What NAVILYST MEDICAL, INC. makes

Navilyst Medical produces single-use and reusable medical devices for vascular access and monitoring, including angiography kits, peripherally-inserted central venous catheters, stopcocks, invasive-pressure transducers, peripheral IV cannulas, guidewires, angiographic syringes, and blood pressure monitoring tubing sets.

The company’s portfolio includes Class I and Class II devices, all listed in the FDA’s Unique Device Identifier (UDI) database. The manufacturer is based in the United States.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Frequently asked questions

What types of medical devices does Navilyst Medical, Inc. manufacture?

Navilyst Medical, Inc. specializes in vascular access and monitoring devices. Their portfolio includes single-use and reusable products like angiography kits, peripherally-inserted central venous catheters (PICCs), stopcocks, invasive-pressure external transducers, peripheral intravenous cannulas, guidewires for cardiac and peripheral vascular use, angiographic syringes, and invasive blood pressure monitoring tubing sets. These devices are designed for procedures requiring precise vascular access and real-time monitoring.

Where is Navilyst Medical, Inc. based?

Navilyst Medical, Inc. is headquartered in the United States. The company’s location ensures compliance with U.S. regulatory standards for medical device manufacturing and distribution.

Are Navilyst Medical, Inc.’s devices listed in any regulatory registries?

Yes, Navilyst Medical’s devices are included in the FDA’s Unique Device Identifier (UDI) database. This registry helps track and identify medical devices by assigning them a standardized UDI, which includes details like the manufacturer, device model, and production batch. Listing in the UDI database supports transparency and traceability in the healthcare industry.

How are Navilyst Medical, Inc.’s devices classified under U.S. regulations?

Navilyst Medical’s devices fall into two primary classification categories: Class I and Class II. Class I devices generally pose minimal risk and are subject to general controls, while Class II devices require additional premarket notifications (510(k)s) to demonstrate safety and effectiveness. The classification depends on factors like the device’s intended use, potential risks, and the body part it interacts with, as determined by FDA guidelines.

What distinguishes Navilyst Medical’s angiography-related products from other devices in their portfolio?

Navilyst Medical’s angiography kits and related products, such as angiographic syringes and guidewires, are tailored for procedures requiring detailed visualization of blood vessels. These devices are designed to work together to ensure precise contrast injection, catheter navigation, and real-time imaging during angiographic studies. Unlike general vascular access devices, they incorporate features optimized for interventional radiology and cardiology, where accuracy and reliability are critical to patient outcomes.

AI-generated from registry data and not verified by a human reviewer.

Information

Country
US
Address
—
LinkedIn
—
Facebook
—
Phone
+1(518)795-1676
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
809699023

Catalogue (7554)

Page 2 of 152
DeviceModel / ReferenceRegistriesClassStatus
Namic 60136864
FDA UDI
N/AActive
Namic 60102112
FDA UDI
N/AActive
Namic 60183274
FDA UDI
N/AActive
Namic 65120646
FDA UDI
N/AActive
Namic 65120366
FDA UDI
N/AActive
Namic 806010010
FDA UDI
Class IIActive
Namic 606209318
FDA UDI
N/AActive
Namic 60220946
FDA UDI
N/AActive
Namic 60018382
FDA UDI
N/AActive
Namic 65110292
FDA UDI
N/AActive
Namic 65101544
FDA UDI
N/AActive
Namic 65100491
FDA UDI
N/AActive
Namic 65140235
FDA UDI
N/AActive
BioFlo PKM2103716S
FDA UDI
Class IIActive
Namic 63230191
FDA UDI
N/AActive
Xcela 60M230141
FDA UDI
Class IIActive
Namic 63070071
FDA UDI
N/AActive
Vaxcel 45477
FDA UDI
Class IIUnknown
Xcela 60M100921
FDA UDI
Class IIActive
Navylist 48-105
FDA UDI
Class IIActive
Namic 601312011
FDA UDI
N/AActive
Namic 60070259
FDA UDI
N/AActive
Namic 60211486
FDA UDI
N/AActive
Namic 60050259
FDA UDI
N/AActive
Namic 60041268
FDA UDI
N/AActive
Namic 60052176
FDA UDI
N/AActive
Namic 651605515
FDA UDI
N/AActive
Namic 60110161
FDA UDI
N/AActive
Namic 60140101
FDA UDI
N/AActive
Namic 60202022
FDA UDI
N/AActive
Namic 63191071
FDA UDI
N/AActive
BioFlo Vortex CT80STBAVI
FDA UDI
Class IIUnknown
Namic 6061037731
FDA UDI
N/AActive
Navilyst 97000505
FDA UDI
Class IIUnknown
BioFlo PKM1318422S
FDA UDI
Class IIActive
Namic 65230148
FDA UDI
N/AActive
Namic PK621264S
FDA UDI
N/AActive
Namic 65190013
FDA UDI
N/AActive
Navilyst 97000208
FDA UDI
Class IIUnknown
BioFlo PKM210171S
FDA UDI
Class IIActive
BioFlo Vortex CT60LPBAVI
FDA UDI
Class IIUnknown
Namic PK121765S
FDA UDI
N/AActive
Namic 60180159
FDA UDI
N/AActive
Namic 91300020
FDA UDI
Class IIActive
Xcela 25122
FDA UDI
Class IIActive
BioFlo 60M180262
FDA UDI
Class IIActive
Namic 60060198
FDA UDI
N/AActive
Navilyst Medical PKM223523S
FDA UDI
N/AActive
Xcela 44032
FDA UDI
Class IIUnknown
Namic 60194771
FDA UDI
N/AActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (92)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

Navilyst Medical, Inc. has 92 recall records initiated between August 6, 2008, and January 9, 2015, all marked as terminated. The reasons cited include unsupported expiration dating potentially compromising sterility in Fluid Management Convenience Kits, incomplete sterilization of catheter boxes, non-coring needles causing septum damage, open syringe seals affecting sterility, sterilization barrier breaches, loose plastic particulates in fluid pathways, damaged manifolds risking air injection, and higher-than-expected catheter tip fragmentation rates in biliary drainage catheters. Some recalls were initiated following notifications from external manufacturers regarding their own product issues.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Jan 9, 2015Z-1204-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1187-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1158-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1189-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1205-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1181-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1148-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1156-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1177-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1154-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1155-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1198-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1172-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1195-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1162-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1207-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1197-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1203-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1159-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1153-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1182-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1196-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1199-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1160-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1152-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1191-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1165-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1163-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1192-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1173-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1175-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1200-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1171-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1206-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1150-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1202-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1179-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1178-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1176-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1193-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1169-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1201-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1166-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1183-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1194-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1184-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1186-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1149-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1168-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1185-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1170-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1190-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1151-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1164-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1180-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1161-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1167-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1188-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1174-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1157-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Aug 25, 2014Z-0035-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0027-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0034-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0030-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0028-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0031-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0026-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0036-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0029-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0032-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0033-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0025-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Nov 25, 2013Z-0826-2014—terminated

Navilyst Medical has recalled a Port Kit with Access Infusion Set due to a possible breach of sterilization barrier.

Dec 19, 2011Z-0821-2012—terminated

Navilyst Medical has received reports of damaged manifolds contained in certain NAMIC Convenience Kits which may result in air injection during certain procedures.

Feb 4, 2011Z-0647-2012—terminated

Navilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.

Feb 4, 2011Z-0648-2012—terminated

Navilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.

May 24, 2010Z-0015-2011—terminated

One box containing five catheters did not complete the sterilization process prior to distribution.

May 13, 2010Z-1880-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1882-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1886-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1875-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1876-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1881-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1883-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1874-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1877-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1885-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1878-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1884-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1879-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1887-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

Aug 6, 2008Z-0287-2009—terminated

Firm became aware that a specific batch of the product may contain loose plastic particulates in the fluid pathway.