Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

Skytron, LLC

US

Last updated September 17, 2026

Information

Country
US
Address
5085 Corporate Exchange Blvd SE, Grand Rapids, MI, 49512
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
1825014
DUNS Number
—

Catalogue (1084)

Page 14 of 22
DeviceModel / ReferenceRegistriesClassStatus
Skytron 4-031-02-4WS
FDA UDI
Class IActive
Radiographic top 5-010-19-B
FDA UDI
Class IUnknown
Skytron Acidic Neutralizer SKY-200-1
FDA UDI
Class IIActive
Skytron 1-011-04-23
FDA UDI
Class IActive
Skytron 1-011-15-4
FDA UDI
Class IActive
Skytron 1-011-08-22
FDA UDI
Class IActive
Sterling Washers P1500-B4L-BWL
FDA UDI
Class IIActive
Anesthesia screen 6-020-02-1
FDA UDI
Class IActive
Skytron Streamline Progressa bed mount HRC-PRO
FDA UDI
Class IActive
I-Series Washers I3-PR
FDA UDI
Class IIActive
Skytron 4-031-03-4WS
FDA UDI
Class IActive
Sterling M1000 M1000-D208E
FDA UDI
Class IIActive
Head end uro tray adapter 4-040-06-1
FDA UDI
Class IActive
Auxiliary base connector kit D5-032-00
FDA UDI
Class IActive
Skytron 3-040-12-4WS
FDA UDI
Class IActive
Skytron QUIKLINK
FDA UDI
Class IIActive
SkyVac SES1
FDA UDI
Class IIActive
CleanStart CSTART-9205
FDA UDI
Class IIActive
Skytron 22" shelf PMSH
FDA UDI
Class IActive
TV-3 Camera Control Unit Aurora II, 3, and Four Series B1-710-59
FDA UDI
Class IUnknown
Skytron 1-011-03-5
FDA UDI
Class IActive
Skytron 1-011-08-5
FDA UDI
Class IActive
Sterling M1000 M1000-S480S
FDA UDI
Class IIActive
Sterling C2000 C2000L-E480
FDA UDI
Class IIActive
Sterling P/M1000 High speed module, 480V 1000-HSM-480
FDA UDI
Class IIActive
Skytron 16.5" nitro faceplate H8-300-16-2
FDA UDI
Class IActive
Skytron 5-010-23-1B
FDA UDI
Class IActive
Skytron 2-010-07-3WS
FDA UDI
Class IActive
Lucina Control Wand B5-014-39
FDA UDI
Class IActive
CleanStart CSTART-9605
FDA UDI
Class IIActive
Skytron 4A-EXU-12 Freedom
FDA UDI
Class IActive
Skytron 5-010-18-B
FDA UDI
Class IUnknown
Skytron 4A-SHB-22 Freedom
FDA UDI
Class IActive
Skytron 1-011-08-32
FDA UDI
Class IActive
Skytron 4-030-40-3S
FDA UDI
Class IActive
Skytron warming cabinet W1-653027
FDA UDI
Class IIActive
Skytron 22" pullout KB tray w/mouse H8-300-27
FDA UDI
Class IActive
Skytron 5-010-19-B
FDA UDI
Class IActive
Flex FLEX-CT6
FDA UDI
Class IActive
Skytron 1-011-17-2
FDA UDI
Class IActive
Sterling P/M1000 Rack 1000-BB-250
FDA UDI
Class IIActive
Sterling Washers P1500-AN
FDA UDI
Class IIActive
Skytron 4-041-01-3WS
FDA UDI
Class IActive
Foot rests for split leg (pr.) 4-030-04
FDA UDI
Class IActive
Skytron warming cabinet W1-252421
FDA UDI
Class IIActive
SkyVision Ascend Eco
FDA UDI
Class IActive
Skytron 4-030-40-4S
FDA UDI
Class IActive
Skytron 4-031-04
FDA UDI
Class IActive
E-Z Grip Bar Clamp 6-030-05
FDA UDI
Class IActive
Sterling P1800 1800-HSM-480
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (2)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Aug 1, 2023Z-2469-2023—open, classified

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Aug 1, 2023Z-2470-2023—open, classified

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.