Skip to content

Zimmer Segmental System

by Zimmer, Inc. 2 variants

510(k) K070978

Zimmer Segmental System is a medical device (FDA Class II; FDA product code KRO, Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer) made by Zimmer, Inc., cleared under FDA 510(k) K070978 on 3 Jul 2007, active as of 22 Oct 2026. 6 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

6 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK070978 Date3 Jul 2007ClearedStatusActive

Official records

From the FDA record

Received by the FDA
6 Apr 2007
FDA decision
3 Jul 2007 (Substantially Equivalent)
FDA review time
88 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K070978 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

6 recalls · FDA

    • Terminated Z-0721-2018 FDA Class II initiated 1 Aug 2017

    The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.

  • 3 recalls, same reason

    • Terminated Z-2566-2017 FDA Class II initiated 28 Feb 2017
    • Terminated Z-2574-2017 FDA Class II initiated 28 Feb 2017
    • Terminated Z-2575-2017 FDA Class II initiated 28 Feb 2017

    Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.

  • 2 recalls, same reason

    • Terminated Z-0783-2014 FDA Class II initiated 19 Dec 2013
    • Terminated Z-0784-2014 FDA Class II initiated 19 Dec 2013

    After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.

Timeline

7 dated events

  1. Cleared FDA 510(k) K070978
  2. Recall initiated FDA recalls Z-0783-2014
  3. Recall initiated FDA recalls Z-0784-2014
  4. Recall initiated FDA recalls Z-2566-2017
  5. Recall initiated FDA recalls Z-2574-2017
  6. Recall initiated FDA recalls Z-2575-2017
  7. Recall initiated FDA recalls Z-0721-2018

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
KRO
Device class
Class II
Regulation
21 CFR Part 888.3510
Advisory committee
Orthopedic
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code KRO (Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer).

Other devices from Zimmer, Inc.

See all devices from Zimmer, Inc. →