# Inpeco SA

> Medical device manufacturer record on deviCERTS, compiled from public registries. Inpeco SA is a medical device manufacturer based in Switzerland, with 8 devices listed in EUDAMED and FDA GUDID (EU Class A; FDA Class I) and 33 FDA recalls as of 18 Nov 2026.

- Canonical page: <https://devicerts.com/manufacturers/inpeco-sa>
- Country: Switzerland (from SRN)
- EUDAMED actor status: active
- Role: Manufacturer
- Registrations: EUDAMED, FDA
- EUDAMED SRN: CH-MF-000017833
- FDA FEI: 3010825766
- DUNS: 484880690

## Figures

- Devices: 8
- FDA recalls (matched to the firm by name): 33, 17 ongoing (6 open since before 2021)

## Registry facts

*Every device of the manufacturer. A device listed in two registries counts in both.*

- Registries: EUDAMED 7, FDA GUDID 7
- Device classes: EU Class A 7, FDA Class I 1
- EMDN [W02 IVD instruments](https://devicerts.com/categories/emdn/W02): 7 devices

## Sites and related companies

- Authorised representative: [QbD RepS BV](https://devicerts.com/representatives/qbd-reps-bv-be-ar-000000040) (Belgium)

## Devices

The first 8 of 8, in the order of the canonical page, which lists them all.

- [ProTube PTB](https://devicerts.com/devices/protube-ptb-e40cc2db) · Class I · GUDID · status: Active
- [FlexLab](https://devicerts.com/devices/flexlab-flx) · ref FLX · Class A · EUDAMED, GUDID · status: Active
- [FlexLab X](https://devicerts.com/devices/flexlab-x-fxx) · ref FXX · Class A · EUDAMED, GUDID · status: Active
- [Aptio® Automation](https://devicerts.com/devices/aptio-r-automation-ap2) · ref AP2 · Class A · EUDAMED, GUDID · status: Active
- [ProTube](https://devicerts.com/devices/protube-ptb) · ref PTB · Class A · EUDAMED, GUDID · status: Active
- [Accelerator a3600](https://devicerts.com/devices/accelerator-a3600-acp) · ref ACP · Class A · EUDAMED, GUDID · status: Withdrawn
- [FlexPath Blox](https://devicerts.com/devices/flexpath-blox-fpb) · ref FPB · Class A · EUDAMED, GUDID · status: Active
- [FlexPath Blox](https://devicerts.com/devices/flexpath-blox-fpb-6f81f163) · ref FPB · Class A · EUDAMED · status: Withdrawn

## FDA recalls (matched by name, newest first)

- Z-0595-2026 · Class II · Ongoing · initiated 2025-10-30 · product code CEM: The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results. This error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes.
- Z-0596-2026 · Class II · Ongoing · initiated 2025-10-30 · product code CEM: The Aliquoter Module (mALX) may generate secondary sample tubes without labels or with mismatched labels which may result in delayed results or erroneous results, this error occurs if the operators open the printer head of mALX to perform error recoveries or maintenance procedures and do not empty the printer tubes buffer to remove all the secondary sample tubes
- Z-0106-2026 · Class II · Ongoing · initiated 2025-09-03 · product code CEM: The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
- Z-0107-2026 · Class II · Ongoing · initiated 2025-09-03 · product code CEM: The FlexLab X System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.
- Z-0560-2024 · Class II · Ongoing · initiated 2023-11-03 · product code CEM: Firmware of the Interface Module with Roche Cobas 8000 and Cobas PRO (CO8 IM) may lead to a delayed sample tube processing.
- Z-1559-2023 · Class II · Ongoing · initiated 2023-03-10 · product code CEM: Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
- Z-1560-2023 · Class II · Ongoing · initiated 2023-03-10 · product code CEM: Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
- Z-1561-2023 · Class II · Ongoing · initiated 2023-03-10 · product code CEM: Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
- Z-1562-2023 · Class II · Ongoing · initiated 2023-03-10 · product code CEM: Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
- Z-1563-2023 · Class II · Ongoing · initiated 2023-03-10 · product code CEM: Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this specific scenario, the Tube A may be released by the module as Tube B due to a miscommunication between the module firmware and the Automation software without any error message. The Automation System loses the traceability of Tube A. It manages both Tube A (incorrectly identified as Tube B) and the real Tube B according to the test orders not performed yet on Tube B. Inpeco has released a Technical Service Bulletin with the procedure to correctly update the firmware. The firmware upgrade will be performed by Field Service Engineers. A CAPA (Corrective Action Preventive Action) has been opened to deeply investigate the root cause and identify possible process lack in order to avoid reoccurrences.
- Z-0087-2022 · Class II · Ongoing · initiated 2021-09-10 · product code CEM: The Firmware (FW) of the Automation System Interface Module to ADVIA Centaur XP/XPT may lead to an incorrect association of test results to sample ID. Improper management of the sample may bring an incorrect patient result.
- Z-2973-2020 · Class II · Terminated · initiated 2020-08-07 · product code CEM: Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
- Z-2974-2020 · Class II · Terminated · initiated 2020-08-07 · product code CEM: Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
- Z-2975-2020 · Class II · Terminated · initiated 2020-08-07 · product code CEM: Firmware of the Sysmex XN-9000/XN-9100 Interface Module (IM) and of the Track to Rack Module has the potential to mis-associate sample IDs leading to incorrect results or delayed results
- Z-0432-2021 · Class II · Terminated · initiated 2020-07-14 · product code CEM: When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
- Z-0433-2021 · Class II · Terminated · initiated 2020-07-14 · product code CEM: When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
- Z-0434-2021 · Class II · Terminated · initiated 2020-07-14 · product code CEM: When a Clot Detection Error is generated during the sample aspiration from the Primary Sample Tube the aspirated volume is dispensed into the first empty Secondary Sample Tube. This Secondary Sample Tube is flagged with error 2132 or 1442 and sent to IOM Priority Output Racks to be manually managed. The current error message associated to error 2132 or 1442 recommends that the operator manages these Secondary Sample Tubes according to Laboratory Practice, but it does not clarify that these Secondary Sample Tubes may be diluted with water from the hydraulic circuit of the Aliquoter Module this may lead to reporting of erroneous results if the diluted sample is used for testing.
- Z-1887-2020 · Class II · Open since 2020 · initiated 2020-04-07 · product code JJE: The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
- Z-1888-2020 · Class II · Open since 2020 · initiated 2020-04-07 · product code JJE: The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
- Z-1889-2020 · Class II · Open since 2020 · initiated 2020-04-07 · product code JJE: The FlexLab Centrifuge Module loading algorithm may lead to unbalanced loads which could damage the centrifuge.
- Z-2003-2020 · Class II · Terminated · initiated 2020-04-07 · product code CEM: When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
- Z-2004-2020 · Class II · Terminated · initiated 2020-04-07 · product code CEM: When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.
- Z-2114-2020 · Class II · Terminated · initiated 2020-04-07 · product code CEM: The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
- Z-2115-2020 · Class II · Terminated · initiated 2020-04-07 · product code JQP: The sensors which activate the safety switches may be affected by delayed activation. In 1% of the cases, the response delay can be up to a maximum delay of 2 seconds, instead of 40ms expected by design.
- Z-2140-2020 · Class II · Terminated · initiated 2020-04-07 · product code JQP: The identified problem is an erroneous association between the carrier and the sample tube caused by a communication error between the firmware of the Interface Modules and the Automation software. This problem can occur only when one of these Interface Modules is put off-line after a carrier in their secondary lane is physically returned on the main track and then put back on-line when the carrier is used to transport another tube.

## Source and citation

- Cite as: Inpeco SA. deviCERTS medical device catalogue. https://devicerts.com/manufacturers/inpeco-sa (data as of 2026-11-18).
- Data sources: [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA GUDID](https://accessgudid.nlm.nih.gov/), [FDA registration and listing](https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfrl/rl.cfm), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- EUDAMED record synced: 2026-11-07
- FDA recall data as of: 2026-11-01
- Data as of: 2026-11-18
- JSON: <https://devicerts.com/manufacturers/inpeco-sa.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/manufacturers/inpeco-sa
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.
