  "format": "devicerts.manufacturer.v1",
  "cite_as": "Beta Bionics, Inc. deviCERTS medical device catalogue. https://devicerts.com/manufacturers/beta-bionics-inc (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/manufacturers/beta-bionics-inc",
    "json_url": "https://devicerts.com/manufacturers/beta-bionics-inc.json",
    "markdown_url": "https://devicerts.com/manufacturers/beta-bionics-inc.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA GUDID",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA registration and listing",
        "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfrl/rl.cfm"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "eudamed_record_synced": null,
    "fda_recalls_as_of": "2026-11-01",
    "registry_facts_computed": "2026-11-20",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/manufacturers/beta-bionics-inc",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "manufacturer": {
    "name": "Beta Bionics, Inc.",
    "url": "https://devicerts.com/manufacturers/beta-bionics-inc",
    "registered_as": null,
    "abbreviated_name": null,
    "also_registered_as": [],
    "country": "United States",
    "country_basis": "FDA applicant, unverified",
    "eudamed_status": null,
    "verified": false,
    "roles": [
      "Manufacturer"
    ],
    "registrations": [
      "FDA"
    ],
    "identifiers": [
      {
        "label": "FDA FEI",
        "value": "3019004087"
      },
      {
        "label": "DUNS",
        "value": "835742136"
      }
    ],
    "summary": "Beta Bionics, Inc. is a medical device manufacturer, with 41 devices listed in FDA GUDID and FDA 510(k) (FDA Class II and I) and 3 FDA recalls as of 20 Nov 2026.",
    "noindex": false
  },
  "counts": {
    "devices": 41,
    "devices_note": null,
    "certificates": 0,
    "fda_recalls_matched_by_name": 3,
    "fda_recalls_ongoing": 3,
    "fda_recalls_ongoing_recent": 3,
    "fda_recalls_ongoing_before": null,
    "fda_establishments": 0,
    "sites": 0
  },
  "registry_facts": {
    "covers": "Every device of the manufacturer. A device listed in two registries counts in both.",
    "registries": [
      {
        "registry": "FDA GUDID",
        "devices": 33
      },
      {
        "registry": "FDA 510(k)",
        "devices": 8
      }
    ],
    "device_classes": [
      {
        "class": "FDA Class II",
        "devices": 40
      },
      {
        "class": "FDA Class I",
        "devices": 1
      }
    ],
    "top_emdn_categories": []
  },
  "related": {
    "site_of": null,
    "part_of": null,
    "sites": [],
    "authorised_representatives": []
  },
  "devices": {
    "shown": 41,
    "total": 41,
    "all_devices_url": "https://devicerts.com/manufacturers/beta-bionics-inc",
    "items": [
      {
        "name": "iLet Insulin Infusion Kit Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-insulin-only-bb1017",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Starter Kit Insulin Only",
        "url": "https://devicerts.com/devices/ilet-starter-kit-insulin-only-bb1018",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-insulin-only-bb1019",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Flex Insulin Infusion Set 5-Pack (23\", 6mm) (BB1020) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-flex-insulin-infusion-set-5-pack-23-6mm-bb1020-insulin",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Clinical Data and Visualization Application BOM000042",
        "url": "https://devicerts.com/devices/clinical-data-and-visualization-application-bom000042-bb1029",
        "model_reference": null,
        "device_class": "Class I",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Flex Insulin Infusion Set 2-Pack (32\", 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-flex-insulin-infusion-set-2-pack-32-6mm-insulin-only-bb1028",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Dosing Decision Software",
        "url": "https://devicerts.com/devices/ilet-dosing-decision-software-qji",
        "model_reference": "K262763",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2026-09-11",
        "open_recall": false
      },
      {
        "name": "Mint ACE Pump",
        "url": "https://devicerts.com/devices/mint-ace-pump-qfg",
        "model_reference": "K260001",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2026-09-11",
        "open_recall": false
      },
      {
        "name": "iLet Bionic Pancreas Insulin Only",
        "url": "https://devicerts.com/devices/ilet-bionic-pancreas-bb1003-68f580af",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Starter Kit - Contact Detach Insulin Only",
        "url": "https://devicerts.com/devices/ilet-starter-kit-contact-detach-bb1008",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit – Inset™ (23\", 6mm) 10-Pack Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-inset-23-6mm-10-pack-bb4200",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit – Inset™ (32\") 10-Pack Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-inset-32-10-pack-bb4313",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Bionic Pancreas Insulin Only",
        "url": "https://devicerts.com/devices/ilet-bionic-pancreas-bb1002-32b7a238",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit - Inset (23”, 6mm) 15-Pack (Pharmacy Kit) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-inset-23-6mm-15-pack-pharmacy-kit-bb4210",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Bionic Circle App, Android BOM000083",
        "url": "https://devicerts.com/devices/bionic-circle-app-android-bb5002",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Connect Adapter 15-Pack (BB3052) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-connect-adapter-15-pack-bb3052-bb3052",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Bionic Pancreas Insulin Only",
        "url": "https://devicerts.com/devices/ilet-bionic-pancreas-bb1004-dc07bf13",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Inset Insulin Infusion Set (32”, 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-inset-insulin-infusion-set-32-6mm-bb4317",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Bionic Pancreas Insulin Only",
        "url": "https://devicerts.com/devices/ilet-bionic-pancreas-bb1002",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Cartridge Kit 15-Pack (1.6mL/160 Units) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-cartridge-kit-15-pack-1-6ml-160-units-bb2052",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Inset Insulin Infusion Set (23”, 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-inset-insulin-infusion-set-23-6mm-bb4311",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit – Contact Detach™ (23\"\", 6mm) - 20-Pack Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-contact-detach-23-6mm-20-pack-bb4310",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Contact Detach Insulin Infusion Set (23”, 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-contact-detach-insulin-infusion-set-23-6mm-bb1012",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Bionic Circle App, iOS BOM000082",
        "url": "https://devicerts.com/devices/bionic-circle-app-ios-bb5001",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Connect Adapter Insulin Only",
        "url": "https://devicerts.com/devices/ilet-connect-adapter-bb3010",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Cartridge Kit Insulin Only",
        "url": "https://devicerts.com/devices/ilet-cartridge-kit-bb2010",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Inset Insulin Infusion Set (23”, 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-inset-insulin-infusion-set-23-6mm-bb1010",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit – Contact Detach (23”, 6mm) 15-Pack (Pharmacy Kit) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-contact-detach-23-6mm-15-pack-phar-bb4300",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Contact Detach Insulin Infusion Set (23”, 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-contact-detach-insulin-infusion-set-23-6mm-bb4312",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Starter Kit - Inset 23\" Insulin Only",
        "url": "https://devicerts.com/devices/ilet-starter-kit-inset-23-bb1005",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Bionic Pancreas Insulin Only",
        "url": "https://devicerts.com/devices/ilet-bionic-pancreas-bb1003",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Insulin Infusion Kit - Inset (32”, 6mm) 15-Pack (Pharmacy Kit) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-insulin-infusion-kit-inset-32-6mm-15-pack-pharmacy-kit-bb4314",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Bionic Pancreas Insulin Only",
        "url": "https://devicerts.com/devices/ilet-bionic-pancreas-bb1004",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Starter Kit - Inset 32\" Insulin Only",
        "url": "https://devicerts.com/devices/ilet-starter-kit-inset-32-bb1006",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Listed",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet Inset Insulin Infusion Set (32”, 6mm) Insulin Only",
        "url": "https://devicerts.com/devices/ilet-inset-insulin-infusion-set-32-6mm-bb1011",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "iLet ACE Pump",
        "url": "https://devicerts.com/devices/ilet-ace-pump-qfg-k253976",
        "model_reference": "K253976",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2026-04-30",
        "open_recall": false
      },
      {
        "name": "iLet® ACE Pump",
        "url": "https://devicerts.com/devices/ilet-r-ace-pump-qfg-k231485",
        "model_reference": "K231485",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2023-06-21",
        "open_recall": true
      },
      {
        "name": "iLet® Dosing Decision Software",
        "url": "https://devicerts.com/devices/ilet-r-dosing-decision-software-qji-k232224",
        "model_reference": "K232224",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2023-09-22",
        "open_recall": true
      },
      {
        "name": "iLet® Dosing Decision Software",
        "url": "https://devicerts.com/devices/ilet-r-dosing-decision-software-qji",
        "model_reference": "K220916",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2023-05-19",
        "open_recall": false
      },
      {
        "name": "iLet® ACE Pump",
        "url": "https://devicerts.com/devices/ilet-r-ace-pump-qfg",
        "model_reference": "K223846",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2023-05-19",
        "open_recall": false
      },
      {
        "name": "iLet ACE Pump",
        "url": "https://devicerts.com/devices/ilet-ace-pump-qfg",
        "model_reference": "K252770",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2025-09-29",
        "open_recall": true
      }
    ]
  },
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1996-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2026-03-12",
      "product_code": "QFG",
      "reason": "insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous\nGlucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for\nsensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia."
    },
    {
      "number": "Z-2513-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2025-09-23",
      "product_code": "QFG",
      "reason": "The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen  transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia."
    },
    {
      "number": "Z-1809-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2025-08-13",
      "product_code": "QFG",
      "reason": "it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery."
    }
  ]
}