# YSIO X.pree 11107464

> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/ysio-x-pree-11107464>
- Manufacturer: [Siemens Healthcare GmbH](https://devicerts.com/manufacturers/siemens-healthcare-gmbh) (Germany)
- Device class: Class IIb (EU MDR); Class II (US FDA)
- FDA product code: KPR
- EMDN: Z11031101 MULTIFUNCTIONAL SYSTEMS FOR DIRECT DIGITAL RADIOLOGY
- GMDN: Stationary basic diagnostic x-ray system, digital (37645)

## Registrations and authorizations

- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 0405686901981WH · UDI-DI 04056869232904 · registered 2026-05-26 · status: Active
- **GUDID** · FDA 510(k) · United States · K233543 · decision 2024-05-21 · decision SESE · status: Active · from GUDID
- **GUDID** · GUDID listing · United States · UDI-DI 04056869232904 · listed 2020-11-01 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K201670 · decision 2020-10-21 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0730-2022 · Class II · ongoing · initiated 2022-03-07: For the automated multi-image-acquisition procedure Ortho x-ray collimation is set in a preparative stage for the entire examination area prior to the exam. During acquisition of each individual x-ray image, the x-ray collimator is automatically positioned in a way that the subsequent series of acquisition covers the defined field of view needed for each step. However, during the acquisition the collimation area displayed to the operator on the User Interface does not correctly represent the collimation area specified by the system. It indicates to the user an open collimator instead, e.g., the information displayed on the User Interface shows wider area of collimation than values preset prior to the examination. However, the collimation of the x-ray is performed correctly and always matches the examination area predefined by the user.
- Z-1346-2021 · Class II · terminated · initiated 2021-03-10: A patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted

## Identity

- Trade name: YSIO X.pree [EUDAMED, FDA GUDID (AccessGUDID)]
- Generic name: Stationary basic diagnostic x-ray system, digital [FDA GUDID (AccessGUDID)]
- Device name: MDR\_16\_II b\_ERLN H 127\_Rad\_digital [EUDAMED]
- Model / reference: 11107464 [EUDAMED, FDA GUDID (AccessGUDID)]; MDR\_16\_II b\_ERLN H 127\_Rad\_digital [EUDAMED]

## Identifiers

- Primary DI / UDI-DI: 04056869232904 [EUDAMED, FDA GUDID (AccessGUDID)]
- Basic UDI-DI: 0405686901981WH [EUDAMED]
- 510(k) Number: K201670 [FDA GUDID (AccessGUDID)]; K233543 [FDA GUDID (AccessGUDID)]
- FDA product code: KPR [FDA GUDID (AccessGUDID)]
- EMDN code: Z11031101 [EUDAMED]
- GMDN term: Stationary basic diagnostic x-ray system, digital (37645) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class IIb [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.1680 [FDA GUDID (AccessGUDID)]
- EU registration type: Standard device [EUDAMED]
- EU market of registration: Germany [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Measuring function: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: YSIO X.pree 11107464, Siemens Healthcare GmbH. deviCERTS medical device catalogue. https://devicerts.com/devices/ysio-x-pree-11107464 (data as of 2026-11-27).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-12
- Statuses computed as of: 2026-11-27
- JSON: <https://devicerts.com/devices/ysio-x-pree-11107464.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/ysio-x-pree-11107464
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.