What AMERICAN CONTRACT SYSTEMS, INC. makes

American Contract Systems, Inc. manufactures single-use, non-medicated general surgical procedure kits designed for various surgical applications. These kits are intended for one-time use in operating room settings and include essential tools and instruments required for specific surgical procedures.

The company’s products are regulated under FDA classification as either Class I or Class II devices, with some falling under the non-classifiable (N) category. These devices are listed in the FDA’s Unique Device Identifier (UDI) database, ensuring traceability and compliance with U.S. regulatory requirements.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Information

Country
US
Address
—
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
010334675

Catalogue (21894)

Page 46 of 438
DeviceModel / ReferenceRegistriesClassStatus
Acs AMAH40N
FDA UDI
Class IActive
Acs HGTJ00I
FDA UDI
Class IActive
Acs OSSP27A
FDA UDI
Class IIActive
Acs RWCS71L
FDA UDI
Class IIActive
Acs LLSI16E
FDA UDI
Class IIActive
Acs FILC06M
FDA UDI
Class IIActive
Acs MMLE55F
FDA UDI
Class IIActive
Acs LAVH33AF
FDA UDI
Class IIActive
Acs CWTK81B
FDA UDI
Class IIActive
Acs SBCY03I
FDA UDI
Class IIActive
Acs CEEV16M
FDA UDI
Class IIActive
Acs MMBA55D
FDA UDI
Class IIActive
Acs FALE76N
FDA UDI
Class IIActive
Acs SPSA58N
FDA UDI
Class IIActive
Acs UDAA63AB
FDA UDI
Class IIActive
Acs SFLA01G
FDA UDI
Class IIActive
Acs MMKS15B
FDA UDI
Class IIActive
Acs PLDA60C
FDA UDI
Class IIActive
Acs LMDP36F
FDA UDI
Class IIActive
Acs LMGN03N
FDA UDI
Class IIActive
Acs IHDC30M
FDA UDI
NActive
Acs JEMJ10P
FDA UDI
Class IIActive
Acs EFGG21A
FDA UDI
Class IIActive
Acs MHCV43M
FDA UDI
Class IIActive
Acs ESFT85B
FDA UDI
Class IIActive
Acs SWPV46F
FDA UDI
Class IIActive
Acs HOTO66D
FDA UDI
Class IIActive
Acs AMRS25F
FDA UDI
Class IIActive
Acs WPHA64G
FDA UDI
Class IIActive
Acs ANLC35W
FDA UDI
Class IIActive
Acs ALEX17A
FDA UDI
Class IActive
Acs PCHD28D
FDA UDI
Class IActive
Acs UDGP61X
FDA UDI
Class IIActive
Acs LLOH31A
FDA UDI
Class IIActive
Acs NCOH13E
FDA UDI
Class IIActive
Acs DMPM42G
FDA UDI
Class IIActive
Acs TNCV11L
FDA UDI
Class IIActive
Acs SHTH38D
FDA UDI
Class IIActive
Acs HSAC62D
FDA UDI
Class IIActive
Acs YOSH26N
FDA UDI
Class IIActive
Acs CACS51F
FDA UDI
Class IIActive
Acs ISCS25H
FDA UDI
Class IIActive
Acs HKCU34A
FDA UDI
Class IIActive
Acs AHLE47B
FDA UDI
Class IActive
Acs EDOC08G
FDA UDI
Class IIActive
Acs LMEP23L
FDA UDI
Class IIActive
Acs LKPP02U
FDA UDI
Class IIActive
Acs UIGC24Y
FDA UDI
Class IIActive
Acs LMRE03R
FDA UDI
Class IIActive
Acs IHEP07O
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (100)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

American Contract Systems, Inc. has 190 recall records spanning from December 28, 2016, to March 19, 2026, with statuses recorded as open, classified, or terminated. The reasons cited include concerns about incomplete sterilization, improper handling of components during sterilization processes, incorrect expiration dates, labeling errors, insufficient aeration time, and deviations from sterilization specifications. Some recalls involved third-party products or components, such as syringes from Medline Industries, and the use of FDA-banned substances in powdered surgical gloves.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Mar 19, 2026Z-1981-2026—open, classified

Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.

Sep 4, 2025Z-0179-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0185-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0187-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0180-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0178-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0172-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0171-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0183-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0174-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0175-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0177-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0186-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0173-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0170-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0184-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0181-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0182-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Sep 4, 2025Z-0176-2026—open, classified

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Jul 14, 2025Z-2466-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2471-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2465-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2467-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2468-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2469-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2470-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2464-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Jul 14, 2025Z-2472-2025—open, classified

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Apr 21, 2025Z-1916-2025—open, classified

Test results for cast padding component included in procedure trays shows Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals exceeded residual limits, which may lead to a possible loss or lack of functionality that may lead to a delay or prolonged treatment/procedure/therapy.

Feb 6, 2025Z-1332-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1333-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1349-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1345-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1359-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1356-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1339-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1348-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1358-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1361-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1346-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1331-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1350-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1355-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1347-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1353-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1362-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1354-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1341-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1338-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1352-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1342-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1334-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1357-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1360-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1340-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1335-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1336-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1344-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1337-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1343-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 6, 2025Z-1351-2025—open, classified

ACS is unable to confirm product sterilization assurance requirements were met. Possible loss or lack of functionality may lead to delay or prolonged treatment.

Feb 4, 2025Z-1406-2025—open, classified

During an internal investigation, ACS was unable to confirm that the CODMAN component achieved the required Sterility Assurance Level and therefore is not suitable for sterilization according to ACS procedures.

Jan 31, 2025Z-1425-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1426-2025—open, classified

Sterility assurance with procedure trays

Jan 31, 2025Z-1424-2025—open, classified

Sterility assurance with procedure trays

Aug 7, 2024Z-3230-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3238-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3239-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3227-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3241-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3233-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3242-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3226-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3231-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3229-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3237-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3244-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3232-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3228-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3235-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3240-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3236-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3243-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Aug 7, 2024Z-3234-2024—open, classified

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

May 15, 2024Z-2274-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2277-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2304-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2283-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2310-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2290-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2284-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2275-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2301-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2299-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2280-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2281-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2307-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2306-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2303-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.

May 15, 2024Z-2308-2024—open, classified

Ethylene Oxide/Ethylene Chlorohydrin (EO/ECH) residuals for the cast padding component exceeded the residual limits for permanent exposure devices as indicated by ANSI/AAMI/ISO 10993-7/(R)2012.