  "format": "devicerts.device.v1",
  "cite_as": "YSIO X.pree 11107464, Siemens Healthcare GmbH. deviCERTS medical device catalogue. https://devicerts.com/devices/ysio-x-pree-11107464 (data as of 2026-11-27).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/ysio-x-pree-11107464",
    "json_url": "https://devicerts.com/devices/ysio-x-pree-11107464.json",
    "markdown_url": "https://devicerts.com/devices/ysio-x-pree-11107464.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID (AccessGUDID)",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "EUDAMED",
        "url": "https://ec.europa.eu/tools/eudamed"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-11-12",
    "status_as_of": "2026-11-27",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/ysio-x-pree-11107464",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "YSIO X.pree 11107464",
    "registry_name": "YSIO X.pree 11107464",
    "url": "https://devicerts.com/devices/ysio-x-pree-11107464",
    "manufacturer": {
      "name": "Siemens Healthcare GmbH",
      "url": "https://devicerts.com/manufacturers/siemens-healthcare-gmbh",
      "country": "Germany",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "European Union and United States"
    },
    "classification": [
      {
        "class": "Class IIb",
        "system": "EU MDR"
      },
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "Primary DI / UDI-DI",
        "values": [
          {
            "value": "04056869232904",
            "registries": [
              "EUDAMED",
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      },
      {
        "label": "Basic UDI-DI",
        "values": [
          {
            "value": "0405686901981WH",
            "registries": [
              "EUDAMED"
            ]
          }
        ]
      },
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K201670",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          },
          {
            "value": "K233543",
            "registries": [
              "FDA GUDID (AccessGUDID)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KPR"
    ],
    "emdn": [
      {
        "code": "Z11031101",
        "term": "MULTIFUNCTIONAL SYSTEMS FOR DIRECT DIGITAL RADIOLOGY"
      }
    ],
    "gmdn": [
      {
        "code": "37645",
        "term": "Stationary basic diagnostic x-ray system, digital"
      }
    ],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "EUDAMED",
      "region": "European Union",
      "type": "CE/MDR",
      "number": "0405686901981WH",
      "number_label": "Basic UDI-DI",
      "device_udi_di": "04056869232904",
      "date": "2026-05-26",
      "date_kind": "registered",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K233543",
      "number_label": null,
      "device_udi_di": null,
      "date": "2024-05-21",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "GUDID listing",
      "number": "04056869232904",
      "number_label": "UDI-DI",
      "device_udi_di": null,
      "date": "2020-11-01",
      "date_kind": "listed",
      "decision_code": null,
      "expires": null,
      "withdrawn": null,
      "status": "Listed",
      "status_note": "GUDID: in commercial distribution",
      "derived_from_gudid": false
    },
    {
      "registry": "GUDID",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K201670",
      "number_label": null,
      "device_udi_di": null,
      "date": "2020-10-21",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": true
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0730-2022",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2022-03-07",
      "reason": "For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.",
      "applies": "this device",
      "via": "K201670"
    },
    {
      "number": "Z-1346-2021",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-03-10",
      "reason": "A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted",
      "applies": "this device",
      "via": "K201670"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identity",
      "fields": [
        {
          "label": "Trade name",
          "values": [
            {
              "value": "YSIO X.pree",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Generic name",
          "values": [
            {
              "value": "Stationary basic diagnostic x-ray system, digital",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Device name",
          "values": [
            {
              "value": "MDR_16_II b_ERLN H 127_Rad_digital",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Model / reference",
          "values": [
            {
              "value": "11107464",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "MDR_16_II b_ERLN H 127_Rad_digital",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "Primary DI / UDI-DI",
          "values": [
            {
              "value": "04056869232904",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Basic UDI-DI",
          "values": [
            {
              "value": "0405686901981WH",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K201670",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            },
            {
              "value": "K233543",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KPR",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EMDN code",
          "values": [
            {
              "value": "Z11031101",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GMDN term",
          "values": [
            {
              "value": "Stationary basic diagnostic x-ray system, digital",
              "code": "37645",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class IIb",
              "registries": [
                "EUDAMED"
              ]
            },
            {
              "value": "Class II",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Regulation number",
          "values": [
            {
              "value": "892.1680",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "EU registration type",
          "values": [
            {
              "value": "Standard device",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "EU market of registration",
          "values": [
            {
              "value": "Germany",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "GUDID distribution status",
          "values": [
            {
              "value": "In Commercial Distribution",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Properties",
      "fields": [
        {
          "label": "Prescription use",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Serial number",
          "values": [
            {
              "value": "Yes",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Supplied sterile",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Sterilise before use",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Single use",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Implantable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reusable",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains latex",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED",
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Reprocessed single-use device",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction)",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a substance derived from human blood or plasma",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains human tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains animal tissues or cells",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Contains endocrine-disrupting substances",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Administering medicine",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Measuring function",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Incorporates a medicinal substance",
          "values": [
            {
              "value": "No",
              "registries": [
                "EUDAMED"
              ]
            }
          ]
        },
        {
          "label": "Over-the-counter",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Convenience kit",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Combination product",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "HCT/P",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Premarket exempt",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Lot / batch number",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Expiration date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        },
        {
          "label": "Manufacturing date",
          "values": [
            {
              "value": "No",
              "registries": [
                "FDA GUDID (AccessGUDID)"
              ]
            }
          ]
        }
      ]
    }
  ]
}