FDA Recalls (2)
Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.
| Date | Recall No. | Class | Status | Reason |
|---|---|---|---|---|
| Aug 9, 2012 | Z-2321-2012 | — | terminated | The firm was notified that a malfunction of the device was caused by an incorrect fuse 32 volt 15 amp fuses instead of 250 volt 15 amp fuses, that was installed in the water system. The incorrect fuse may result in excessive heat in the fuse holder eventually resulting in failure of the device to operate. |
| Sep 3, 2004 | Z-0004-05 | — | terminated | The device has a potential risk of failure caused by faulty wiring within the electrical component. The wires can overheat and eventually break in two. |