FDA Recalls (39)
Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.
| Date | Recall No. | Class | Status | Reason |
|---|---|---|---|---|
| Nov 7, 2024 | Z-0677-2025 | — | open, classified | Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result. |
| Dec 13, 2023 | Z-0815-2024 | — | open, classified | DiaSorin has identified that lots 136344, 136344A, 136344B, 136344C, 136520, 136520A, 136520B, 136520C, 136604, 136703, 136703A, 136703B, 136703C 136775of LIAISON HSV-2 Type Specific IgG assay may produce increased equivocal and/or false positive results. |
| Nov 23, 2020 | Z-1342-2021 | — | terminated | Potential for system to produce falsely elevated parathyroid hormone (PTH) results from patient samples. |
| Jul 17, 2020 | Z-2936-2020 | — | open, classified | Printed label inside the box lid of the Parvovirus has the incorrect Lot Specific Constant value |
| Jul 13, 2016 | Z-0645-2017 | — | terminated | DiaSorin is issuing an urgent medical device correction for the LIAISON Estradiol II Gen assay. Due to the risk of a recently identified cross reactivity, the LIAISON Estradiol II Gena assay should not be used to test patients being treated with fulvestrant (Faslodex). |
| Mar 16, 2015 | Z-1464-2015 | — | terminated | Instability in the PRO-Trac II Tacrolimus ELISA kit which may cause falsely elevated patient results. |
| Aug 19, 2010 | Z-1744-2011 | — | terminated | Through internal testing, DiaSorin has become aware that the kit lots identified (ETI-AB-AUK PLUS kits, catalog number P001931, lot numbers 0650270A and/or 0650270A/1) may not consistently exhibit stability to defined expiration dating, resulting in decreased binding and lower absorbance. Internal testing has confirmed that when defined assay validation criteria pass, analytical performance is co |
| Dec 16, 2009 | Z-0922-2010 | — | terminated | DiaSorin Rubella IgG kits of lot 123127A were packed with one coated microtiter plate instead of two microtiter plates as indicated in the product labeling. The risk is the possibility of delayed patient test results. |
| Oct 12, 2009 | Z-1914-2010 | — | terminated | The kit control provided with the GammaCoat Plasma Renin Activity kit was labeled with the incorrect range. |
| Dec 5, 2008 | Z-1042-2009 | — | terminated | DiaSorin determined that there was a potential for a short to occur in a a residual starter pump electrical cable in the LIAISON (R) Analyzer. |
| Feb 1, 2008 | Z-1240-2008 | — | terminated | Inaccurate measurements: Kits may sporadically recover kit and external control values out of range high. |
| Aug 21, 2007 | Z-1237-2007 | — | terminated | 25 Hydroxyvitamin D RIA kits may contain the incorrect lot of tracer. |
| Mar 14, 2007 | Z-0755-2007 | — | terminated | A lot to lot inconsistency was noted on kits of ETI EBNA G. |
| Feb 15, 2007 | Z-0748-2007 | — | terminated | Assay Failure: Assay failures may occur due to the kit control running outside of the established range (invalidating the assay) |
| Jun 5, 2006 | Z-0707-2007 | — | terminated | Potential for the DiaSorin Kit 1,25-Dihydroxyvitamin D RIA Kit Control 1 (lot 548520) & 2 (lot 548521) to recover out of the defined range, LOW. (If one or both of the kit controls recover outside the defined range, the run is considered invalid) |
| Aug 8, 2005 | Z-0557-06 | — | terminated | Potential for false negative results. A software anomaly may occur when performing the ETI-Mumps IgG assay on the ETI-Max 3000 automated platform. The error occurs when running a combined plate on the ETI-Max 3000, with the ETI-Mumps IgG assay as the second assay on a combined plate with ETI-Measles IgG and/or ETI-VZV IgG. |
| Apr 8, 2005 | Z-0776-05 | — | terminated | The package insert for Hepatitis A Virus IgM antibody kit (ETI-IgMK Plus, catalog P001925), may inadvertently contain pages 5 through 12 from the instructions for Hepatitus A virus Total antibody test (ETI-AB-HAVK Plus, catalog P001926). |
| Mar 16, 2005 | Z-0784-05 | — | terminated | Package insert incorrectly states that specimen diluent is ready to use, however specimen diluent bottle in kit states that the diluent must be diluted to one liter. |
| Feb 14, 2005 | Z-0637-05 | — | terminated | Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values. |
| Feb 14, 2005 | Z-0638-05 | — | terminated | An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data. |
| Jan 18, 2005 | Z-0706-05 | — | terminated | Expiration date of 2005/1/6 stamped on vial label of Anti-Human igG, Fluorescein Conjugate is incorrect. Correct expiration date is 2006/1/6. |
| Jan 10, 2005 | Z-0636-05 | — | terminated | QC testing of the Antibody Reagent Set for Lp(a) SPQ Test System it was discovered that the background was failing high of the specification. Through substitution testing it was determined that the Lp(a) diluent, part #238017 was the cause of the failing background specification. |
| Nov 1, 2004 | Z-0927-05 | — | terminated | DiaSorin Anti-Cardiolipin IgA Kit, catalog no. 7800, positive control was on the low limit of the acceptable range (16-35 U/mL). |
| Nov 1, 2004 | Z-0658-05 | — | terminated | The DiaSorin ETI-EA-G kits contained the wrong microtiter plate. The plate provided in the kits was for a different product, the ProTrac II Tacrolimus kit. |
| Oct 26, 2004 | Z-0674-05 | — | terminated | Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml. |
| Oct 20, 2004 | Z-0633-05 | — | terminated | 2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50. The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail. |
| Sep 24, 2004 | Z-0807-05 | — | terminated | A problem with software for the PolyTiter Immunofluorescent Titration System in that no flags or warnings appear when PolyTiter Calibrator Values are re-entered or changed (including values changed to unacceptable values) after the initial curve has been generated, if the 'refresh button' has not been pressed. The graph may appear valid; however erroneous results may occur. |
| Jul 15, 2004 | Z-0634-05 | — | terminated | Standards used in the GammaCoat Androstenedione Radioimmunoassay Kit appeared yellowish in color and were gelatinous upon reconstitution. |
| Jun 30, 2004 | Z-0676-05 | — | terminated | The program for a specific immunofluorescent titration system (Polymedco PolyTiter immunofluorescent Titration System) was found faulty in that the calibrator 2 value was incorrectly programmed as 1:120 instead of 1:160. |
| Jun 30, 2004 | Z-0705-05 | — | terminated | An incorrect expiration date of 2005/08/18 was placed on the kit box label for 25-hydroxyvitamin D [125I] RIA kits, Catalog 68100E, Lot 113300B. Correct label expiration date is 2004/08/18. |
| May 25, 2004 | Z-0704-05 | — | terminated | Standards contained in the GammaCoat Androstenedione RIA Kit appeared yellowish in color and were gelatinous upon reconstitution. The assay validation criteria failed when the gelatinous standards were used. |
| May 20, 2004 | Z-0703-05 | — | terminated | When software was updated for the Washer program of the Bio Tek Automated Microplate Washer Model ELx50, the software did not automatically clear old versions of the wash program when new ones were added. |
| Apr 20, 2004 | Z-0672-05 | — | terminated | An error in the Bio-Tek ELx800 microplate reader programming specifications for the ENA screen Assay exists in that an equivocal (borderline) zone is missing. |
| Apr 12, 2004 | Z-0675-05 | — | terminated | The data reduction program written for a specific automated microtiter plate reader (Bio-Tek ELx800) was found to be faulty. This program is used to calculate the cut off values for the Diastat Anti-Thyroglobulin Kit assay. The negative Control Upper Limit should be less than 0.95, rather than less than 1.0. |
| Apr 12, 2004 | Z-0702-05 | — | terminated | The software program written for use in the automated microtiter plate reader (Bio-Tek ELx800) for use with the Pro-Trac II Tacrolimus ELISA kit was found to be faulty. Assay software program did not include the Calibrator Zero OD specification (STDO>=1.500) as specified in the product insert. |
| Apr 12, 2004 | Z-0673-05 | — | terminated | The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed. |
| Mar 9, 2004 | Z-1576-05 | — | terminated | An error in the acceptance specification for the substrate blank optical density was discovered in the Bio-Tek ELx800 programming specifications for the ORTHO HCV Version 3.0 ELISA Test System Assay. |
| Jan 23, 2004 | Z-0632-05 | — | terminated | DiaSorin Anti-Cardiolipin IgA kit's positive control is out of acceptable limits. |
| Sep 10, 2003 | Z-0572-05 | — | terminated | Varicella Zoster ELISA kit assay was producing invalid runs on the ETI-Max 3000 automated microtiter plate instrument and/or false negative results. |