Etac A/S

Active in EUDAMED

Denmarkfrom SRN EUDAMED FDA SRN DK-MF-000017724 FEI 3004137175

Etac A/S is a medical device manufacturer based in Denmark, with 373 devices listed in EUDAMED and FDA GUDID (EU Class I; FDA Class I) and 9 FDA recalls as of 16 Nov 2026.

Devices
373
Registrations
EUDAMEDFDA
FDA recalls
94 ongoing since 2021 (6 open in total)
2 open since before 2021

Registry facts

Counted by deviCERTS from registry records, 16 Nov 2026

Registries
  • EUDAMED 351
  • FDA GUDID 175

A device listed in two registries counts in both.

Device classes
  • EU Class I 351
  • FDA Class I 22
Identifiers
  • EUDAMED SRN DK-MF-000017724
  • FDA FEI 3004137175
  • DUNS 305724502

At a glance

Counted by deviCERTS from registry records

Market reach

European Union
EUDAMED actor
United States
FDA FEI, FDA recalls

What Etac A/S makes

Etac A/S manufactures standing transfer platforms (manual), reusable device sterilization/disinfection containers, general-purpose patient lifting system slings/harnesses, and mobile patient lifting systems that are electrically powered. These products are marketed under the Molift brand name.

The devices are classified as Class I and are listed in the EUDAMED and FDA UDI registries. The company is based in Denmark.

AI-generated from registry data, 26 Oct 2026. Not verified by deviCERTS or the manufacturer.

Catalogue373

Newest first, rows 1-50 of 373

Search and filter all 373 devices
Skip device list

Rows 1-50 of 373

Showing 1-50 of 373

Company details

EUDAMED record synced 7 Nov 2026

EUDAMED SRN
DK-MF-000017724
FDA FEI number
3004137175
DUNS number
305724502
EU VAT number
10313082
EUDAMED status
Active
EUDAMED registered name
Etac A/S
Registered address
3, Parallelvej, DK-8751, Gedved, Denmark
Website
www.etac.dk
Phone
+45 79685833
PRRC contact
Vibeke Damgaard Nielsen(public in EUDAMED)

FDA recalls9

Recalls are matched to the firm by name. FDA recall data as of 1 Nov 2026.

FDA recall records list 4 recalls initiated by Etac A/S, initiated on 2025-11-17. By recorded status: 4 ongoing. The reasons filed fall into these groups: other reasons as filed. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.

Generated automatically from FDA recall records, as a summary of the records listed below. It describes what the records state and is not an assessment of this manufacturer.

Recalls initiated per year

  • 2017 0
  • 2018 0
  • 2019 2
  • 2020 0
  • 2021 0
  • 2022 0
  • 2023 0
  • 2024 0
  • 2025 4
  • 2026 0

3 initiated before 2017

  • 2025

    • 17 Nov 2025 Z-1308-2026 Class II Ongoing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment. Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland,…
      Full distribution text Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
    • 17 Nov 2025 Z-1309-2026 Class II Ongoing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment. Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland,…
      Full distribution text Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
    • 17 Nov 2025 Z-1310-2026 Class II Ongoing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment. Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland,…
      Full distribution text Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
    • 17 Nov 2025 Z-1311-2026 Class II Ongoing A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment. Distribution: Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland,…
      Full distribution text Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.
  • 2019

    • 4 Feb 2019 Z-1351-2019 Class II Open since 2019 The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall. Distribution: Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington,…
      Full distribution text Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
    • 4 Feb 2019 Z-1352-2019 Class II Open since 2019 The firm has received 7 reports of malfunction for the device in which the hooks have broken off, potentially leading the sling to slip out of the sling bar and causing the patient to fall. Distribution: Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington,…
      Full distribution text Domestic distribution to Arizona, Delaware, Florida, Georgia, Massachusetts, Ohio, South Carolina, Texas, Washington, and Wisconsin. Foreign distribution to Canada, Asia, Europe, and Australia.
  • 2016

    • 24 Jun 2016 Z-2291-2016 Class II Terminated Potential for the bolt on the suspension coupling on the Molift Air to come loose from the housing. Distribution: Worldwide Distribution - US (nationwide) and Internationally to SZ, TR, BE, CH, DE, DK, ES, FI, FR, GB, IS LV, NL, NO,…
      Full distribution text Worldwide Distribution - US (nationwide) and Internationally to SZ, TR, BE, CH, DE, DK, ES, FI, FR, GB, IS LV, NL, NO, SE, SI, AU, CN, IL, JP, NZ, IN, QA, PE, AE
  • 2013

    • 15 Feb 2013 Z-1342-2013 Class II Terminated The leg spreading mechanism will fail to hoist and collapse if assemble incorrectly by the user. Distribution: US Nationwide Distribution: CA, FL, GA, ID, IL, KY, MA, ME, MI, MN, NC, NH, NJ, NM, NV, NY, OH, OK, PA, TN, TX, VT, WA…
      Full distribution text US Nationwide Distribution: CA, FL, GA, ID, IL, KY, MA, ME, MI, MN, NC, NH, NJ, NM, NV, NY, OH, OK, PA, TN, TX, VT, WA and Hawaii. Foreign Distribution:: Australia, Austria; Belgium, Denmark, Finland, France, Germany, Iceland, Israel, Italy, Japan, Netherlands, New Zealand, Norway, Spain, Sverige, Switzerland, United Kingdom.
    • 15 Feb 2013 Z-1343-2013 Class II Terminated The leg spreading mechanism will fail to hoist and collapse if assemble incorrectly by the user. Distribution: US Nationwide Distribution: CA, FL, GA, ID, IL, KY, MA, ME, MI, MN, NC, NH, NJ, NM, NV, NY, OH, OK, PA, TN, TX, VT, WA…
      Full distribution text US Nationwide Distribution: CA, FL, GA, ID, IL, KY, MA, ME, MI, MN, NC, NH, NJ, NM, NV, NY, OH, OK, PA, TN, TX, VT, WA and Hawaii. Foreign Distribution:: Australia, Austria; Belgium, Denmark, Finland, France, Germany, Iceland, Israel, Italy, Japan, Netherlands, New Zealand, Norway, Spain, Sverige, Switzerland, United Kingdom.

Frequently asked questions

How many devices does Etac A/S have in the registry?

The registry currently lists 373 devices for Etac A/S. Devices recorded under merged duplicate company entries are counted once.

Which registries list Etac A/S's devices?

Among its 50 most recently registered devices, Etac A/S has devices listed in EUDAMED (28 devices) and FDA GUDID (38 devices). A device can be listed in more than one registry, so these counts overlap.

What risk classes do Etac A/S's devices fall under?

Among its 50 most recently registered devices, the classes recorded for Etac A/S's devices are Class I. Each device page shows its own class and authorisation.

Are there FDA recall records for Etac A/S?

FDA recall records list 9 recalls initiated by Etac A/S, the most recent initiated on 17 Nov 2025. 6 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.

What are the newest devices from Etac A/S in the registry?

The most recently registered devices listed for Etac A/S include Molift Partner 255, Molift Assist Belt S/M, Molift UnoSling HighBack S and Molift Soft Handles 2 pcs.