What I-Flow Corp. makes

I-Flow Corp. manufactures a range of infusion and catheter products, including the Soaker Catheter, Painbuster Infusion Kit, Titan Driver 60cc administration set, Paragon Infusion System, Sidekick Infusion Kit, elastomeric pumps such as the I-Flow Elastomeric Pump with Bolus and various models (On-Q, Painbuster, C-Block, HomePump, EasyPump), HomePump C-Series and One-Step KVO devices, fixed flow rate gravity administration sets, a multi‑drug infusion system, and personal infusers VIVUS 50 and VIVUS 100.

All of these devices are classified as Class II medical devices and are listed in the FDA 510(k) database. I-Flow Corp. is based in the United States.

AI-generated from registry data, 26 Oct 2026. Not verified by deviCERTS or the manufacturer.

Catalogue42

Newest first, rows 1-42 of 42

Search and filter all 42 devices
Skip device list

Rows 1-42 of 42

FDA recalls18

Recalls are matched to the firm by name. FDA recall data as of 22 Oct 2026.

FDA recall records list 18 recalls initiated by I-Flow Corp., the earliest on 2002-08-19 and the most recent on 2012-06-01. By recorded status: 18 terminated. The reasons filed fall into these groups: labeling or instructions for use, sterility or contamination, packaging and manufacturing or quality process. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.

Generated automatically from FDA recall records, as a summary of the records listed below. It describes what the records state and is not an assessment of this manufacturer.

Showing the 10 most recent of 18 recalls.

  • 2012

    • 1 Jun 2012 Z-1444-2013 Class II Terminated I-Flow initiated two voluntary product removals of the ON-Q Pain Relief System: T-Block Needle/Catheter Tray and T-Block Needle/Catheter Kit because they may have potentially malfunctioned. Distribution: Nationwide distribution: USA including states of: CA, CT, FL, IL, IN, KY, NY, OH, OR, TN, TX, VA and WI. Submission number: K904440
    • 1 Jun 2012 Z-1445-2013 Class II Terminated I-Flow initiated two voluntary product removals of the ON-Q Pain Relief System: T-Block Needle/Catheter Tray and T-Block Needle/Catheter Kit because they may have potentially malfunctioned. Distribution: Nationwide distribution: USA including states of: CA, CT, FL, IL, IN, KY, NY, OH, OR, TN, TX, VA and WI. Submission number: K904440
    • 8 May 2012 Z-2253-2012 Class I Terminated It was determined that in a small quantity of ON-Q pumps with ONDEMAND, the bolus button may not lock in the down position when depressed and/or the orange bolus refill indicator may stay in the lowermost position. As a result, the patient may receive continuous infusion at a rate greater than expected. Distribution: Worldwide Distribution -- USA, Canada, Australia, the EU (Germany), United Arab Emirates, and Cyprus. Submission number: K023318
  • 2011

    • 15 Apr 2011 Z-2860-2011 Unclassified (pre-2012) Terminated The recall was initiated because I Flow became aware that a small percentage of ON-Q Pump with Select A Flow may develop a minor leak with a flow rate controller. There may be consequences for the patient, such as under delivery of pain medication or the need to prematurely remove the pump, which would result in loss of the pain treatment provided by the pump. Distribution: Nationwide (USA) Distribution and worldwide to: Australia, Canada, Cyprus, Italy, Germany, Finland, Greece, Israel,…
      Full distribution text Nationwide (USA) Distribution and worldwide to: Australia, Canada, Cyprus, Italy, Germany, Finland, Greece, Israel, South Africa, Sweden, UAE, and UK.
      Submission number: K063530
  • 2008

    • 1 Aug 2008 Z-2249-2009 Unclassified (pre-2012) Terminated I-Flow identified a B. Braun syringe package that had a tear/hole in the paper portion of the primary package. Distribution: France Submission number: K071459
    • 1 Aug 2008 Z-2250-2009 Unclassified (pre-2012) Terminated I-Flow identified a B. Braun syringe package that had a tear/hole in the paper portion of the primary package. Used to provide an optional filling device for the Easypump. Distribution: France Submission number: K071459
    • 14 Feb 2008 Z-2133-2008 Unclassified (pre-2012) Terminated Two lots of ON-Q C-bloc, Model Number CB005 may contain an incorrect fill port label. Distribution: Class II Recall - Nationwide Distribution --- including states of IL, TX, PA, OK, NC, GA, MI, FL, OH, and CA. Submission number: K063530
    • 12 Feb 2008 Z-1869-2008 Unclassified (pre-2012) Terminated The ON-Q PainBuster, REF PS6504 (lot number 772166), box may contain the wrong product, ON-Q PainBuster, REF PS6502. If the wrong product is used, patients would receive a 2 ml/hr infusion rather than a 5 ml/hr infusion, which could result in inadequate pain relief. Distribution: Internationally - United Kingdom, Germany, Spain, Denmark, Australia, and Italy Submission number: K063530
    • 11 Feb 2008 Z-2111-2009 Unclassified (pre-2012) Terminated I-Flow Corporation received two complaints involving Easypump LT 60-24 (Lot 612996) containing a 100 ml x 2 ml/hr pump in the sterile package instead of a 60 ml x 2 ml/hr pump (Illustration was included to show consignee what both the correct and incorrect part looks like). Lot 612996 was manufactured on January 6, 2006. As of February 7, 2008, there have been no further complaints received on t Distribution: Nationwide Distribution to B. Braun, Spain, who then further distributed to Italy, Germany, Netherland, Latvia, Spain,…
      Full distribution text Nationwide Distribution to B. Braun, Spain, who then further distributed to Italy, Germany, Netherland, Latvia, Spain, Portugal and Singapore.
      Submission number: K052117
  • 2007

    • 27 Mar 2007 Z-0057-2008 Unclassified (pre-2012) Terminated Misassembly: The ON-Q PainBuster PMB01 package may contain a Soaker catheter instead of a Non-Soaker catheter. As a result, the bolus button may remain latched in the down position which may cause the the device to deliver continuously at 7 ml/hr instead of 2 ml/hr. Distribution: Worldwide, USA and Australia Submission number: K932740
Show all 18 recalls
  • 2006

    • 29 Nov 2006 Z-0358-2007 Unclassified (pre-2012) Terminated The pump flow rate labeling may not match the package labeling. The label on filter may not match the label on the top of the pump or the package labeling. Distribution: Nationwide Submission number: K932740
    • 17 Jan 2006 Z-0584-06 Unclassified (pre-2012) Terminated The sterile pack may contain the wrong infusion pump inside. Distribution: France only Submission numbers: K052117, K932740
    • 17 Jan 2006 Z-0585-06 Unclassified (pre-2012) Terminated The sterile pack may contain the wrong infusion pump inside. Distribution: France only Submission numbers: K052117, K932740
  • 2003

    • 23 Aug 2003 Z-0313-04 Unclassified (pre-2012) Terminated Some pumps run at 2 ml/hr instead of 0.5 ml/hr. Distribution: Only to France, and one customer.
    • 18 Mar 2003 Z-0741-03 Unclassified (pre-2012) Terminated Overinfusion due to misassembly. Distribution: Only in France to distributor B. Braun Medical, France Submission number: K991513
    • 5 Mar 2003 Z-0709-03 Unclassified (pre-2012) Terminated Wrong component may result in over infusion of medications. Distribution: States of NY and MA, and internationally to England, Spain, Mexico, Canada, Italy, Austrailia, South Africa. Submission number: K923875
    • 4 Mar 2003 Z-0687-03 Unclassified (pre-2012) Terminated Product infuses too fast at rate of 30 minutes instead of 2 hours. Component mix-up. Distribution: CA, OH, WA, TX, UT, SC, MI, IL. Submission number: K932740
  • 2002

    • 19 Aug 2002 Z-0312-04 Unclassified (pre-2012) Terminated Slower than expected infusion of medications, such as antibiotics. Submission number: K020862

Frequently asked questions

How many devices does I-Flow Corp. have in the registry?

The registry currently lists 42 devices for I-Flow Corp.. Devices recorded under merged duplicate company entries are counted once.

Which registries list I-Flow Corp.'s devices?

Among its 42 most recently registered devices, I-Flow Corp. has devices listed in FDA 510(k) (42 devices).

What risk classes do I-Flow Corp.'s devices fall under?

Among its 42 most recently registered devices, the classes recorded for I-Flow Corp.'s devices are Class II. Each device page shows its own class and authorisation.

Are there FDA recall records for I-Flow Corp.?

FDA recall records list 18 recalls initiated by I-Flow Corp., the most recent initiated on 2012-06-01. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.

What are the newest devices from I-Flow Corp. in the registry?

The most recently registered devices listed for I-Flow Corp. include SOAKER CATHETER, PAINBUSTER INFUSION KIT, TITAN DRIVER 60CC, 30 MINUTE ADMINISTRATION SET and PARAGON INFUSION SYSTEM.

Generated automatically from registry data and not verified by a human reviewer.