Beckman Coulter Mishima K.K.
Country not established FDA establishment FEI 3002807663
- Devices
- 0
- Registrations
- FDA
- FDA recalls
- 53 ongoing
Company details
- FDA FEI number
- 3002807663
- Address
- 454-32 HIGASHINO,NAGAIZUMI-CHO, SUNTO-GUN Shizuoka
FDA recalls
Recalls initiated per year
- 2017 0
- 2018 0
- 2019 0
- 2020 0
- 2021 0
- 2022 0
- 2023 0
- 2024 2
- 2025 2
- 2026 1
2026
- 3 Aug 2026 Z-3332-2026 Class II Ongoing Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
Full distribution text
- 3 Aug 2026 Z-3332-2026 Class II Ongoing Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.
2025
- 30 Apr 2025 Z-1907-2025 Class II Completed A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.
Full distribution text
- 30 Apr 2025 Z-1908-2025 Class II Completed Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.
Full distribution text
- 30 Apr 2025 Z-1907-2025 Class II Completed A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.
2024
- 10 Jul 2024 Z-2691-2024 Class II Ongoing Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Full distribution text
- 5 Jun 2024 Z-2455-2024 Class II Ongoing There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.
Full distribution text
- 10 Jul 2024 Z-2691-2024 Class II Ongoing Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.