Not available yet.
FDA Recalls (3)
Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.
| Date | Recall No. | Class | Status | Reason |
|---|---|---|---|---|
| Jun 27, 2024 | Z-2457-2024 | — | open, classified | If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue. |
| Feb 16, 2024 | Z-1381-2024 | — | open, classified | Neonatal pads are experiencing reduced water flow, which may reduce or prevent water heating or cooling performance of the pads that are part of a temperature management system. |
| Jan 5, 2005 | Z-0436-05 | — | terminated | Potential for an inaccuracy of the primary patient temperature reading that is used to monitor and control patient temperature. |