## Figures

- Devices: 5
- FDA recalls (matched to the firm by name): 3, 2 ongoing (2 open since before 2021)
- FDA establishments: 1

## Registry facts

*Every device of the manufacturer. A device listed in two registries counts in both.*

- Registries: EUDAMED 4, FDA 510(k) 1
- Device classes: EU Class I 2, EU Class IIa 1, EU Class IIb 1, FDA Class II 1
- EMDN [Z12 Instruments for functional explorations and therapeutic interventions](https://devicerts.com/categories/emdn/Z12): 4 devices

## Devices

The first 5 of 5, in the order of the canonical page, which lists them all.

- [Needle Holder](https://devicerts.com/devices/needle-holder-a10010) · ref A10010 · Class I · EUDAMED · status: Active
- [IMACTIS CT-Navigation system](https://devicerts.com/devices/imactis-ct-navigation-system-jak) · ref K162314 · Class II · 510(k) · status: Active · decision 2018-04-24 · open recall
- [IMACTIS Navigation station](https://devicerts.com/devices/imactis-navigation-station-j02000) · ref J02000 · Class IIb · EUDAMED · status: Active
- [Needle Holder (Not sterile)](https://devicerts.com/devices/needle-holder-not-sterile-a10200) · ref A10200 · Class I · EUDAMED · status: Active
- [IMACTIS® NaviKit™](https://devicerts.com/devices/imactis-r-navikit-i10100) · ref I10100 · Class IIa · EUDAMED · status: Active

## FDA recalls (matched by name, newest first)

- Z-1264-2021 · Class II · Terminated · initiated 2021-02-11 · product code JAK: Defect in the mechanical assembly of the needle holder contained in the navigation kit. One of the 2 jaws of the needle holder does not provide gripping functionality; if the defect is not detected before use, it can cause poor needle guidance and a lack of precision during navigation causing an inaccurate needle position which could cause damage to organs or vasculature adjacent to the trajectory
- Z-2972-2020 · Class II · Open since 2020 · initiated 2020-08-24 · product code JAK: The firm has received reports concerning the patient fiducial disk separating from the body and users attempting to re-assemble the device. The use of a re-assembled patient fiducial could lead to inability to navigate or inaccuracies such as improper needle trajectory.
- Z-0838-2020 · Class II · Open since 2019 · initiated 2019-08-01 · product code JAK: The firm has become aware that the sensor cover, a component of the single-use navigation kit may contain microbial contamination. A source of microbial contamination has been identified in a raw product used to manufacture the tip of the cover.This could result in direct exposure of the intra-cardiovascular, intra-lymphatic or neurological system to a level of endotoxin sufficient to produce a pyrogenic response.

## Source and citation

- Cite as: Imactis SAS. deviCERTS medical device catalogue. https://devicerts.com/manufacturers/imactis-sas (data as of 2026-11-26).
- Data sources: [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA registration and listing](https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfrl/rl.cfm), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- EUDAMED record synced: 2026-11-07
- FDA recall data as of: 2026-11-01
- Data as of: 2026-11-26
- JSON: <https://devicerts.com/manufacturers/imactis-sas.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.