercial distribution

Official records

From the FDA record

MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 12 Nov 2026, as the labeller entered it.

Across registries

3 registries list this device

  • EUDAMED Basic UDI-DI 0405686901981WH registered 26 May 2026 Active
  • FDA 510(k) K233543 decided 21 May 2024 Active· +1 more in the status table
  • FDA GUDID UDI-DI 04056869232904 listed 1 Nov 2020 Listed

Recalls

2 recalls · 1 ongoing · FDA

    • Ongoing Z-0730-2022 FDA Class II initiated 7 Mar 2022

    For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.

    • Terminated Z-1346-2021 FDA Class II initiated 10 Mar 2021

    A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted

Timeline

6 dated events

  1. Cleared FDA 510(k) K201670
  2. Listed FDA GUDID 04056869232904
  3. Recall initiated FDA recalls Z-1346-2021
  4. Recall initiated FDA recalls Z-0730-2022
  5. Cleared FDA 510(k) K233543
  6. Registered EUDAMED 0405686901981WH

Coverage

2 jurisdictions

United States of AmericaSwedenFinlandLatviaEstoniaDenmarkLithuaniaIrelandNetherlandsGermanyPolandFranceBelgiumLuxembourgCzech RepublicSlovakiaAustriaHungaryRomaniaSloveniaBulgariaPortugalSpainItalyCroatiaGreeceMaltaCyprus
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered
  • European Union CE/MDR Active
  • United States FDA 510(k) Active

Specification

EUDAMED · GUDID

Trade name
YSIO X.pree
Generic name
Stationary basic diagnostic x-ray system, digital
Device name
MDR_16_II b_ERLN H 127_Rad_digital
Model / reference
11107464, MDR_16_II b_ERLN H 127_Rad_digital
Primary DI / UDI-DI
04056869232904
Basic UDI-DI
0405686901981WH
510(k) Number
FDA product code
KPR
EMDN code
Z11031101 Multifunctional systems for direct digital radiology
GMDN term
Stationary basic diagnostic x-ray system, digital
Device class
Class IIb EUDAMED
Class II GUDID
Regulation number
892.1680
EU registration type
Standard device
EU market of registration
Germany
GUDID distribution status
In Commercial Distribution

Truth per registry

2 sources · 1 field differs

FieldEUDAMEDGUDID
Device classEUDAMED
Class IIb
GUDID
Class II

Device class is stated on each registry's own scale: Class IIb in the EU (EUDAMED), Class II in the US (FDA GUDID). EU and US classes are different schemes, so this is not a disagreement.

Similar devices in this category

Other devices filed under MULTIFUNCTIONAL SYSTEMS FOR DIRECT DIGITAL RADIOLOGY.

Other devices from Siemens Healthcare GmbH

See all devices from Siemens Healthcare GmbH →

Properties

  • Prescription use
  • Serial number
Other properties (23, all answered No)