Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

OAKWORKS, INC.

US

Last updated September 17, 2026

Information

Country
US
Address
923 E Wellspring Rd, NEW FREEDOM, PA, 17349
Website
—
LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
1000121370
DUNS Number
130256837

Catalogue (398)

Page 1 of 8
DeviceModel / ReferenceRegistriesClassStatus
Oakworks Medical DTPM300
FDA UDI
Class IIActive
Oakworks Medical CFPM301
FDA UDI
Class IIActive
Ultrasound Ergonomic Advantage product line Ultrasound EA General 3-Section Table
FDA UDI
Class IActive
Oakworks Medical CFPM300
FDA UDI
Class IIActive
Imaging and Pain Management CFPM 201
FDA UDI
Class IIActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Oakworks Medical CFPM300
FDA UDI
Class IIActive
Oakworks Medical CFUR301
FDA UDI
Class IIActive
Oakworks Medical CFPM300
FDA UDI
Class IIActive
Imaging and Pain Management CFPM201
FDA UDI
Class IIActive
Imaging and Pain Management CFPM200
FDA UDI
Class IIActive
Oakworks Medical CFPMB301
FDA UDI
Class IIActive
Oakworks Medical DTPM300
FDA UDI
Class IIActive
Oakworks Medical DTPM300
FDA UDI
Class IIActive
Oakworks Medical CFPMFXH
FDA UDI
Class IIUnknown
Physical Therapy (PT) product line Accessory 72178
FDA UDI
Class IActive
Oakworks Medical Fluoro Extender
FDA UDI
Class IIActive
PCG series 3100 Procedure Chair / PCG Chair
FDA UDI
Class IActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Imaging and Pain Management CFPM 200
FDA UDI
Class IIActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Oakworks Medical CFPM300
FDA UDI
Class IIActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Oakworks Medical CFPMFXH
FDA UDI
Class IIUnknown
Imaging and Pain Management CFPM 302
FDA UDI
Class IIActive
Ultrasound Ergonomic Advantage product line Ultrasound EA General 3-Section Top Table
FDA UDI
Class IActive
Oakworks Medical Carbon Fiber Arm Board
FDA UDI
Class IIUnknown
Ultrasound Ergonomic Advantage product line EA Vascular Table with Fowler (V2 Model)
FDA UDI
Class IActive
Oakworks Medical CFPM300
FDA UDI
Class IIActive
Oakworks Medical DTPM300
FDA UDI
Class IIActive
Imaging and Pain Management CFPM 200
FDA UDI
Class IIActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Imaging and Pain Management CFPM 201
FDA UDI
Class IIActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Powerline Powerline Flat Top Exam Table
FDA UDI
Class IActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Physical Therapy (PT) product line PT250 Table
FDA UDI
Class IActive
Imaging and Pain Management CFPM 201
FDA UDI
Class IIActive
Physical Therapy (PT) product line PT400 Table
FDA UDI
Class IUnknown
Oakworks Medical CFPM301
FDA UDI
Class IIActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Oakworks Medical CFPM400
FDA UDI
Class IIActive
Physical Therapy (PT) product line PT300 Table
FDA UDI
Class IActive
Oakworks Medical CFPM401
FDA UDI
Class IIActive
Imaging and Pain Management CFPM 200
FDA UDI
Class IIActive
Imaging and Pain Management CFPM 302 Standard
FDA UDI
Class IIActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (9)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
Dec 5, 2023Z-0730-2024—open, classified

Backrest platform stress fracture - the table is not operational in an inclined position

Dec 5, 2023Z-0728-2024—open, classified

Backrest platform stress fracture - the table is not operational in an inclined position

Dec 5, 2023Z-0729-2024—open, classified

Backrest platform stress fracture - the table is not operational in an inclined position

Dec 5, 2023Z-0727-2024—open, classified

Backrest platform stress fracture - the table is not operational in an inclined position

Apr 21, 2022Z-1238-2022—terminated

Product label was printed without the UDI number and warning label was printed with the incorrect weight capacity. The correct weight limit is 400 lbs., but the label states 500 lbs.

Apr 21, 2022Z-1239-2022—terminated

The warning label was printed with the incorrect weight capacity. The correct weight limit is 400 lbs., but the label states 500 lbs.

Jun 29, 2021Z-2349-2021—terminated

Weld failure- the bracket attaches the tabletop tilting actuator to the tabletop frame. If the bracket weld fails, the tabletop tilting action is non-functional

May 2, 2019Z-1411-2019—terminated

Table assembly error with incorrect actuators for the lateral and longitudinal table movements

Jul 31, 2017Z-0703-2018—terminated

The foot control used on the OAKWORKS Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table.