  • 2025

    • 6 Nov 2025 Z-0940-2026 Class II Ongoing Following a component change in the DPOAE probe (speaker replacement), technical distortions may occur at high stimulus levels combined with unusually low OAE amplitude acceptance thresholds. These distortions can mimic true DPOAE responses and may result in a false PASS outcome. Distribution: Domestic: WI. International: Ireland. Submission number: K100661
  • 2023

    • 26 Sep 2023 Z-0401-2024 Class II Ongoing Affected lots of product may experience technical distortions to a greater extent than expected. Distribution: US States: FL, MO, OH, WI Submission numbers: K133012, K213345
    • 26 Sep 2023 Z-0402-2024 Class II Ongoing Affected lots of product may experience technical distortions to a greater extent than expected. Distribution: US States: FL, MO, OH, WI Submission numbers: K133012, K213345
    • 26 Sep 2023 Z-0403-2024 Class II Ongoing Affected lots of product may experience technical distortions to a greater extent than expected. Distribution: US States: FL, MO, OH, WI Submission numbers: K133012, K213345
    • 26 Sep 2023 Z-0404-2024 Class II Ongoing Affected lots of product may experience technical distortions to a greater extent than expected. Distribution: US States: FL, MO, OH, WI Submission numbers: K133012, K213345

This is an FDA establishment record without devices in deviCERTS.