AVID MEDICAL, INC.

US

Last updated May 24, 2026

What AVID MEDICAL, INC. makes

Avid Medical, Inc. makes various types of surgical procedure kits, including general surgical, orthopaedic, obstetrical/gynaecological, cardiothoracic, ENT, vascular catheterization support, urological, and neurosurgical kits. These kits can be medicated or non-medicated and may be designed for single-use. Examples of their devices include the AVID TruCustom and Halyard kits.

The company's portfolio is regulated under Class I and Class II devices and is listed in the FDA UDI registry. Avid Medical, Inc. is based in the US.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Frequently asked questions

What types of products does Avid Medical, Inc. manufacture?

Avid Medical, Inc. makes various surgical procedure kits, including general surgical, orthopaedic, and obstetrical/gynaecological kits, which can be medicated or non-medicated and may be designed for single-use.

Where is Avid Medical, Inc. based?

Avid Medical, Inc. is based in the US, as indicated by its country of origin.

Which regulatory registries list Avid Medical, Inc.'s devices?

Avid Medical, Inc.'s devices are listed in the FDA UDI registry, which tracks unique device identifiers for medical devices.

How are Avid Medical, Inc.'s devices classified?

Avid Medical, Inc.'s devices are classified under Class I and Class II, indicating the level of regulatory control required for each device type.

AI-generated from registry data and not verified by a human reviewer.

Information

Country
US
Address
—
LinkedIn
—
Facebook
—
Phone
+1(888) 564-7153
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
015623119

Catalogue (43260)

Page 3 of 500
DeviceModel / ReferenceRegistriesClassStatus
Halyard KAFB005-21_0002
FDA UDI
Class IIActive
Halyard SHND410-02_0003
FDA UDI
Class IIActive
Halyard NAMC448-01_0011
FDA UDI
Class IIActive
Halyard LMAT006-01_0004
FDA UDI
Class IIActive
AVID TruCustom SINA222-01_0002
FDA UDI
Class IIActive
AVID TruCustom BTDS003-02_0003
FDA UDI
Class IIActive
AVID TruCustom VANA019-01_0011
FDA UDI
Class IIActive
AVID TruCustom ADVC208-03_0003
FDA UDI
Class IIActive
AVID TruCustom MAAH010-06_0003
FDA UDI
Class IIActive
AVID TruCustom EDHN012-04_0004
FDA UDI
Class IIActive
AVID TruCustom SOME003-02_0009
FDA UDI
Class IActive
AVID TruCustom VARV033-02_0002
FDA UDI
Class IIActive
AVID TruCustom VTAM2500-03_0008
FDA UDI
Class IIActive
AVID TruCustom LGLF003-05_0001
FDA UDI
Class IIActive
AVID TruCustom ELLH045-07_0001
FDA UDI
Class IActive
AVID TruCustom LANK010-04_0002
FDA UDI
Class IIActive
Halyard USAR032-13_0007
FDA UDI
Class IIActive
AVID TruCustom ADVG049-07_0002
FDA UDI
Class IIActive
AVID TruCustom SAMM034-13_0008
FDA UDI
Class IIActive
AVID TruCustom VAMK012-05_0003
FDA UDI
Class IIActive
AVID TruCustom PORS004_0007
FDA UDI
Class IActive
AVID TruCustom ADBR006-08_0007
FDA UDI
Class IIActive
AVID TruCustom MRCM002-06_0004
FDA UDI
Class IIActive
Halyard RNTJ25-03_0002
FDA UDI
Class IIActive
Halyard WILF002-18_0026
FDA UDI
Class IIActive
Halyard BLPD65P_0001
FDA UDI
Class IIActive
Halyard SJGH112-04_0005
FDA UDI
Class IIActive
Halyard WRNM037-12_0009
FDA UDI
Class IIActive
AVID TruCustom FBCH130_0007
FDA UDI
Class IIActive
AVID TruCustom HAYS800_0007
FDA UDI
Class IIActive
AVID TruCustom KLRH011-03_0009
FDA UDI
Class IIActive
AVID TruCustom MAPS0004_0009
FDA UDI
Class IIActive
Halyard OGCV00-02_0002
FDA UDI
Class IIActive
Halyard VATX011-07_0017
FDA UDI
Class IIActive
AVID TruCustom SSBL003-02_0007
FDA UDI
Class IIActive
AVID TruCustom MMNP005-01_0007
FDA UDI
Class IIActive
AVID TruCustom KAFB023-01_0009
FDA UDI
Class IIActive
AVID TruCustom DGMC110-04_0003
FDA UDI
Class IIActive
AVID TruCustom OSUM148-04_0003
FDA UDI
Class IIActive
Halyard FBCH124-01_0005
FDA UDI
Class IIActive
AVID TruCustom EACH1001-20_0002
FDA UDI
Class IIActive
Halyard GVVA006-04_0004
FDA UDI
Class IIActive
Halyard MMNP005-01_0009
FDA UDI
Class IIActive
AVID TruCustom JFSN031-01_0005
FDA UDI
Class IIActive
AVID TruCustom MTHD004-06_0001
FDA UDI
Class IActive
AVID TruCustom OHIO006-13_0002
FDA UDI
Class IIActive
AVID TruCustom RDLD008-04_0014
FDA UDI
Class IActive
AVID TruCustom ADVT022-19_0004
FDA UDI
Class IIActive
AVID TruCustom SAMM033-11_0003
FDA UDI
Class IIActive
AVID TruCustom HUMC010-02_0009
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (100)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

The FDA recall records for Avid Medical, Inc. consist of 164 records, with the earliest initiated on 2005-09-09 and the most recent on 2026-03-13. The recorded statuses of these recalls include open, classified, and terminated. The reasons given for the recalls include issues with sterility, packaging, and component defects, such as recalled surgical gowns, loose metal flakes on components, and potential breaches in package integrity. The records also mention issues with specific products, including a latex-containing dressing and a swab stick collection device with an expiration date prior to the kit expiration date.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Mar 13, 2026Z-1971-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1978-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1976-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1973-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1975-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1979-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1977-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1974-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1972-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Mar 13, 2026Z-1970-2026—open, classified

Potential risk for Medline syringe rotating adaptor unwinding during use, which may result in a loose connection and/or full disconnection between the syringe and manifold.

Jan 2, 2026Z-1338-2026—open, classified

Tyvek bag seal issue which may compromise sterility of the kit.

Nov 14, 2025Z-0926-2026—open, classified

Potential for incomplete seal on header bag.

Nov 14, 2025Z-0927-2026—open, classified

Potential for incomplete seal on header bag.

Nov 14, 2025Z-0930-2026—open, classified

Potential for incomplete seal on header bag.

Nov 14, 2025Z-0929-2026—open, classified

Potential for incomplete seal on header bag.

Nov 14, 2025Z-0928-2026—open, classified

Potential for incomplete seal on header bag.

Nov 11, 2025Z-1000-2026—open, classified

Devices are not suitable for organ transplant.

Nov 11, 2025Z-1001-2026—open, classified

Devices are not suitable for organ transplant.

Nov 11, 2025Z-1002-2026—open, classified

Devices are not suitable for organ transplant.

Aug 4, 2025Z-2531-2025—open, classified

Potential for open header bag seals, compromising sterility.

Aug 4, 2025Z-2530-2025—open, classified

Potential for open header bag seals, compromising sterility.

Jun 11, 2025Z-2252-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2242-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2259-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2250-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2245-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2247-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2240-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2243-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2257-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2248-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2258-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2246-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2256-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2249-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2251-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2241-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2262-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2263-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2253-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2255-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2254-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2261-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2260-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Jun 11, 2025Z-2244-2025—open, classified

CLEARIFY VISUALIZATION SYSTEM was inadvertently resterilized in kits, causing discoloration.

Mar 31, 2025Z-1741-2025—open, classified

An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.

Mar 31, 2025Z-1743-2025—open, classified

An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.

Mar 31, 2025Z-1742-2025—open, classified

An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.

Sep 20, 2024Z-0433-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0436-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0389-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0427-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0394-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0397-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0416-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0452-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0418-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0432-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0415-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0398-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0383-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0446-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0453-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0451-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0408-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0437-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0434-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0442-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0428-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0425-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0423-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0454-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0390-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0419-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0395-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0411-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0405-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0447-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0439-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0450-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0403-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0417-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0388-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0387-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0391-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0429-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0385-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0438-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0449-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0401-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0407-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0413-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0445-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0441-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0400-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0455-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0414-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0404-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0435-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions

Sep 20, 2024Z-0402-2025—open, classified

Sponge forceps and towel clamps associated with small loose metal flakes potentially could detach from the components. Metal flakes on components may enter a patients surgical site undetected causing local reactions or possible foreign body reactions