Fisher & Paykel Healthcare Limited
New ZealandFDA applicant, unverified
- Devices
- 1
- FDA recalls
- 63 ongoing
At a glance
Market reach
- United States
- FDA recalls
Catalogue
Optiflow Oxygen Kit (AA484J) Ref K262268
Regimes: 510(k) Registry: 510(k)
FDA recalls
Recalls initiated per year
- 2017 0
- 2018 0
- 2019 0
- 2020 0
- 2021 0
- 2022 0
- 2023 0
- 2024 2
- 2025 1
- 2026 0
2025
- 18 Sep 2025 Z-0158-2026 Class II Ongoing When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
Full distribution text
- 18 Sep 2025 Z-0158-2026 Class II Ongoing When unintentionally disconnected from power source, humidifier device (used to deliver high flow warmed and humidified respiratory gases) may sound audible alert for less than 120 seconds, which may delay user awareness to reconnect power, which could lead to oxygen desaturation, hypoxia. Disinfection kit user manual updated to include checking the audible alert prior to use on a new patient.
2024
- 7 Aug 2024 Z-3262-2024 Class II Ongoing Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
Full distribution text
- 2 Apr 2024 Z-1823-2024 Class II Ongoing Humidifier devices, used to deliver high flow respiratory therapy to patients, manufactured before 14 August 2017, have a speaker configuration that may result in distorted, intermittent or inaudible alarm sound levels. In the absence of an audible alarm, if there is an interruption to therapy, a patient may experience oxygen desaturation.
Full distribution text
- 7 Aug 2024 Z-3262-2024 Class II Ongoing Due to a software issue, affected devices that are set up with High Pressure Oxygen (HPO), if the flow alignment alarm occurs, the device will deliver room air only. If this happens, a patient my experience oxygen desaturation that could lead to hypoxia.
2014
- 6 Nov 2014 Z-0210-2015 Class II Terminated Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.
Full distribution text
- 15 Apr 2014 Z-1539-2014 Class I Terminated The Infant Nasal CPAP Prong has the potential to detach from the Nasal Tubing and therapy is likely to be interrupted.
Full distribution text
- 6 Nov 2014 Z-0210-2015 Class II Terminated Breakage of the nut that secures the heater head in place, which can cause the heater head to become partially detached and swing towards the bassinet. There is the potential that a heater head could contact a patient situated on the bassinet and cause a serious injury.
2006
- 9 Oct 2006 Z-0485-2007 Unclassified (pre-2012) Terminated This firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height. A bent pivot plate generally results in the Warmer being unable to elevate properly. Other component failures may result.
Frequently asked questions
How many devices does Fisher & Paykel Healthcare Limited have in the registry?
The registry currently lists 1 device for Fisher & Paykel Healthcare Limited. Devices recorded under merged duplicate company entries are counted once.
Which registries list Fisher & Paykel Healthcare Limited's devices?
Among its 1 most recently registered device, Fisher & Paykel Healthcare Limited has devices listed in FDA 510(k) (1 device).
What risk classes do Fisher & Paykel Healthcare Limited's devices fall under?
Among its 1 most recently registered device, the classes recorded for Fisher & Paykel Healthcare Limited's devices are Class II. Each device page shows its own class and authorisation.
Are there FDA recall records for Fisher & Paykel Healthcare Limited?
FDA recall records list 6 recalls initiated by Fisher & Paykel Healthcare Limited, the most recent initiated on 18 Sep 2025. 3 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Fisher & Paykel Healthcare Limited in the registry?
The most recently registered devices listed for Fisher & Paykel Healthcare Limited include Optiflow Oxygen Kit (AA484J).