AMERICAN CONTRACT SYSTEMS, INC.

US

Last updated May 23, 2026

What AMERICAN CONTRACT SYSTEMS, INC. makes

American Contract Systems, Inc. manufactures single-use, non-medicated general surgical procedure kits designed for various surgical applications. These kits are intended for one-time use in operating room settings and include essential tools and instruments required for specific surgical procedures.

The company’s products are regulated under FDA classification as either Class I or Class II devices, with some falling under the non-classifiable (N) category. These devices are listed in the FDA’s Unique Device Identifier (UDI) database, ensuring traceability and compliance with U.S. regulatory requirements.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Information

Country
US
Address
—
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
010334675

Catalogue (21894)

Page 23 of 438
DeviceModel / ReferenceRegistriesClassStatus
Acs LMMJ24W
FDA UDI
Class IIActive
Acs UILC71AA
FDA UDI
Class IIActive
Acs MCCS46E
FDA UDI
Class IIActive
Acs FHSH28AA
FDA UDI
Class IIActive
Acs JETK06Q
FDA UDI
Class IActive
Acs YOEX27N
FDA UDI
Class IActive
Acs OSBR15D
FDA UDI
Class IActive
Acs TNLP15N
FDA UDI
Class IIActive
Acs JRAB40I
FDA UDI
Class IIActive
Acs UIKN28AF
FDA UDI
Class IActive
Acs CDLD97D
FDA UDI
Class IIActive
Acs LSRENDVASF
FDA UDI
Class IIActive
Acs SWLP81D
FDA UDI
Class IIActive
Acs MMSH31B
FDA UDI
Class IIActive
Acs LLAP22F
FDA UDI
Class IActive
Acs ANPP15AP
FDA UDI
Class IActive
Acs HSAN63N
FDA UDI
Class IIActive
Acs UIOC22J
FDA UDI
Class IIActive
Acs UICR54H
FDA UDI
Class IIActive
Acs SMCS22O
FDA UDI
Class IIActive
Acs EVPV29E
FDA UDI
Class IIActive
Acs UIPC89D
FDA UDI
Class IIActive
Acs BLGY18K
FDA UDI
Class IIActive
Acs HSRO81G
FDA UDI
Class IIActive
Acs BHCA01I
FDA UDI
Class IIActive
Acs LSROHEARTI
FDA UDI
Class IIActive
Acs CKSP56E
FDA UDI
Class IIActive
Acs TNHH26R
FDA UDI
Class IActive
Acs SLVD11I
FDA UDI
Class IIActive
Acs WPMP16E
FDA UDI
Class IIActive
Acs CDMH06E
FDA UDI
Class IIActive
Acs MCSU87D
FDA UDI
Class IIActive
Acs BDCS31A
FDA UDI
Class IIActive
Acs FIKS12A
FDA UDI
Class IActive
Acs SVTH15H
FDA UDI
Class IIActive
Acs CWCP51A
FDA UDI
Class IIActive
Acs SCAR03O
FDA UDI
Class IIActive
Acs CEVS11I
FDA UDI
Class IIActive
Acs CFAR29C
FDA UDI
Class IActive
Acs BLUB95D
FDA UDI
Class IIActive
Acs CCAP12C
FDA UDI
Class IActive
Acs UIOL34AK
FDA UDI
Class IIActive
Acs UDMI78H
FDA UDI
Class IIActive
Acs MMCC14G
FDA UDI
Class IIActive
Acs EMPM16F
FDA UDI
Class IIActive
Acs BFMJ03Z
FDA UDI
Class IIActive
Acs SVNT50D
FDA UDI
Class IIActive
Acs TOPS11B
FDA UDI
Class IIActive
Acs SLSH06C
FDA UDI
Class IIActive
Acs WPHC45G
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (100)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

American Contract Systems, Inc. has 190 recall records spanning from December 28, 2016, to March 19, 2026, with statuses recorded as open, classified, or terminated. The reasons cited include concerns about incomplete sterilization, improper handling of components during sterilization processes, incorrect expiration dates, labeling errors, insufficient aeration time, and deviations from sterilization specifications. Some recalls involved third-party products or components, such as syringes from Medline Industries, and the use of FDA-banned substances in powdered surgical gloves.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Mar 19, 2026Z-1981-2026—open, classified

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Sep 4, 2025Z-0179-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0185-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0187-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0180-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0178-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0172-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0171-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0183-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0174-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0175-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0177-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0186-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0173-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0170-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0184-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0181-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0182-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0176-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Jul 14, 2025Z-2466-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2471-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2465-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2467-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2468-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2469-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2470-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2464-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2472-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Apr 21, 2025Z-1916-2025—open, classified

Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.

Feb 6, 2025Z-1332-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1333-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1349-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1345-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1359-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1356-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1339-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1348-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1358-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1361-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1346-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1331-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1350-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1355-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1347-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1353-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1362-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1354-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1341-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1338-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1352-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1342-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1334-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1357-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1360-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1340-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1335-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1336-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1344-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1337-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1343-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1351-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 4, 2025Z-1406-2025—open, classified

During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.

Jan 31, 2025Z-1425-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1426-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1424-2025—open, classified

Sterility assurance with procedure trays

Aug 7, 2024Z-3230-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3238-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3239-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3227-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3241-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3233-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3242-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3226-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3231-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3229-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3237-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3244-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3232-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3228-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3235-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3240-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3236-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3243-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3234-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

May 15, 2024Z-2274-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2277-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2304-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2283-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2310-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2290-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2284-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2275-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2301-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2299-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2280-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2281-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2307-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2306-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2303-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2308-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.