What Kelyniam Global Inc. makes
Kelyniam Global Inc. manufactures non-alterable cranioplasty plates, designed for cranial reconstruction following trauma, surgery, or defect repair. These devices are intended to restore cranial integrity and protect underlying brain tissue.
The company’s cranioplasty plates are classified as Class II medical devices under FDA regulations and are listed in the FDA’s Unique Device Identifier (UDI) database. The manufacturer is based in the United States.
Catalogue
Search and filter all 1,838 devicesShowing 1,751-1,800 of 1,838
Company details
- FDA FEI number
- 3009424796
- Address
- 97 RIVER RD STE A, COLLINSVILLE, CT, 06019
FDA recalls
FDA recall records list 4 recalls initiated by KELYNIAM GLOBAL INC., initiated on 2017-06-10. By recorded status: 4 terminated. The reasons filed fall into these groups: other reasons as filed. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.
Recalls initiated per year
- 2017 4
- 2018 0
- 2019 0
- 2020 0
- 2021 0
- 2022 0
- 2023 0
- 2024 0
- 2025 0
- 2026 0
2017
- 10 Jun 2017 Z-1870-2018 Class II Terminated Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
- 10 Jun 2017 Z-1871-2018 Class II Terminated Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
- 10 Jun 2017 Z-1872-2018 Class II Terminated Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended
- 10 Jun 2017 Z-1873-2018 Class II Terminated Significant design changes made to the Custom Skull Implant (CSI) devices, such as removing pressure relief holes across the length of the device and adding suture holes along the ridge of the implant, were not validated or submitted to FDA for marketing clearance. The Temporal Suture System (TSS) is no longer being offered and the Integrated Fixation System (IFS) has been temporarily suspended