Biosense Webster Inc
United States US-MF-000014219 Last updated 23 Oct 2026
Is this your brand? Sign in to claim this brand
- Devices
- 673
- Registries
- EUDAMEDGUDIDFDA
- FDA recalls
- 624 ongoing
Catalogue
Webster Electrophysiology Catheter Ref 36S31Q
Webster Electrophysiology Catheter Ref 36E30Q
Carto Vizigo D138505 Ref D138505
Carto Vizigo D138504 Ref D138504
Webster Electrophysiology Catheter Ref 36F30Q
TX eco Cable Ref D140103
nGEN Ref D138401
Webster Electrophysiology Catheter Ref 37U00R
Myostar Injection Catheter Ref 121119S
nGEN Ref D138405
Myostar Injection Catheter Ref 121120S
TRUPULSE Generator Ref D141701
nGEN Pump Ref D139701
Carto 3 System Ref FG540000U
Webster Electrophysiology Catheter Ref 35Y13Q
Webster Electrophysiology Catheter Ref 36S30Q
Webster Electrophysiology Catheter Ref 36C33R
Halo D6T20282RT Ref D6T20282RT
N/A OF63X4C010FS Ref OF63X4C010FS
Celsius D7TCFG252RT Ref D7TCFG252RT
Webster D5S06AL252RT Ref D5S06AL252RT
Webster D5SAL252RT Ref D5SAL252RT
N/A D7EL252RT Ref D7EL252RT
Lasso D7L2015RT Ref D7L2015RT
ThermoCool SF NAV BNI35BBH Ref BNI35BBH
Soundstar SNDSTR10 Ref SNDSTR10
ThermoCool SF NAV D131504 Ref D131504
Qdot Micro D139403 Ref D139403
EZ Steer ThermoCool NAV BNI75TCJJH Ref BNI75TCJJH
ThermoCool SmartTouch D132705 Ref D132705
Webster F5QD252RT Ref F5QD252RT
Webster 107875S Ref 107875S
N/A D7DL252PS Ref D7DL252PS
N/A D7AL252PS Ref D7AL252PS
Pentaray Nav D128212 Ref D128212
Webster F6BF010RT Ref F6BF010RT
N/A D7PSL252RT Ref D7PSL252RT
Thermocool Smarttouch Sf D134703 Ref D134703
Celsius DS D7TCF8L162RT Ref D7TCF8L162RT
Thermocool Smarttouch Sf D134805 Ref D134805
OCTARAY Mapping Catheter D160906 Ref D160906
Webster Cs D135305 Ref D135305
Showing 1-50 of 673
EUDAMED certificates
Notified body 0344 2 certificates
- 3903009CE01 Reissued
Quality management system EU MDR
- Issued
- 1 Sep 2026
- Expires
- 1 Sep 2031
- 3903009CE07 Reissued
Quality management system EU MDR
- Issued
- 1 Jun 2026
- Expires
- 1 Jun 2031
Notified body 2797 1 certificate
- MDR 825396 Supplemented
Technical documentation EU MDR
- Issued
- 22 Jun 2026
- Expires
- 26 Jan 2031
- Reason
- Change approved
Related companies
Authorised representatives
Notified bodies
- Notified body 0344
- Notified body 2797
FDA recalls
2025
- 5 Jan 2025 Z-1205-2025 Class I Completed Due to an observed trend of neurovascular events
2024
- 16 Feb 2024 Z-1290-2024 Class II Ongoing Due to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.
2023
- 21 Feb 2023 Z-1319-2023 Class II Ongoing Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch
- 21 Feb 2023 Z-1320-2023 Class II Ongoing Incorrect labeling; specifically, the product name, lot number and product code on the carton label for Bi-Directional Guiding Sheaths does not match the inner label pouch
2022
- 29 Jun 2022 Z-1677-2022 Class II Ongoing Outside packaging for a device indicated for introducing cardiovascular catheters into the heart, sustained water damage, which may have affected the sterile barrier of the product inside. Devices with a compromised sterile barrier could be a source for infection.
2021
- 26 Feb 2021 Z-1285-2021 Class II Completed There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism
- 26 Feb 2021 Z-1286-2021 Class II Completed There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism
- 26 Feb 2021 Z-1287-2021 Class II Completed There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism
2020
- 22 Apr 2020 Z-2371-2020 Class II Terminated Mapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
- 22 Apr 2020 Z-2372-2020 Class II Terminated Mapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.
Show the 25 most recent of 62 recalls
2018
- 7 May 2018 Z-2878-2018 Class I Terminated When attempting to recapture the 32-mm device, the tip of the delivery sheath may fold or buckle, resulting in increased retraction force and difficulty or failure to recapture the device.
2017
- 6 Sep 2017 Z-0054-2018 Class II Terminated Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- 6 Sep 2017 Z-0055-2018 Class II Terminated Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- 6 Sep 2017 Z-0056-2018 Class II Terminated Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- 6 Sep 2017 Z-0057-2018 Class II Terminated Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- 6 Sep 2017 Z-0058-2018 Class II Terminated Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.
- 14 Jun 2017 Z-0210-2018 Class II Terminated Generator may inadvertently deliver RF energy if the foot pedal is placed in close proximity and oriented toward the Stereotaxis magnets.
- 17 May 2017 Z-0211-2018 Class II Terminated Contraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
- 17 May 2017 Z-0212-2018 Class II Terminated Contraindication language updated in the Instructions For Use (IFU) to indicate that Pentaray Catheters should not be used in patients with prosthetic valves.
2016
- 15 Aug 2016 Z-2776-2016 Class II Terminated Biosense Webster, Inc. is initiating a voluntary correction to re-install identified CARTO 3 Systems that were moved and installed in China without testing and approval by Biosense Webster Technical Services, as described in the Instructions for Use (IFU).
- 14 Apr 2016 Z-1590-2016 Class II Terminated Biosense Webster is recalling the Pentaray Catheters to clarify the contraindication language in the Instructions For Use (IFU) and product labeling for this catheter relative to patients with prosthetic valves.
- 14 Apr 2016 Z-1591-2016 Class II Terminated Biosense Webster is recalling the Pentaray Catheters to clarify the contraindication language in the Instructions For Use (IFU) and product labeling for this catheter relative to patients with prosthetic valves.
2015
- 10 Sep 2015 Z-0002-2016 Class II Terminated Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
- 10 Sep 2015 Z-0003-2016 Class II Terminated Image disappeared from the cardiac ultrasound system when the CARTO 3 EP Navigation System needed restarting while the patient was experiencing pericardial effusion. Affects the CARTOSOUND Module of the CARTO 3 EP Navigation System when used with the SOUNDSTAR eco 8F and 10F Diagnostic Ultrasound Catheters. New precautions added.
- 17 Apr 2015 Z-1525-2015 Class II Terminated The Webster HIS catheter, 4Pole Fixed Curve with Auto ID, is not being recognized by the CARTO 3 navigation System.
FDA establishments
- Biosense Webster, Inc. JUAREZ Chihuahua · MX FEI 3003248396
Information
- Country
- United States
- Address
- 31, Technology Drive, Suite 200, Irvine, California, United States
- Website
- its.jnj.com
- [email protected]
- Phone
- +1(800)729-9010
- PRRC contact
- Tucky Wong
- EUDAMED SRN
- US-MF-000014219
- FDA FEI number
- 3022451073
- DUNS number
- 020163218