Skip to content

YUKON OCT Spinal System

by K2M, Inc. 2 variants

510(k) K171444

YUKON OCT Spinal System is a medical device (FDA Class II; FDA product code NKG, Posterior Cervical Screw System) made by K2M, Inc., cleared under FDA 510(k) K171444 on 8 Aug 2017, active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

Recall on shared 510(k) K171444; other models may be affected

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK171444 Date8 Aug 2017ClearedStatusActive

Official records

From the FDA record

Received by the FDA
16 May 2017
FDA decision
8 Aug 2017 (Substantially Equivalent)
FDA review time
84 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K171444 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

3 recalls · 1 ongoing · FDA

    • Ongoing Z-1307-2022 FDA Class II initiated 13 Jun 2022 on shared 510(k) K171444

    Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.

    • Terminated Z-0931-2021 FDA Class II initiated 14 Dec 2020

    The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).

    • Terminated Z-0110-2020 FDA Class II initiated 24 Jul 2018

    Screws had a manufacturing error that could prevent the screw from mating with the screw inserter or set screw as intended. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; the action was completed in 2018.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

4 dated events

  1. Cleared FDA 510(k) K171444
  2. Recall initiated FDA recalls Z-0110-2020
  3. Recall initiated FDA recalls Z-0931-2021
  4. Recall initiated FDA recalls Z-1307-2022

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
NKG
Device class
Class II
Regulation
21 CFR Part 888.3075
Advisory committee
Orthopedic
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code NKG (Posterior Cervical Screw System).

Other devices from K2M, Inc.