Nextremity Solutions, Inc.

Active in EUDAMED
Also registered as Nextremity Solutions, Nextremity Solutions, LLC

United Statesfrom EUDAMED address EUDAMED SRN US-MF-000008441

Devices
20
Registrations
EUDAMED
FDA recalls
15

At a glance

Counted by deviCERTS from registry records

Market reach

European Union
EUDAMED actor
United States
FDA recalls

Catalogue20

Newest first, rows 1-20 of 20

Search and filter all 20 devices
Skip device list

Rows 1-20 of 20

Company details

EUDAMED SRN
US-MF-000008441
EUDAMED status
Active
EUDAMED registered name
Nextremity Solutions, Inc.
EUDAMED abbreviated name
Nextremity Solutions
PRRC contact
Travis Christman(public in EUDAMED)

FDA recalls15

Recalls are matched to the firm by name. FDA recall data as of 22 Oct 2026.

Recalls initiated per year

  • 2017 0
  • 2018 0
  • 2019 5
  • 2020 1
  • 2021 2
  • 2022 7
  • 2023 0
  • 2024 0
  • 2025 0
  • 2026 0

Showing the 10 most recent of 15 recalls.

  • 2022

    • 10 Feb 2022 Z-0910-2022 Class II Terminated Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery Distribution: Distribution US Nationwide, Japan, and the Netherlands Submission number: K110445
    • 10 Feb 2022 Z-0911-2022 Class II Terminated Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery Distribution: Distribution US Nationwide, Japan, and the Netherlands Submission number: K110445
    • 10 Feb 2022 Z-0912-2022 Class II Terminated Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery Distribution: Distribution US Nationwide, Japan, and the Netherlands Submission number: K110445
    • 10 Feb 2022 Z-0913-2022 Class II Terminated Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery Distribution: Distribution US Nationwide, Japan, and the Netherlands Submission number: K110445
    • 10 Feb 2022 Z-0914-2022 Class II Terminated Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery Distribution: Distribution US Nationwide, Japan, and the Netherlands Submission number: K110445
    • 13 Jan 2022 Z-0739-2022 Class II Terminated During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision Distribution: US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA. Submission numbers: K213530, K200840
    • 13 Jan 2022 Z-0740-2022 Class II Terminated During insertion of the middle phalanx implant into bone, the driver may deform the implant locking mechanism and result in inadequate mating force between the middle and proximal phalanx implants, resulting in surgical delay, in-situ removal of the implant, use of alternative surgical approach, disassembly of the implants post-operatively, and revision Distribution: US Nationwide distribution in the states of AZ, IL, GA, LA, MD, NY, PA, TX, WA. Submission numbers: K213530, K200840
  • 2021

    • 26 Mar 2021 Z-1500-2021 Class II Terminated Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended Distribution: AL, AK, CA, IA,IL, IN, KY ,MD. MI, NJ, NY, OR, PA, SC, TX, VA
    • 26 Mar 2021 Z-1501-2021 Class II Terminated Product could potentially contain an undersized inner diameter of the 40mm drill bushing which result in not allowing the drill bit to pass through the drill bushing, causing inability of surgeon to complete surgical procedure as intended Distribution: AL, AK, CA, IA,IL, IN, KY ,MD. MI, NJ, NY, OR, PA, SC, TX, VA
  • 2020

    • 12 Feb 2020 Z-1458-2020 Class II Terminated Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from the proximal Nextra implant Distribution: International distribution in the countries of Spain, United Kingdom, Germany, and Switzerland. Submission number: K110445
Show all 15 recalls
  • 2019

    • 16 Dec 2019 Z-1127-2020 Class II Terminated The product contains a different plate than indicated by the package labeling. The product is labeled as a right plate but contains a left plate. Distribution: Domestic distribution to AL, CA, FL, GA, ID, IL, IA, KY, LA, ME, MA, MI, MN, MS, NE, NJ, NY, OH ,OK, OR, TX, and WA. Submission number: K182201
    • 11 Nov 2019 Z-0776-2020 Class II Terminated Components 2mm shorter than indicated on the package label Distribution: Nationwide Submission number: K182201
    • 11 Nov 2019 Z-0777-2020 Class II Terminated Components 2mm shorter than indicated on the package label Distribution: Nationwide Submission number: K182201
    • 11 Nov 2019 Z-0778-2020 Class II Terminated Components 2mm shorter than indicated on the package label Distribution: Nationwide Submission number: K182201
    • 9 Oct 2019 Z-0534-2020 Class II Terminated Product labeled as STRM-NL-3550ST, NL LP Screw 3.5x50mm ST, lot M1151 may contain the incorrect screw type and length Distribution: US Nationwide distribution. Submission number: K182201

Frequently asked questions

How many devices does Nextremity Solutions, Inc. have in the registry?

The registry currently lists 20 devices for Nextremity Solutions, Inc.. Devices recorded under merged duplicate company entries are counted once.

Which registries list Nextremity Solutions, Inc.'s devices?

Among its 20 most recently registered devices, Nextremity Solutions, Inc. has devices listed in FDA 510(k) (20 devices).

What risk classes do Nextremity Solutions, Inc.'s devices fall under?

Among its 20 most recently registered devices, the classes recorded for Nextremity Solutions, Inc.'s devices are Class II. Each device page shows its own class and authorisation.

Are there FDA recall records for Nextremity Solutions, Inc.?

FDA recall records list 15 recalls initiated by Nextremity Solutions, Inc., the most recent initiated on 10 Feb 2022. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.