  "format": "devicerts.manufacturer.v1",
  "cite_as": "Mobius Imaging. deviCERTS medical device catalogue. https://devicerts.com/manufacturers/mobius-imaging (data as of 2026-11-23).",
  "source": {
    "canonical_url": "https://devicerts.com/manufacturers/mobius-imaging",
    "json_url": "https://devicerts.com/manufacturers/mobius-imaging.json",
    "markdown_url": "https://devicerts.com/manufacturers/mobius-imaging.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA GUDID",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA registration and listing",
        "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfrl/rl.cfm"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "eudamed_record_synced": null,
    "fda_recalls_as_of": "2026-11-01",
    "registry_facts_computed": "2026-11-22",
    "status_as_of": "2026-11-23",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/manufacturers/mobius-imaging",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "manufacturer": {
    "name": "Mobius Imaging",
    "url": "https://devicerts.com/manufacturers/mobius-imaging",
    "registered_as": null,
    "abbreviated_name": null,
    "also_registered_as": [
      "Mobius Imaging, LLC"
    ],
    "country": "United States",
    "country_basis": "from the FDA registration address",
    "eudamed_status": null,
    "verified": false,
    "roles": [
      "Manufacturer"
    ],
    "registrations": [
      "FDA"
    ],
    "identifiers": [
      {
        "label": "FDA FEI",
        "value": "3010151377"
      }
    ],
    "summary": "Mobius Imaging is a medical device manufacturer, with 4 devices listed in FDA 510(k) and FDA GUDID (FDA Class II) and 7 FDA recalls as of 23 Nov 2026.",
    "noindex": false
  },
  "counts": {
    "devices": 4,
    "devices_note": null,
    "certificates": 0,
    "fda_recalls_matched_by_name": 7,
    "fda_recalls_ongoing": 3,
    "fda_recalls_ongoing_recent": 3,
    "fda_recalls_ongoing_before": null,
    "fda_establishments": 0,
    "sites": 0
  },
  "registry_facts": {
    "covers": "Every device of the manufacturer. A device listed in two registries counts in both.",
    "registries": [
      {
        "registry": "FDA 510(k)",
        "devices": 3
      },
      {
        "registry": "FDA GUDID",
        "devices": 1
      }
    ],
    "device_classes": [
      {
        "class": "FDA Class II",
        "devices": 4
      }
    ],
    "top_emdn_categories": []
  },
  "related": {
    "site_of": null,
    "part_of": null,
    "sites": [],
    "authorised_representatives": []
  },
  "devices": {
    "shown": 4,
    "total": 4,
    "all_devices_url": "https://devicerts.com/manufacturers/mobius-imaging",
    "items": [
      {
        "name": "Airo® MobiCT-32",
        "url": "https://devicerts.com/devices/airo-r-mobict-32",
        "model_reference": null,
        "device_class": "Class II",
        "registries": [
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Airo Ct System",
        "url": "https://devicerts.com/devices/airo-ct-system-jak",
        "model_reference": "K131431",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2013-09-26",
        "open_recall": false
      },
      {
        "name": "AIRO Computed Tomography (CT) X-ray System",
        "url": "https://devicerts.com/devices/airo-computed-tomography-ct-x-ray-system-jak-k180393",
        "model_reference": "K180393",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2018-03-15",
        "open_recall": true
      },
      {
        "name": "AIRO Computed Tomography (CT) X-ray System",
        "url": "https://devicerts.com/devices/airo-computed-tomography-ct-x-ray-system-jak",
        "model_reference": "K160126",
        "device_class": "Class II",
        "registries": [
          "510(k)"
        ],
        "status": "Active",
        "decision_date": "2016-04-19",
        "open_recall": false
      }
    ]
  },
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1040-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2025-12-04",
      "product_code": "JAK",
      "reason": "AIRO pendant replacement kits and upgrade do not possess the \"Safe Hand Position Warning\" label."
    },
    {
      "number": "Z-1041-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2025-12-04",
      "product_code": "JAK",
      "reason": "AIRO pendant replacement kits and upgrade do not possess the \"Safe Hand Position Warning\" label."
    },
    {
      "number": "Z-0957-2023",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2022-12-07",
      "product_code": "JAK",
      "reason": "Stryker received one report of an injury to an untrained user that occurred while the untrained user was moving an Airo TruCT unit in reverse. The firm will increase visibility of instructions for users to consider when transporting the unit in reverse by providing labels to be affixed onto the AIRO."
    },
    {
      "number": "Z-0344-2023",
      "classification": "Class II",
      "status": "Terminated",
      "initiated": "2022-11-18",
      "product_code": "JAK",
      "reason": "Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing."
    },
    {
      "number": "Z-2991-2018",
      "classification": "Class II",
      "status": "Terminated",
      "initiated": "2018-08-01",
      "product_code": "JAK",
      "reason": "The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed."
    },
    {
      "number": "Z-0715-2018",
      "classification": "Class II",
      "status": "Terminated",
      "initiated": "2017-09-26",
      "product_code": "JAK",
      "reason": "If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan."
    },
    {
      "number": "Z-1016-2015",
      "classification": "Class II",
      "status": "Terminated",
      "initiated": "2014-12-08",
      "product_code": "JAK",
      "reason": "There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to the curve."
    }
  ]
}