AMERICAN CONTRACT SYSTEMS, INC.

US

Last updated May 23, 2026

What AMERICAN CONTRACT SYSTEMS, INC. makes

American Contract Systems, Inc. manufactures single-use, non-medicated general surgical procedure kits designed for various surgical applications. These kits are intended for one-time use in operating room settings and include essential tools and instruments required for specific surgical procedures.

The company’s products are regulated under FDA classification as either Class I or Class II devices, with some falling under the non-classifiable (N) category. These devices are listed in the FDA’s Unique Device Identifier (UDI) database, ensuring traceability and compliance with U.S. regulatory requirements.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Information

Country
US
Address
—
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
010334675

Catalogue (21894)

Page 35 of 438
DeviceModel / ReferenceRegistriesClassStatus
Acs CALC98G
FDA UDI
Class IIActive
Acs FTKS02O
FDA UDI
Class IActive
Acs SLCY61H
FDA UDI
Class IIActive
Acs FTGS08M
FDA UDI
Class IIActive
Acs LATH01AE
FDA UDI
Class IActive
Acs PRTO31X
FDA UDI
Class IActive
Acs AHCR93D
FDA UDI
Class IActive
Acs SAHP83P
FDA UDI
Class IActive
Acs ANCV78AW
FDA UDI
Class IIActive
Acs AHCY29O
FDA UDI
Class IIActive
Acs BLAR01E
FDA UDI
Class IIActive
Acs HOVH19C
FDA UDI
Class IIActive
Acs ALEX17D
FDA UDI
Class IActive
Acs SFMB05B
FDA UDI
Class IIActive
Acs UINC07A
FDA UDI
Class IIActive
Acs PKML85A
FDA UDI
Class IIActive
Acs EGKA02D
FDA UDI
Class IIActive
Acs NRCC98C
FDA UDI
Class IIActive
Acs HSSK90H
FDA UDI
Class IIActive
Acs BETK07J
FDA UDI
Class IIActive
Acs RWCS71G
FDA UDI
Class IIActive
Acs SACY80N
FDA UDI
Class IIActive
Acs UDAC66AA
FDA UDI
Class IActive
Acs UDKN68AH
FDA UDI
Class IActive
Acs SAAC27M
FDA UDI
Class IIActive
Acs SFAH67E
FDA UDI
Class IIActive
Acs LSRSMPIVINA
FDA UDI
Class IIActive
Acs BOCP70K
FDA UDI
Class IActive
Acs AHTS05A
FDA UDI
Class IActive
Acs MPCH19A
FDA UDI
Class IIActive
Acs QPTH29B
FDA UDI
Class IActive
Acs TSTH54C
FDA UDI
Class IActive
Acs UHSH77AF
FDA UDI
Class IActive
Acs MHRB88AE
FDA UDI
Class IIActive
Acs HEDV62AJ
FDA UDI
Class IIActive
Acs LLDC40A
FDA UDI
Class IIActive
Acs LKLP32F
FDA UDI
Class IIActive
Acs UHSP33V
FDA UDI
Class IActive
Acs BLOT19A
FDA UDI
Class IIActive
Acs AHCR93E
FDA UDI
Class IActive
Acs CEAT19I
FDA UDI
Class IIActive
Acs PWNP56O
FDA UDI
Class IIActive
Acs LMLP08Z
FDA UDI
Class IIActive
Acs UDMJ62Y
FDA UDI
Class IIActive
Acs AGMI14I
FDA UDI
Class IIActive
Acs HSWS80E
FDA UDI
Class IIActive
Acs AGLA66F
FDA UDI
Class IIActive
Acs LLDC40K
FDA UDI
NActive
Acs AGMM60N
FDA UDI
Class IIActive
Acs ANLM40X
FDA UDI
Class IActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (100)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

American Contract Systems, Inc. has 190 recall records spanning from December 28, 2016, to March 19, 2026, with statuses recorded as open, classified, or terminated. The reasons cited include concerns about incomplete sterilization, improper handling of components during sterilization processes, incorrect expiration dates, labeling errors, insufficient aeration time, and deviations from sterilization specifications. Some recalls involved third-party products or components, such as syringes from Medline Industries, and the use of FDA-banned substances in powdered surgical gloves.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Mar 19, 2026Z-1981-2026—open, classified

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Sep 4, 2025Z-0179-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0185-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0187-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0180-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0178-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0172-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0171-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0183-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0174-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0175-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0177-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0186-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0173-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0170-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0184-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0181-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0182-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0176-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Jul 14, 2025Z-2466-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2471-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2465-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2467-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2468-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2469-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2470-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2464-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2472-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Apr 21, 2025Z-1916-2025—open, classified

Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.

Feb 6, 2025Z-1332-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1333-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1349-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1345-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1359-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1356-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1339-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1348-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1358-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1361-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1346-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1331-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1350-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1355-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1347-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1353-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1362-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1354-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1341-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1338-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1352-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1342-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1334-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1357-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1360-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1340-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1335-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1336-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1344-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1337-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1343-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1351-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 4, 2025Z-1406-2025—open, classified

During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.

Jan 31, 2025Z-1425-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1426-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1424-2025—open, classified

Sterility assurance with procedure trays

Aug 7, 2024Z-3230-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3238-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3239-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3227-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3241-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3233-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3242-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3226-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3231-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3229-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3237-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3244-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3232-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3228-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3235-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3240-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3236-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3243-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3234-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

May 15, 2024Z-2274-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2277-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2304-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2283-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2310-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2290-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2284-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2275-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2301-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2299-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2280-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2281-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2307-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2306-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2303-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2308-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.