  "format": "devicerts.manufacturer.v1",
  "cite_as": "NOxBOX Limited. deviCERTS medical device catalogue. https://devicerts.com/manufacturers/noxbox-limited (data as of 2026-11-29).",
  "source": {
    "canonical_url": "https://devicerts.com/manufacturers/noxbox-limited",
    "json_url": "https://devicerts.com/manufacturers/noxbox-limited.json",
    "markdown_url": "https://devicerts.com/manufacturers/noxbox-limited.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "EUDAMED",
        "url": "https://ec.europa.eu/tools/eudamed"
      },
      {
        "registry": "FDA GUDID",
        "url": "https://accessgudid.nlm.nih.gov/"
      },
      {
        "registry": "FDA registration and listing",
        "url": "https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfrl/rl.cfm"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "eudamed_record_synced": "2026-11-07",
    "fda_recalls_as_of": "2026-11-01",
    "registry_facts_computed": "2026-11-29",
    "status_as_of": "2026-11-29",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/manufacturers/noxbox-limited",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "manufacturer": {
    "name": "NOxBOX Limited",
    "url": "https://devicerts.com/manufacturers/noxbox-limited",
    "registered_as": null,
    "abbreviated_name": null,
    "also_registered_as": [],
    "country": "United Kingdom",
    "country_basis": "from SRN",
    "eudamed_status": {
      "status": "active",
      "since_year": null
    },
    "verified": false,
    "roles": [
      "Manufacturer"
    ],
    "registrations": [
      "EUDAMED",
      "FDA"
    ],
    "identifiers": [
      {
        "label": "EUDAMED SRN",
        "value": "GB-MF-000009734"
      },
      {
        "label": "FDA FEI",
        "value": "3012796317"
      }
    ],
    "summary": "NOxBOX Limited is a medical device manufacturer based in the United Kingdom, with 8 devices listed in EUDAMED and FDA GUDID (EU Class IIa and IIb) and 6 FDA recalls as of 29 Nov 2026.",
    "noindex": false
  },
  "counts": {
    "devices": 8,
    "devices_note": null,
    "certificates": 0,
    "fda_recalls_matched_by_name": 6,
    "fda_recalls_ongoing": 5,
    "fda_recalls_ongoing_recent": 5,
    "fda_recalls_ongoing_before": null,
    "fda_establishments": 1,
    "sites": 0
  },
  "registry_facts": {
    "covers": "Every device of the manufacturer. A device listed in two registries counts in both.",
    "registries": [
      {
        "registry": "EUDAMED",
        "devices": 8
      },
      {
        "registry": "FDA GUDID",
        "devices": 5
      }
    ],
    "device_classes": [
      {
        "class": "EU Class IIa",
        "devices": 7
      },
      {
        "class": "EU Class IIb",
        "devices": 1
      }
    ],
    "top_emdn_categories": [
      {
        "code": "Z12",
        "name": "Instruments for functional explorations and therapeutic interventions",
        "url": "https://devicerts.com/categories/emdn/Z12",
        "devices": 8
      }
    ]
  },
  "related": {
    "site_of": null,
    "part_of": null,
    "sites": [],
    "authorised_representatives": [
      {
        "name": "Emergo Europe B.V.",
        "url": "https://devicerts.com/representatives/emergo-europe-bv-nl-ar-000000116",
        "country": "Netherlands"
      }
    ]
  },
  "devices": {
    "shown": 8,
    "total": 8,
    "all_devices_url": "https://devicerts.com/manufacturers/noxbox-limited",
    "items": [
      {
        "name": "Single-use delivery and sampling kit for Bunnell Life Pulse",
        "url": "https://devicerts.com/devices/single-use-delivery-and-sampling-kit-for-bunnell-life-pulse-noxkit-hfjv",
        "model_reference": "NOXKIT-HFJV",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED",
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Single-use 10mm delivery and sampling kit",
        "url": "https://devicerts.com/devices/single-use-10mm-delivery-and-sampling-kit-noxkit-1-10",
        "model_reference": "NOXKIT-1-10",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED",
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Single-use kit for adapting 10 mm kit to 12 mm circuits",
        "url": "https://devicerts.com/devices/single-use-kit-for-adapting-10-mm-kit-to-12-mm-circuits-noxkit-2-12",
        "model_reference": "NOXKIT-2-12",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED",
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Single-use 15mm delivery and sampling kit",
        "url": "https://devicerts.com/devices/single-use-15mm-delivery-and-sampling-kit-noxkit-2-15",
        "model_reference": "NOXKIT-2-15",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED",
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "NOxFLOW sensor for use with NOxBOXi system",
        "url": "https://devicerts.com/devices/noxflow-sensor-for-use-with-noxboxi-system-noxflow-2-f7f636da",
        "model_reference": "NOXFLOW-2",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "NOxBOXi",
        "url": "https://devicerts.com/devices/noxboxi-noxbox-i-noxdc-i",
        "model_reference": null,
        "device_class": "Class IIb",
        "registries": [
          "EUDAMED"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Emergency Backup Kit for the NOxBOXi system",
        "url": "https://devicerts.com/devices/emergency-backup-kit-for-the-noxboxi-system-noxkit-emer-bb523acf",
        "model_reference": "NOXKIT-EMER",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      },
      {
        "name": "Single-use 22mm delivery and sampling kit",
        "url": "https://devicerts.com/devices/single-use-22mm-delivery-and-sampling-kit-noxkit-2-22-f1f428bf",
        "model_reference": "NOXKIT-2-22",
        "device_class": "Class IIa",
        "registries": [
          "EUDAMED",
          "GUDID"
        ],
        "status": "Active",
        "decision_date": null,
        "open_recall": false
      }
    ]
  },
  "certificates": [],
  "recalls": [
    {
      "number": "Z-3306-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2026-09-14",
      "product_code": "MRN",
      "reason": "Potential for irregular oxygen readings due to a manufacturing issue affecting certain oxygen sensors."
    },
    {
      "number": "Z-3165-2026",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2026-08-19",
      "product_code": "MRN",
      "reason": "Potential for oxygen tubing leaks due to failure of the tubing or tubing connection, which may lead to oxygen desaturation."
    },
    {
      "number": "Z-0028-2026",
      "classification": "Class I",
      "status": "Ongoing",
      "initiated": "2025-09-09",
      "product_code": "MRN",
      "reason": "Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz)."
    },
    {
      "number": "Z-0029-2026",
      "classification": "Class I",
      "status": "Ongoing",
      "initiated": "2025-09-09",
      "product_code": "MRN",
      "reason": "The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts."
    },
    {
      "number": "Z-2544-2025",
      "classification": "Class II",
      "status": "Ongoing",
      "initiated": "2023-07-19",
      "product_code": "MRN",
      "reason": "Replacement of the internal Sample Pump due to high failure rates."
    },
    {
      "number": "Z-1667-2023",
      "classification": "Class I",
      "status": "Terminated",
      "initiated": "2023-05-03",
      "product_code": "MRN",
      "reason": "Device may experience malfunctions due to misalignment of the check valve in the manifold. If the device fails, there is a risk of nitric oxide leak, oxygen gas leak, or earlier cylinder gas changeover than expected or not at all. In a worst case scenario, there may be an interruption in therapy which could lead to oxygen desaturation of the patient and associated complications."
    }
  ]
}