What AMERICAN CONTRACT SYSTEMS, INC. makes

American Contract Systems, Inc. manufactures single-use, non-medicated general surgical procedure kits designed for various surgical applications. These kits are intended for one-time use in operating room settings and include essential tools and instruments required for specific surgical procedures.

The company’s products are regulated under FDA classification as either Class I or Class II devices, with some falling under the non-classifiable (N) category. These devices are listed in the FDA’s Unique Device Identifier (UDI) database, ensuring traceability and compliance with U.S. regulatory requirements.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Information

Country
US
Address
—
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
010334675

Catalogue (21894)

Page 4 of 438
DeviceModel / ReferenceRegistriesClassStatus
Acs CDCO11K
FDA UDI
Class IIActive
Acs CIPO16D
FDA UDI
Class IIActive
Acs JEPI90J
FDA UDI
Class IIActive
Acs SAPL07G
FDA UDI
Class IIActive
Acs OSHN17D
FDA UDI
Class IIActive
Acs AMCP14E
FDA UDI
Class IIActive
Acs CXGU11AF
FDA UDI
Class IIActive
Acs HGTV60C
FDA UDI
Class IIActive
Acs LLLC26E
FDA UDI
Class IActive
Acs PCEX64G
FDA UDI
Class IActive
Acs AHMD52C
FDA UDI
Class IActive
Acs CLSH99H
FDA UDI
Class IActive
Acs PRGY82D
FDA UDI
Class IIActive
Acs MRST14J
FDA UDI
Class IIActive
Acs ALAL04A
FDA UDI
Class IIActive
Acs UIHY16AD
FDA UDI
Class IIActive
Acs JETH05Q
FDA UDI
Class IActive
Acs CATH71D
FDA UDI
Class IIActive
Acs BFCT04X
FDA UDI
Class IIActive
Acs BDPD87B
FDA UDI
Class IActive
Acs SJUL18I
FDA UDI
Class IIActive
Acs LLSI16G
FDA UDI
Class IIActive
Acs CGPD52C
FDA UDI
Class IActive
Acs BHTA02J
FDA UDI
Class IIActive
Acs WBHP78K
FDA UDI
Class IActive
Acs UDIN23AI
FDA UDI
Class IIActive
Acs FDGP05AC
FDA UDI
Class IIActive
Acs DVTR48E
FDA UDI
Class IIActive
Acs UILP16AE
FDA UDI
Class IIActive
Acs FHKA19AA
FDA UDI
Class IActive
Acs UITC53AI
FDA UDI
Class IIActive
Acs AMJL22X
FDA UDI
Class IIActive
Acs MHCV93F
FDA UDI
Class IIActive
Acs SJCP39D
FDA UDI
Class IIActive
Acs EGTK10F
FDA UDI
Class IIActive
Acs AMOH35AA
FDA UDI
Class IIActive
Acs MSMT46A
FDA UDI
Class IIActive
Acs SSSP83F
FDA UDI
Class IActive
Acs SHUR48J
FDA UDI
Class IIActive
Acs AHOR34Q
FDA UDI
Class IActive
Acs NUCT01A
FDA UDI
Class IIActive
Acs UICR54N
FDA UDI
Class IIActive
Acs SLDA35H
FDA UDI
Class IIActive
Acs ISAR20B
FDA UDI
Class IIActive
Acs ICTA89D
FDA UDI
Class IIActive
Acs IHLC11AC
FDA UDI
Class IIActive
Acs UTLE27Q
FDA UDI
Class IActive
Acs WVCS02F
FDA UDI
Class IIActive
Acs FDCL22L
FDA UDI
Class IIActive
Acs GILI99D
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (100)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

American Contract Systems, Inc. has 190 recall records spanning from December 28, 2016, to March 19, 2026, with statuses recorded as open, classified, or terminated. The reasons cited include concerns about incomplete sterilization, improper handling of components during sterilization processes, incorrect expiration dates, labeling errors, insufficient aeration time, and deviations from sterilization specifications. Some recalls involved third-party products or components, such as syringes from Medline Industries, and the use of FDA-banned substances in powdered surgical gloves.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Mar 19, 2026Z-1981-2026—open, classified

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Sep 4, 2025Z-0179-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0185-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0187-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0180-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0178-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0172-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0171-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0183-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0174-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0175-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0177-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0186-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0173-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0170-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0184-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0181-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0182-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0176-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Jul 14, 2025Z-2466-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2471-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2465-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2467-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2468-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2469-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2470-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2464-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2472-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Apr 21, 2025Z-1916-2025—open, classified

Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.

Feb 6, 2025Z-1332-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1333-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1349-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1345-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1359-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1356-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1339-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1348-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1358-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1361-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1346-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1331-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1350-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1355-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1347-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1353-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1362-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1354-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1341-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1338-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1352-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1342-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1334-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1357-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1360-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1340-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1335-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1336-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1344-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1337-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1343-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1351-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 4, 2025Z-1406-2025—open, classified

During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.

Jan 31, 2025Z-1425-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1426-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1424-2025—open, classified

Sterility assurance with procedure trays

Aug 7, 2024Z-3230-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3238-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3239-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3227-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3241-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3233-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3242-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3226-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3231-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3229-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3237-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3244-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3232-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3228-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3235-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3240-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3236-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3243-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3234-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

May 15, 2024Z-2274-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2277-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2304-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2283-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2310-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2290-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2284-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2275-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2301-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2299-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2280-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2281-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2307-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2306-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2303-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2308-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.