International Medsurg Connection

Also registered as International Medsurg Connection, Inc.

Country not established

Devices
13
FDA recalls
65

At a glance

Counted by deviCERTS from registry records

Market reach

United States
FDA recalls

What International Medsurg Connection makes

International Medsurg Connection manufactures single-use surgical and medical devices, including irrigation sets, hypodermic and insulin syringes, blood collection needles, surgical drapes, hypodermic needles, surgical face masks, and equipment covers for medical procedures.

The company’s portfolio consists of Class II medical devices, all of which are listed under the FDA’s 510(k) pathway. These devices are regulated under U.S. federal law and are intended for use in surgical and procedural settings.

AI-generated from registry data, 26 Oct 2026. Not verified by deviCERTS or the manufacturer.

Catalogue13

Newest first, rows 1-13 of 13

Search and filter all 13 devices
Skip device list
  • Irrigation Sets Ref K023808

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Surgical Drape Ref K100991

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Hypodermic Needle Ref K102584

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Medsurg Equipment Covers Ref K014239

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Piston Syringes Ref K102969 Open recall

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Insulin Syringe Ref K110882

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Surgical Drapes Ref K052169

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Piston Syringe & Hypodermic Needle Ref K022159

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Surgical Face Mask Ref K100846

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Surgical Drape Ref K100208

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Under Buttocks Drape Ref K042893

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Surgical Drapes Ref K971159

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Imc Blood Collection Needle Ref K110656

    Active Regimes: 510(k) Registry: 510(k) Class:II

Rows 1-13 of 13

FDA recalls65

Recalls are matched to the firm by name. FDA recall data as of 22 Oct 2026.

FDA recall records list 65 recalls initiated by International Medsurg Connection, the earliest on 2006-08-16 and the most recent on 2017-06-23. By recorded status: 65 terminated. The reasons filed fall into these groups: labeling or instructions for use and sterility or contamination. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.

Generated automatically from FDA recall records, as a summary of the records listed below. It describes what the records state and is not an assessment of this manufacturer.

Recalls initiated per year

  • 2017 59
  • 2018 0
  • 2019 0
  • 2020 0
  • 2021 0
  • 2022 0
  • 2023 0
  • 2024 0
  • 2025 0
  • 2026 0

6 initiated before 2017

Showing the 10 most recent of 65 recalls.

  • 2017

    • 23 Jun 2017 Z-2804-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
    • 23 Jun 2017 Z-2805-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2806-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2807-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2808-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2809-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2810-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2811-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2812-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2813-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
Show the 25 most recent of 65 recalls
  • 2017

    • 23 Jun 2017 Z-2814-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2815-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2816-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2817-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2818-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2819-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2820-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2821-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2822-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2823-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2824-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102584
    • 23 Jun 2017 Z-2825-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102969
    • 23 Jun 2017 Z-2826-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102969
    • 23 Jun 2017 Z-2827-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102969
    • 23 Jun 2017 Z-2828-2017 Class II Terminated Insufficient documentation to support product sterility claims. Distribution: Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN,…
      Full distribution text Nationwide Distribution including OH, MA, GA, FL, NC, MS, MO, PA, IN, NJ, TX, KS, LA, NE, WA, AL, UT, AK, CA, IL, MN, MD, NY, and MI.
      Submission number: K102969

Frequently asked questions

How many devices does International Medsurg Connection have in the registry?

The registry currently lists 13 devices for International Medsurg Connection. Devices recorded under merged duplicate company entries are counted once.

Which registries list International Medsurg Connection's devices?

Among its 13 most recently registered devices, International Medsurg Connection has devices listed in FDA 510(k) (13 devices).

What risk classes do International Medsurg Connection's devices fall under?

Among its 13 most recently registered devices, the classes recorded for International Medsurg Connection's devices are Class II. Each device page shows its own class and authorisation.

Are there FDA recall records for International Medsurg Connection?

FDA recall records list 65 recalls initiated by International Medsurg Connection, the most recent initiated on 2017-06-23. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.

What are the newest devices from International Medsurg Connection in the registry?

The most recently registered devices listed for International Medsurg Connection include IRRIGATION SETS, IMC SURGICAL DRAPE, IMC HYPODERMIC NEEDLE and MEDSURG EQUIPMENT COVERS.

Generated automatically from registry data and not verified by a human reviewer.