# YUKON OCT Spinal System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/yukon-oct-spinal-system-nkg>
- Manufacturer: [K2M, Inc.](https://devicerts.com/manufacturers/k2m-inc) (United States)
- Device class: Class II (US FDA)
- FDA product code: NKG

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K171444 · decision 2017-08-08 · decision SESE · status: Active

## Recalls

- Z-1307-2022 · Class II · ongoing · initiated 2022-06-13 · matched through shared submission K171444, may concern other models: Some of the rods in this lot were observed to have discoloration, later confirmed to be indication of trace amounts of inorganic phosphorous compound. This may produce a mild irritating or inflammatory response; adverse local tissue reactions may occur related to cytotoxicity with the nonconforming rod and surrounding tissue.
- Z-0931-2021 · Class II · terminated · initiated 2020-12-14: The Yukon polyaxial screw used in the Yukon Spinal System was incorrectly laser marked with size 5.0 mm (diameter) & 24 mm (length).
- Z-0110-2020 · Class II · terminated · initiated 2018-07-24: Screws had a manufacturing error that could prevent the screw from mating with the screw inserter or set screw as intended. This event was reported to FDA as part of Stryker's retrospective review of K2M product field actions; the action was completed in 2018.

## Identifiers

- 510(k) Number: K171444 [FDA 510(k)]
- FDA product code: NKG [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 888.3075 [FDA 510(k)]
- Advisory committee: Orthopedic [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: YUKON OCT Spinal System, K2M, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/yukon-oct-spinal-system-nkg (data as of 2026-11-29).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-29
- JSON: <https://devicerts.com/devices/yukon-oct-spinal-system-nkg.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/yukon-oct-spinal-system-nkg
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.