PENTAX of America, Inc.

United States·US-MF-000008135

Last updated September 17, 2026

Information

Country
United States
Address
3, Paragon Drive, Montvale, United States
LinkedIn
—
Facebook
—
Phone
—
PRRC Contact
Paresh Patel
EUDAMED SRN
US-MF-000008135
FDA FEI Number
1000080301
DUNS Number
—

Catalogue (368)

Page 8 of 8
DeviceModel / ReferenceRegistriesClassStatus
PENTAX Medical Video Colonoscope EC34-i10T Series K180285
FDA 510(k)
Class IIActive
Pentax Eg-3870utk Ultrasound Video Gastroscope + Hi Vision Preirus K130247
FDA 510(k)
Class IIActive
PENTAX Medical Video Upper GI Scopes (EG Family) K190799
FDA 510(k)
Class IIActive
PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Upper GI Scope EG29-i20c, PENTAX Medical Video Colonoscope EC38-i20cL K231249
FDA 510(k)
Class IIActive
C2 CryoBalloon Catheter (Pear), C2 CryoBalloon Catheter (Standard), C2 CryoBalloon Controller, C2 CryoBalloon Foot Pedal, C2 CryoBalloon Cartridge K203024
FDA 510(k)
Class IIActive
PENTAX Medical Video Upper GI Scopes (EG Family) (EG Family); PENTAX Medical Video Colonoscopes (EC Family) (EC Family) K251256
FDA 510(k)
Class IIActive
PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U K183516
FDA 510(k)
Class IIActive
PENTAX Medical Video Upper GI Scope (EG27-i20c); PENTAX Medical Video Colonoscope (EC34-i20cL) K241213
FDA 510(k)
Class IIActive
PENTAX Medical Video Upper GI Scopes (EG Family), PENTAX Medical Video Colonoscopes (EC Family) K210485
FDA 510(k)
Class IIActive
Pentax Eg-3670urk Ultrasound Video Gastroscope + Hi Vision Preirus K130206
FDA 510(k)
Class IIActive
Pentax Medical Video Duodenoscope ED32-i10 K202365
FDA 510(k)
Class IIActive
PENTAX Medical EPK-3000 Imaging System with Camera Head PVK-J10, PENTAX Medical ENT Imaging System with Camera Head PVK-J10 K220465
FDA 510(k)
Class IIActive
PENTAX Medical Video Bronchoscope EB-J10 Series K200678
FDA 510(k)
Class IIActive
Nasometer 6500
EUDAMEDFDA UDI
Class IActive
Laryngeal Strobe 9400 9400
EUDAMEDFDA UDI
Class IActive
Computerized Speech Lab (CSL) 4500b 4500b
EUDAMEDFDA UDI
Class IActive
3950c Visi-Pitch 3950c
EUDAMEDFDA UDI
Class IActive
Laryngeal Strobe 9500 9500
EUDAMEDFDA UDI
Class IActive

Related Companies

Authorised Representatives

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (33)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Jul 16, 2025Z-2329-2025—open, classified

During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.

Jan 29, 2025Z-1237-2025—open, classified

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Jan 29, 2025Z-1238-2025—open, classified

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Jan 29, 2025Z-1236-2025—open, classified

During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.

Jul 14, 2021Z-2449-2021—open, classified

Updated Reprocessing Instruction For Use (rIFU): Impacted PENTAX Medical Video Upper GI Scopes (EG) and Video Colonoscopes (EC) families

Jul 14, 2021Z-2448-2021—open, classified

Updated Reprocessing Instruction For Use (rIFU): Impacted PENTAX Medical Video Upper GI Scopes (EG) and Video Colonoscopes (EC) families

Apr 21, 2021Z-1682-2021—terminated

IFU updated: The warning section of the IFU for both the OE-A63 distal end cap and ED34-i10T2 duodenoscope has been updated to notify users of the associated risks with the distal end cap (OE-A63) unexpectedly becoming detached during a procedure and updated to notify users of what immediate actions should be taken in case the event occurs. This can result in unforseen events such as mucosal injury, lacerations, or bleeding of the patient. Detachment of the distal end cap (OE-A63) into the oral cavity of the patient may also result in aspiration

Sep 22, 2020Z-0293-2021—terminated

There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.

Sep 22, 2020Z-0292-2021—terminated

There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.

Jan 3, 2020Z-1268-2020—terminated

Distributed in the USA without an approved 510K

Jan 3, 2020Z-1267-2020—terminated

Distributed in the USA without an approved 510K

Dec 3, 2018Z-0970-2019—terminated

The Controller does not detect overpressure in the balloon during the application of non-dosing puffs of Nitrous Oxide, which can contribute to balloon over pressurization, if the intended vent lumen of the catheter is significantly occluded to prevent relieving balloon pressure due to a kinked catheter condition.If a patient is exposed to higher than physiologic pressures, adverse events such as perforation or mucosal laceration may occur.

Dec 3, 2018Z-1326-2019—terminated

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Dec 3, 2018Z-1325-2019—terminated

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Dec 3, 2018Z-1327-2019—terminated

Incorrect default dose and dose increments may be transmitted to the controller if the catheter RFID tag is not correct or is not being correctly or completely readable by the controller.

Mar 30, 2018Z-2256-2018—terminated

The catheter alignment arrow printed onto the strain relief is not aligned with the detachment feature of the catheter connector, resulting in the potential for the catheter to be attached to the controller in the detach orientation. In this orientation, the catheter is not securely attached to the controller and may be partially (~1cm) ejected from controller.

Feb 13, 2018Z-0643-2018—terminated

The duodenoscopes are being recalled in order to replace the forceps elevator mechanism, the O-rings, and the distal end covering to be consistent with the updated design as well as provide an updated periodic inspection as part of the Operation Manual in order to mitigate the potential risk of infection in flexible endoscopy.

Jul 13, 2017Z-0615-2018—terminated

The video cytoscopes lack 510(k) premarket notification clearance.

May 23, 2017Z-3068-2017—terminated

Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

May 23, 2017Z-3067-2017—terminated

Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

Apr 13, 2017Z-2742-2017—terminated

Pentax Medical did not always provide transformers with 9175 isolation transformers are used with 7245C, 7245C/E, 7245D, 9200Cs, 9200Ds, 9310HDs, and 9400s computer systems.

Apr 13, 2017Z-2744-2017—terminated

Pentax Medical did not always provide transformers with 9175 isolation transformers are used with 7245C, 7245C/E, 7245D, 9200Cs, 9200Ds, 9310HDs, and 9400s computer systems.

Apr 13, 2017Z-2743-2017—terminated

Pentax Medical did not always provide transformers with 9175 isolation transformers are used with 7245C, 7245C/E, 7245D, 9200Cs, 9200Ds, 9310HDs, and 9400s computer systems.

Apr 13, 2017Z-2745-2017—terminated

Pentax Medical did not always provide transformers with 9175 isolation transformers are used with 7245C, 7245C/E, 7245D, 9200Cs, 9200Ds, 9310HDs, and 9400s computer systems.

Jan 17, 2017Z-2713-2017—terminated

Pentax initiated a field correction/safety alert for two (2) models of the Video Duodenoscope to determine how soiling may have occurred on the surface of the suction cylinder and under the distal cap during testing.

Dec 12, 2016Z-1075-2017—terminated

Pentax is initiating a Field Correction to revise the Instructions for Use of the Rigid Laryngostroboscopes Model 9106 and 9108.

Aug 15, 2016Z-0318-2017—terminated

PENTAX Medical is initiating this field action to provide customers that have purchased the affected endoscopes with the most recent Operation and Reprocessing IFUs.

Aug 15, 2016Z-0636-2017—terminated

Pentax is initiating the product correction of several model numbers of Pentax's Colonoscope that were shipped between the clearance date of April 2014 and September 2015 because they contain the operation and reprocessing IFU's that were in use prior to the 2014 clearances and therefore did not include the cleared operation IFU, the updated reprocessing IFU or the dedicated reprocessing adaptor (for applicable models).

May 31, 2016Z-2875-2016—terminated

Pentax America Inc. is recalling various bronchoscopes which contain an incorrect Operation and Reprocessing IFUS.

May 31, 2016Z-0757-2017—terminated

The EB-1970UK Ultrasound Video Bronchoscope shipped between the clearance date of April 2014 and September 2015 contained the Operation and Reprocessing IFUs that were in use prior to the 2014 clearance, and therefore did not include the cleared Operation and Reprocessing IFUs.

Jan 11, 2011Z-2325-2017—terminated

This field action serves to retrospectively document the actions that were taken by Pentax to correct the loose suction arms on eleven (11) models of endoscopes and to notify FDA of these actions taken in 2010-2011.

Jan 11, 2011Z-2326-2017—terminated

This field action serves to retrospectively document the actions that were taken by Pentax to correct the loose suction arms on eleven (11) models of endoscopes and to notify FDA of these actions taken in 2010-2011.

Jan 11, 2011Z-2324-2017—terminated

This field action serves to retrospectively document the actions that were taken by Pentax to correct the loose suction arms on eleven (11) models of endoscopes and to notify FDA of these actions taken in 2010-2011.