Gendex Dental Systems
Country not established
- Devices
- 2
- FDA recalls
- 7
At a glance
Market reach
- United States
- FDA recalls
Catalogue
Vix Win Pro Ref K060178
Regimes: 510(k) Registry: 510(k)- Regimes: 510(k) Registry: 510(k)
FDA recalls
2011
- 29 Dec 2011 Z-1358-2012 Unclassified (pre-2012) Terminated Improper hardware was used during installation of the eXpert DC Intraoral X-Ray system units causing drifting of the tubehead assembly.
- 4 Feb 2011 Z-2040-2011 Unclassified (pre-2012) Terminated There is an anomaly in the Orthoralix firmware which can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped, negative, reversed or had lines across them) is most likely to occur during periods of high network traffic and is significant enough that it is clear and obviou
Full distribution text
- 4 Feb 2011 Z-2041-2011 Unclassified (pre-2012) Terminated There is an anomaly in the Orthoralix firmware which can cause a message corruption which will distort the patient image being transferred from the X-Ray system to the PC. This image distortion/corruption (distorted, fragmented, flipped, negative, reversed or had lines across them) is most likely to occur during periods of high network traffic and is significant enough that it is clear and obviou
Full distribution text
2010
- 21 Dec 2010 Z-1373-2011 Unclassified (pre-2012) Terminated The Automatic Exposure Control (AEC) feature located on the control panel of the Orthoralix 8500 Panoramic Dental X-Ray System was functional, but was subsequently disabled in units manufactured starting 5/17/10. North American units manufactured between 12/4/09 and 5/16/10 had functioning AEC buttons, but their manuals stated they were disabled. Units intended for International distribution bet
Full distribution text
- 20 Dec 2010 Z-1559-2011 Unclassified (pre-2012) Terminated The X-Ray unit may break loose from the mounting plate and fall off the wall due to defects in the mounting plate.
- 17 Dec 2010 Z-1248-2011 Unclassified (pre-2012) Terminated The intraoral x-ray systems were installed improperly with incorrect screws which allow the unit to pull away from the wall when the arm is fully extended. This condition may harm the patient.
- 21 Dec 2010 Z-1373-2011 Unclassified (pre-2012) Terminated The Automatic Exposure Control (AEC) feature located on the control panel of the Orthoralix 8500 Panoramic Dental X-Ray System was functional, but was subsequently disabled in units manufactured starting 5/17/10. North American units manufactured between 12/4/09 and 5/16/10 had functioning AEC buttons, but their manuals stated they were disabled. Units intended for International distribution bet
2009
- 6 Mar 2009 Z-1548-2009 Unclassified (pre-2012) Terminated There is a compatibility issue when the VixWin Platinum versions 1.0/1.1/1.2 software is used in conjunction with Easy Dental, PracticeWorks, EagleSoft and SoftDent practice management software, resulting in patient files that may contain images that are mixed between patients.
Frequently asked questions
How many devices does Gendex Dental Systems have in the registry?
The registry currently lists 2 devices for Gendex Dental Systems. Devices recorded under merged duplicate company entries are counted once.
Which registries list Gendex Dental Systems's devices?
Among its 2 most recently registered devices, Gendex Dental Systems has devices listed in FDA 510(k) (2 devices).
What risk classes do Gendex Dental Systems's devices fall under?
Among its 2 most recently registered devices, the classes recorded for Gendex Dental Systems's devices are Class II. Each device page shows its own class and authorisation.
Are there FDA recall records for Gendex Dental Systems?
FDA recall records list 7 recalls initiated by Gendex Dental Systems, the most recent initiated on 29 Dec 2011. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Gendex Dental Systems in the registry?
The most recently registered devices listed for Gendex Dental Systems include VIX WIN PRO and ARTHORALIX 8500, MODELS 110-0199G1, 110-0199G2, 110-0199G3, 110-0199G4.