Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

ULRICH MEDICAL USA, INC.

US

Last updated September 17, 2026

Information

Country
US
Address
3700 East Plano Parkway, Suite 200, Plano, TX, 75074
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
3005532635
DUNS Number
—

Catalogue (1623)

Page 27 of 33
DeviceModel / ReferenceRegistriesClassStatus
Momentum MIS UU092-30-0065
FDA UDI
Class IIActive
Solidity Vertebral Body Replacement UU042-03-04
FDA UDI
Class IIActive
uNion Cervical Plate System UM141-40-20
FDA UDI
Class IIActive
Momentum UU051-02-7550
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-52-1412-10
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU071-11-4022
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-82-1412-06
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU071-00-4040
FDA UDI
Class IIActive
Momentum UU052-60-0010
FDA UDI
Class IIActive
uNion Cervical Plate System UM152-45-14
FDA UDI
Class IIActive
uNion Cervical Plate System UM101-01-12
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU073-20-06
FDA UDI
Class IIActive
Momentum MIS UU091-55-5120
FDA UDI
Class IIActive
Momentum UU051-72-4055
FDA UDI
Class IIActive
uNion MAX UU201-22-4012
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU081-12-1916-10
FDA UDI
Class IIActive
Momentum UU051-01-7535
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU071-00-4032
FDA UDI
Class IIActive
Momentum UU051-20-9090
FDA UDI
Class IIActive
Momentum MIS UU091-20-1110
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-82-1412-07
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU081-00-1916-06
FDA UDI
Class IIActive
Momentum MIS UU091-12
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-50-1211-05
FDA UDI
Class IIActive
uNion Cervical Plate System UU994-47-02
FDA UDI
Class IActive
Momentum UU051-55-0085-L
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU081-06-1211-12
FDA UDI
Class IIActive
uNion Cervical Plate System UM102-02-30
FDA UDI
Class IIActive
Momentum UU051-11-5545
FDA UDI
Class IIActive
Momentum MIS UU091-11-8560
FDA UDI
Class IIActive
Momentum UU053-06-01
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU071-11-4028
FDA UDI
Class IIActive
Momentum UU051-01-6560
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU071-31-4010
FDA UDI
Class IIActive
Momentum MIS UU091-20-8070
FDA UDI
Class IIActive
Cortium Universal OCT Spinal Fixation System UU073-10-01
FDA UDI
Class IIActive
uNion Cervical Plate System UM190-00-90
FDA UDI
Class IIActive
Momentum MIS UU091-55-5250-L
FDA UDI
Class IIActive
Momentum UU051-02-8555
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-56-1211-09
FDA UDI
Class IIActive
Momentum UU051-02-5535
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-52-1211-10
FDA UDI
Class IIActive
uNion MAX UU201-01-0017
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU082-50-1614-12
FDA UDI
Class IIActive
Momentum UU051-11-8550
FDA UDI
Class IIActive
Momentum MIS UU091-11-7550
FDA UDI
Class IIActive
Momentum UU052-70-0020
FDA UDI
Class IIActive
Momentum MIS UU092-31-0100
FDA UDI
Class IIActive
Momentum MIS UU091-55-5140
FDA UDI
Class IIActive
Flux-C Cervical Interbody UU081-00-1211-06
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (3)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Dec 31, 2019Z-1350-2020—terminated

Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.

Mar 8, 2019Z-1126-2019—terminated

Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.

Mar 7, 2017Z-1878-2017—terminated

Surgical technique was revised to prevent static plate screws blocking mechanisms from disassociating.