Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

Skytron, LLC

US

Last updated September 17, 2026

Information

Country
US
Address
5085 Corporate Exchange Blvd SE, Grand Rapids, MI, 49512
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
1825014
DUNS Number
—

Catalogue (1084)

Page 10 of 22
DeviceModel / ReferenceRegistriesClassStatus
Skytron 4-030-10-B
FDA UDI
Class IActive
Skytron 1-011-03-4
FDA UDI
Class IActive
Skytron Moderate Alkaline Detergent SKY-110SCLF-1
FDA UDI
Class IIActive
Skytron 4A-PUS-17 Freedom
FDA UDI
Class IActive
Skytron 5-010-21-2B
FDA UDI
Class IActive
Skytron Smoke Evacuation ELVV220
FDA UDI
Class IIActive
Skytron warming cabinet W2-652427
FDA UDI
Class IIActive
Skytron Scope-Zyme Multi Enzymatic Cleaner SKY-405SCLF-15
FDA UDI
Class IIActive
Skytron 4-030-10-3WS
FDA UDI
Class IActive
Hyperlight Transfer Board 4-030-44
FDA UDI
Class IActive
X-Ray Head Section 5-010-02-1B
FDA UDI
Class IUnknown
Skytron 1-011-06-21
FDA UDI
Class IActive
Skytron 4A-BUD Freedom
FDA UDI
Class IActive
Skytron 3-041-01
FDA UDI
Class IActive
Skytron 24" back cover E3 H8-300-09-A
FDA UDI
Class IActive
Skytron 1-011-13-4
FDA UDI
Class IActive
Radiographic top 5-010-21-B
FDA UDI
Class IUnknown
Skytron 4A-ERBA Freedom
FDA UDI
Class IActive
CleanStart CSTART-9625
FDA UDI
Class IIActive
Sterling P/M1000 Flexible Rack for washing in 4 levels 1000-FLX4
FDA UDI
Class IIActive
Skytron vac slide rail mt H8-200-07
FDA UDI
Class IActive
Sterling P1800 1800-CON
FDA UDI
Class IIActive
Skytron 1-011-06-3-BOS
FDA UDI
Class IActive
Skytron F222 SERIES Freedom
FDA UDI
Class IActive
Skytron 4A-CS-49 Freedom
FDA UDI
Class IActive
Skytron 5-010-21-BC
FDA UDI
Class IActive
CleanStart CSTART-7125
FDA UDI
Class IIActive
Skytron front rail 10" E3 H8-300-08-A
FDA UDI
Class IActive
Skytron 4-030-44-3WS
FDA UDI
Class IActive
Skytron Cart Rinse and Drying Agent SKY-600-5
FDA UDI
Class IIActive
SkyVision SDS
FDA UDI
Class IActive
CleanStart CSTART-8501
FDA UDI
Class IIActive
Lucina 4 Examination Lights LCN4-PKG1
FDA UDI
Class IActive
Skytron warming cabinet W1-252427
FDA UDI
Class IIActive
Skytron 4A-HAB Freedom
FDA UDI
Class IActive
Sterling C2000 2002205
FDA UDI
Class IIActive
Skytron 4A-CA-UB Freedom
FDA UDI
Class IActive
Kub Board 4-040-09-7
FDA UDI
Class IActive
Sterling P1800 1800-FLX-SH
FDA UDI
Class IIActive
Skytron Sterling Lubricant SKY-300SCLF-15
FDA UDI
Class IIActive
Spectra Examination Light Series AUT1B
FDA UDI
Class IActive
Skytron 1-011-08-33
FDA UDI
Class IActive
Skytron 1-011-06-3B
FDA UDI
Class IActive
Skytron 4-041-01
FDA UDI
Class IActive
Uro drain tray with mounting brackets 4-040-07
FDA UDI
Class IActive
Skytron 5-010-02-2B
FDA UDI
Class IUnknown
Aurora Surgical Light Series Aurora II
FDA UDI
Class IIUnknown
Skytron 4-041-01-5WS
FDA UDI
Class IActive
Skytron warming cabinet W1-362421
FDA UDI
Class IIActive
Skytron 16.5" pedal holder H8-300-13
FDA UDI
Class IActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (2)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Aug 1, 2023Z-2469-2023—open, classified

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Aug 1, 2023Z-2470-2023—open, classified

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.