Zenition 50
by Philips Medical Systems Nederland B.V.
510(k) K183101
Active
Valid until
Not listed
FDA recalls
5 ongoingTrust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE 510(k) | NumberK183101 | Date31 Dec 2018 | StatusActive |
| European Union | CE/MDR EUDAMED | Number— | Date— | StatusActive |
Recalls
-
- Ongoing Z-0241-2026
Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.
-
- Ongoing Z-1580-2025
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation.
-
- Ongoing Z-0352-2025
Potential for the Mains Control Unit board fuse may blow out during system start-up or during a procedure.
-
- Ongoing Z-0226-2024
Philips has become aware of an issue with the Philips Integris-Allura, Allura Xper and Azurion systems that have been delivered to Puerto Rico without a warning label and the Source-to-Skin Distance Spacer (Tube-spacer) is missing.
-
- Ongoing Z-0647-2022
Firm discovered that the wireless foot switch can suddenly stop responding due to a firmware issue. If the wireless foot switch fails, there could be a delay or interruption of procedure if the hand switch is not used or if the wired foot switch is not connected to the system.
-
- Terminated Z-1932-2020
When strain relief is lost at the stand side of the Stand MVS Trolley Cable, over time one or more electrical contacts could break due to pulling, twisting, or dropping of the connector. If this happens, live parts may be exposed (230 V) and/or the video signal can be lost.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
- European Union
- United States
Specification
- 510(k) Number
- FDA Product Code
- OWB
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1650
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 18 Sep 2026
The Zenition 50 is an Interventional Fluoroscopic X-Ray System, designed for real-time imaging during surgical and interventional procedures. It employs a mobile C-arm configuration with an integrated image intensifier to provide high-quality fluoroscopic views, enabling precise guidance for procedures such as orthopedics, cardiology, and vascular interventions.
This system is intended for use in clinical settings where portability and flexibility are required, such as operating rooms, catheterization labs, and interventional radiology suites. Supplied as a reusable medical device, it adheres to FDA 510(k) clearance and MDR regulatory standards. The Zenition 50 operates within Class II risk classification and includes multiple FEI numbers and registration numbers for compliance tracking. Storage and handling follow standard protocols for fluoroscopic equipment, with no indication of single-use or MRI compatibility.
Other devices from Philips Medical Systems Nederland B.V.
See all devices from Philips Medical Systems Nederland B.V. →
Questions
AI-generated answers, 18 Sep 2026
What types of procedures is the Zenition 50 designed to support with its real-time imaging capabilities?
The Zenition 50 is specifically designed for real-time imaging during surgical and interventional procedures in fields like orthopedics, cardiology, and vascular interventions. Its mobile C-arm configuration with an integrated image intensifier provides high-quality fluoroscopic views, making it ideal for guiding precise interventions in settings such as operating rooms, catheterization labs, and interventional radiology suites.
Is the Zenition 50 a single-use or reusable device, and what does this mean for maintenance and costs?
The Zenition 50 is a reusable medical device, meaning it is intended for repeated use after proper cleaning, disinfection, and maintenance. This design reduces per-procedure costs but requires adherence to manufacturer guidelines for reprocessing to ensure safety, sterility, and performance consistency over multiple uses.
How is the Zenition 50 regulated, and what does this imply for its use in clinical settings?
The Zenition 50 is authorized under both FDA 510(k) clearance and the EU’s Medical Device Regulation (MDR), classifying it as a Class II device. This regulatory pathway ensures it meets stringent safety and performance standards for interventional fluoroscopic systems, allowing its use in clinical settings where portability and flexibility are critical, such as ORs and catheterization labs.
What storage and handling requirements should be followed for the Zenition 50 to maintain its functionality?
While specific storage details aren’t provided, as a reusable fluoroscopic system, the Zenition 50 should be stored in a clean, dry environment protected from physical damage, extreme temperatures, and electromagnetic interference. Manufacturer guidelines likely include regular maintenance checks and calibration to ensure optimal imaging quality and safety during use.
Does the Zenition 50 come with multiple configurations or models, and how might this affect procedure planning?
The Zenition 50 is described as a single interventional fluoroscopic system with a mobile C-arm design, but it includes multiple FEI (Facility Equipment Identifier) and registration numbers, suggesting compliance tracking across various institutional setups. While no specific models are mentioned, its adaptability for orthopedics, cardiology, and vascular procedures implies versatility in procedure planning, though exact configurations would depend on institutional customization or accessories.