# Z-800 Infusion System

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/z-800-infusion-system-frn>
- Manufacturer: [Zyno Medical, LLC](https://devicerts.com/manufacturers/fda-fei-3006575795) (FDA registration: United States)
- Name in the registry: Z-800 INFUSION SYSTEM
- Device class: Class II (US FDA)
- FDA product code: FRN

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K130690 · decision 2013-06-12 · decision SESE · status: Active

## Recalls

- Z-1865-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1866-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1867-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1868-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-0506-2025 · Class II · ongoing · initiated 2024-10-18: Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
- Z-0523-2025 · Class III · ongoing · initiated 2024-09-30: For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- Z-0005-2025 · Class I · ongoing · initiated 2024-09-13: There is a defect in the air-in-line software algorithm.
- Z-2263-2024 · Class II · ongoing · initiated 2024-06-04: Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
- Z-1183-2024 · Class II · ongoing · initiated 2023-12-20: When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death

## Identifiers

- 510(k) Number: K130690 [FDA 510(k)]
- FDA product code: FRN [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 880.5725 [FDA 510(k)]
- Advisory committee: General Hospital [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Z-800 Infusion System, Zyno Medical, LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/z-800-infusion-system-frn (data as of 2026-11-23).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/z-800-infusion-system-frn.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/z-800-infusion-system-frn
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.