- **GUDID** · GUDID listing · United States · UDI-DI 00814371020006 · listed 2015-10-13 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K130690 · decision 2013-06-12 · decision SESE · status: Active · from GUDID

## Recalls

- Z-0506-2025 · Class II · ongoing · initiated 2024-10-18: Potential for the battery to deplete over time from self-discharge and internal parasitic leakage current despite being continuously plugged into AC mains.
- Z-0523-2025 · Class III · ongoing · initiated 2024-09-30: For the Z-800WF pumps, the WiFi software was not compatible with the pump software that had been loaded in accordance with ongoing correction Z-1183-2024.
- Z-0005-2025 · Class I · ongoing · initiated 2024-09-13: There is a defect in the air-in-line software algorithm.
- Z-1865-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1866-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1867-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-1868-2025 · Class I · ongoing · initiated 2025-05-07: Unreleased software versions were installed on distributed devices without verification or validation.
- Z-2263-2024 · Class II · ongoing · initiated 2024-06-04: Zyno Medical has found that certain Z800, Z800F, Z800W, and Z800WF pumps have faulty power supply modules that will not provide a charge to the rechargeable battery within the pump. This may lead to the battery having insufficient or no charge available to the pump when it is unplugged from the AC mains power.
- Z-1183-2024 · Class II · ongoing · initiated 2023-12-20: When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death

## Identity

- Generic name: Bedside infusion pump, single-channel [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: Z-800 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00814371020006 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K130690 [FDA GUDID (AccessGUDID)]
- FDA product code: FRN [FDA GUDID (AccessGUDID)]
- GMDN term: Bedside infusion pump, single-channel (13215) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 880.5725 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Clinical & Handling

- MRI safety: MR Conditional [FDA GUDID (AccessGUDID)]

## Properties

- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Z-800, Zyno Medical LLC. deviCERTS medical device catalogue. https://devicerts.com/devices/z-800-z-800 (data as of 2026-11-23).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-21
- Statuses computed as of: 2026-11-23
- JSON: <https://devicerts.com/devices/z-800-z-800.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/z-800-z-800
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.