What NAVILYST MEDICAL, INC. makes

Navilyst Medical produces single-use and reusable medical devices for vascular access and monitoring, including angiography kits, peripherally-inserted central venous catheters, stopcocks, invasive-pressure transducers, peripheral IV cannulas, guidewires, angiographic syringes, and blood pressure monitoring tubing sets.

The company’s portfolio includes Class I and Class II devices, all listed in the FDA’s Unique Device Identifier (UDI) database. The manufacturer is based in the United States.

Written from this company's registered devices by an AI model, and not verified by a human reviewer.

Frequently asked questions

What types of medical devices does Navilyst Medical, Inc. manufacture?

Navilyst Medical, Inc. specializes in vascular access and monitoring devices. Their portfolio includes single-use and reusable products like angiography kits, peripherally-inserted central venous catheters (PICCs), stopcocks, invasive-pressure external transducers, peripheral intravenous cannulas, guidewires for cardiac and peripheral vascular use, angiographic syringes, and invasive blood pressure monitoring tubing sets. These devices are designed for procedures requiring precise vascular access and real-time monitoring.

Where is Navilyst Medical, Inc. based?

Navilyst Medical, Inc. is headquartered in the United States. The company’s location ensures compliance with U.S. regulatory standards for medical device manufacturing and distribution.

Are Navilyst Medical, Inc.’s devices listed in any regulatory registries?

Yes, Navilyst Medical’s devices are included in the FDA’s Unique Device Identifier (UDI) database. This registry helps track and identify medical devices by assigning them a standardized UDI, which includes details like the manufacturer, device model, and production batch. Listing in the UDI database supports transparency and traceability in the healthcare industry.

How are Navilyst Medical, Inc.’s devices classified under U.S. regulations?

Navilyst Medical’s devices fall into two primary classification categories: Class I and Class II. Class I devices generally pose minimal risk and are subject to general controls, while Class II devices require additional premarket notifications (510(k)s) to demonstrate safety and effectiveness. The classification depends on factors like the device’s intended use, potential risks, and the body part it interacts with, as determined by FDA guidelines.

What distinguishes Navilyst Medical’s angiography-related products from other devices in their portfolio?

Navilyst Medical’s angiography kits and related products, such as angiographic syringes and guidewires, are tailored for procedures requiring detailed visualization of blood vessels. These devices are designed to work together to ensure precise contrast injection, catheter navigation, and real-time imaging during angiographic studies. Unlike general vascular access devices, they incorporate features optimized for interventional radiology and cardiology, where accuracy and reliability are critical to patient outcomes.

AI-generated from registry data and not verified by a human reviewer.

Information

Country
US
Address
—
LinkedIn
—
Facebook
—
Phone
+1(518)795-1676
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
—
DUNS Number
809699023

Catalogue (7554)

Page 7 of 152
DeviceModel / ReferenceRegistriesClassStatus
Namic PK191086S
FDA UDI
N/AActive
Navilyst Medical PKMR61072273S
FDA UDI
N/AActive
Namic 650501929
FDA UDI
N/AActive
Namic 62010201
FDA UDI
N/AActive
Namic 651800518
FDA UDI
N/AActive
Namic 65520232
FDA UDI
N/AActive
Namic 60195052
FDA UDI
N/AActive
Namic 60160478
FDA UDI
N/AActive
Namic 602500212
FDA UDI
N/AActive
Namic 60160549
FDA UDI
N/AActive
Namic 60122781
FDA UDI
N/AActive
BioFlo 75020
FDA UDI
Class IIActive
Navilyst 97000206
FDA UDI
Class IIUnknown
Namic PK0503710S
FDA UDI
N/AActive
Namic 601201318
FDA UDI
N/AActive
Namic 65030701
FDA UDI
N/AActive
Namic 70037200
FDA UDI
Class IIActive
Namic 601100113
FDA UDI
N/AActive
Navilyst Medical pkm137985s
FDA UDI
N/AActive
Namic 65232485
FDA UDI
N/AActive
Namic 60185473
FDA UDI
N/AActive
Namic 60184643
FDA UDI
N/AActive
Namic 65101991
FDA UDI
N/AActive
Namic 60185902
FDA UDI
N/AActive
Namic 60022025
FDA UDI
N/AActive
Namic 60121223
FDA UDI
N/AActive
Namic PK1326018S
FDA UDI
N/AActive
Namic 60040094
FDA UDI
N/AActive
Namic 60220292
FDA UDI
N/AActive
Namic 60130345
FDA UDI
N/AActive
Namic 653078083
FDA UDI
N/AActive
Navilyst Medical PKM231059S
FDA UDI
N/AActive
Namic 65161498
FDA UDI
N/AActive
Namic 6561041782
FDA UDI
N/AActive
Namic 60071242
FDA UDI
N/AActive
Namic 60140284
FDA UDI
N/AActive
Namic PK071322S
FDA UDI
N/AActive
Namic PK071241S
FDA UDI
N/AActive
Namic 60032392
FDA UDI
N/AActive
Namic 60013681
FDA UDI
N/AActive
Namic 601403919
FDA UDI
N/AActive
Namic 81051103
FDA UDI
Class IIActive
Namic 60110557
FDA UDI
N/AActive
Namic 650305412
FDA UDI
N/AActive
Envi 35510_DI
FDA UDI
Class IIActive
Namic 60150123
FDA UDI
N/AActive
Namic 60011434
FDA UDI
N/AActive
Namic 60050304
FDA UDI
N/AActive
Namic 60033132
FDA UDI
N/AActive
BioFlo 60M131847
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (92)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

Navilyst Medical, Inc. has 92 recall records initiated between August 6, 2008, and January 9, 2015, all marked as terminated. The reasons cited include unsupported expiration dating potentially compromising sterility in Fluid Management Convenience Kits, incomplete sterilization of catheter boxes, non-coring needles causing septum damage, open syringe seals affecting sterility, sterilization barrier breaches, loose plastic particulates in fluid pathways, damaged manifolds risking air injection, and higher-than-expected catheter tip fragmentation rates in biliary drainage catheters. Some recalls were initiated following notifications from external manufacturers regarding their own product issues.

AI-generated summary of the recall records listed below. It describes what the records state and is not an assessment of this manufacturer.

DateRecall No.ClassStatusReason
Jan 9, 2015Z-1204-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1187-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1158-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1189-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1205-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1181-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1148-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1156-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1177-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1154-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1155-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1198-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1172-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1195-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1162-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1207-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1197-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1203-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1159-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1153-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1182-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1196-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1199-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1160-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1152-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1191-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1165-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1163-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1192-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1173-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1175-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1200-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1171-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1206-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1150-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1202-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1179-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1178-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1176-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1193-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1169-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1201-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1166-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1183-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1194-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1184-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1186-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1149-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1168-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1185-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1170-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1190-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1151-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1164-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1180-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1161-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1167-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1188-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1174-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Jan 9, 2015Z-1157-2015—terminated

The labelled expiration dating of the affected product is not fully supported by shelf life verification testing, which may result in a sterility breach. Use of a Fluid Management Convenience Kit, where sterility has been breached, may result in an increased risk of infection.

Aug 25, 2014Z-0035-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0027-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0034-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0030-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0028-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0031-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0026-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0036-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0029-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0032-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0033-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Aug 25, 2014Z-0025-2015—terminated

NMI was notified by the manufacturer of the 10 ml BD PosiFlush SF Saline Flush Syringes, that it had initiated a recall on specific lots of the syringes as some syringe unit packages exhibited open seals which may impact package integrity and potentially affect product sterility. Based on the notification, NMI is recalling specific PICC Kits that contain the affected 10 ml BD PosiFlush SF Saline Flush Syringes.

Nov 25, 2013Z-0826-2014—terminated

Navilyst Medical has recalled a Port Kit with Access Infusion Set due to a possible breach of sterilization barrier.

Dec 19, 2011Z-0821-2012—terminated

Navilyst Medical has received reports of damaged manifolds contained in certain NAMIC Convenience Kits which may result in air injection during certain procedures.

Feb 4, 2011Z-0647-2012—terminated

Navilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.

Feb 4, 2011Z-0648-2012—terminated

Navilyst Medical, Inc. determined that the Exodus Standard Loop Biliary Drainage Catheters exhibited a rate of reported failures (catheter tip fragmentation) higher than anticipated.

May 24, 2010Z-0015-2011—terminated

One box containing five catheters did not complete the sterilization process prior to distribution.

May 13, 2010Z-1880-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1882-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1886-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1875-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1876-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1881-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1883-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1874-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1877-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1885-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1878-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1884-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1879-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

May 13, 2010Z-1887-2010—terminated

FDA testing determined that non-coring needles included in the Vaxcel Port and Vaxcel Port with PASV Valve may result in coring and resulting septum damage which could lead to patient injury.

Aug 6, 2008Z-0287-2009—terminated

Firm became aware that a specific batch of the product may contain loose plastic particulates in the fluid pathway.