# Mobius Imaging

> Medical device manufacturer record on deviCERTS, compiled from public registries. Mobius Imaging is a medical device manufacturer, with 4 devices listed in FDA 510(k) and FDA GUDID (FDA Class II) and 7 FDA recalls as of 23 Nov 2026.

- Canonical page: <https://devicerts.com/manufacturers/mobius-imaging>
- Also registered as: Mobius Imaging, LLC
- Country: United States (from the FDA registration address)
- Role: Manufacturer
- Registrations: FDA
- FDA FEI: 3010151377

## Figures

- Devices: 4
- FDA recalls (matched to the firm by name): 7, 3 ongoing

## Registry facts

*Every device of the manufacturer. A device listed in two registries counts in both.*

- Registries: FDA 510(k) 3, FDA GUDID 1
- Device classes: FDA Class II 4

## Devices

The first 4 of 4, in the order of the canonical page, which lists them all.

- [Airo® MobiCT-32](https://devicerts.com/devices/airo-r-mobict-32) · Class II · GUDID · status: Active
- [Airo Ct System](https://devicerts.com/devices/airo-ct-system-jak) · ref K131431 · Class II · 510(k) · status: Active · decision 2013-09-26
- [AIRO Computed Tomography (CT) X-ray System](https://devicerts.com/devices/airo-computed-tomography-ct-x-ray-system-jak-k180393) · ref K180393 · Class II · 510(k) · status: Active · decision 2018-03-15 · open recall
- [AIRO Computed Tomography (CT) X-ray System](https://devicerts.com/devices/airo-computed-tomography-ct-x-ray-system-jak) · ref K160126 · Class II · 510(k) · status: Active · decision 2016-04-19

## FDA recalls (matched by name, newest first)

- Z-1040-2026 · Class II · Ongoing · initiated 2025-12-04 · product code JAK: AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
- Z-1041-2026 · Class II · Ongoing · initiated 2025-12-04 · product code JAK: AIRO pendant replacement kits and upgrade do not possess the "Safe Hand Position Warning" label.
- Z-0957-2023 · Class II · Ongoing · initiated 2022-12-07 · product code JAK: Stryker received one report of an injury to an untrained user that occurred while the untrained user was moving an Airo TruCT unit in reverse. The firm will increase visibility of instructions for users to consider when transporting the unit in reverse by providing labels to be affixed onto the AIRO.
- Z-0344-2023 · Class II · Terminated · initiated 2022-11-18 · product code JAK: Stryker received a report of a battery fire involving an Airo TruCT unit previously subjected to bypass charging. The firm is notifying customers with units whose battery monitoring systems were bypassed to charge the batteries; the batteries in these units require replacing.
- Z-2991-2018 · Class II · Terminated · initiated 2018-08-01 · product code JAK: The Tube Current Modulation feature (Modulated Scans) is not working in AIRO systems with software version 2.0.0.0, and operators would not be able to detect this fault until after a scan is completed.
- Z-0715-2018 · Class II · Terminated · initiated 2017-09-26 · product code JAK: If the tilt-drive motor has an intermittent connection or broken encoder or sensor wire there is a risk of unintended or unpredictable device motion while the CT system is homing the axis at startup or preparing for a scan.
- Z-1016-2015 · Class II · Terminated · initiated 2014-12-08 · product code JAK: There is a risk that during the transfer of an image and navigation data to the Brainlab Curve Image Guided Surgery Navigation System after a CT scan, an error may occur, causing either no navigation data or incorrect navigation data being transferred to the curve.

## Source and citation

- Cite as: Mobius Imaging. deviCERTS medical device catalogue. https://devicerts.com/manufacturers/mobius-imaging (data as of 2026-11-23).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA GUDID](https://accessgudid.nlm.nih.gov/), [FDA registration and listing](https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfrl/rl.cfm), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- FDA recall data as of: 2026-11-01
- Data as of: 2026-11-23
- JSON: <https://devicerts.com/manufacturers/mobius-imaging.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/manufacturers/mobius-imaging
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.