## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K070978 · decision 2007-07-03 · decision SESE · status: Active

## Recalls

- Z-0721-2018 · Class II · terminated · initiated 2017-08-01: The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
- Z-2566-2017 · Class II · terminated · initiated 2017-02-28: Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2574-2017 · Class II · terminated · initiated 2017-02-28: Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-2575-2017 · Class II · terminated · initiated 2017-02-28: Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.
- Z-0783-2014 · Class II · terminated · initiated 2013-12-19: After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.
- Z-0784-2014 · Class II · terminated · initiated 2013-12-19: After evaluation of 13 complaints(0.4% rate) for hyper-extension of the Zimmer Segmental System, Zimmer is updating the related surgical technique and instructions for use to provide additional instructions and patient conditions that may place excessive loading on the polyethylene insert.

## Identifiers

- 510(k) Number: K070978 [FDA 510(k)]
- FDA product code: KRO [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 888.3510 [FDA 510(k)]
- Advisory committee: Orthopedic [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Zimmer Segmental System, Zimmer, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/zimmer-segmental-system-kro-k070978 (data as of 2026-11-28).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-28
- JSON: <https://devicerts.com/devices/zimmer-segmental-system-kro-k070978.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/zimmer-segmental-system-kro-k070978
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.