OriGen Biomedical Inc.
Active in EUDAMEDUnited Statesfrom SRN EUDAMED FDA SRN US-MF-000018098 FEI 3042484745
- Devices
- 296
- Registrations
- EUDAMEDFDA
- FDA recalls
- 201 ongoing
Catalogue
Search and filter all 296 devicesShowing 251-296 of 296
Company details
- EUDAMED SRN
- US-MF-000018098
- FDA FEI number
- 3042484745
- DUNS number
- 625272232
- EU VAT number
- NL825807803B01
- EUDAMED status
- Active
- EUDAMED registered name
- OriGen Biomedical Inc.
- EUDAMED abbreviated name
- OriGen Biomedical Inc.
- Registered address
- 2215, Grand Avenue Parkway, 78728, Austin, United States
- Website
- www.origen.com/
- facebook.com/OriGenBiomedical
- [email protected]
- Phone
- +1 512-474-7278
- PRRC contact
- Sam Saleh
FDA recalls
Recalls initiated per year
- 2017 7
- 2018 7
- 2019 1
- 2020 0
- 2021 1
- 2022 0
- 2023 1
- 2024 0
- 2025 0
- 2026 0
2023
- 19 Dec 2023 Z-1068-2024 Class II Ongoing Accessory sets contain a syringes not yet qualified for the accessory's intended purpose; due to design changes to the syringe barrels and a polypropylene resin change that resulted in loss of gamma compatibility, for accessory set syringes that were gamma sterilized, which resulted in yellowed syringes.
2021
- 28 Jan 2021 Z-1409-2021 Class III Terminated It was identified that tissue culture bags were incorrectly labeled as free of phthalates.
2019
- 24 Jun 2019 Z-0318-2021 Class II Terminated The firm has become aware that they have distributed product that was expired.
2018
- 21 Aug 2018 Z-0102-2019 Class II Terminated The catheter failed the endotoxin testing.
Full distribution text
- 21 Aug 2018 Z-0179-2019 Class II Terminated These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Full distribution text
- 21 Aug 2018 Z-0180-2019 Class II Terminated These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Full distribution text
- 21 Aug 2018 Z-0181-2019 Class II Terminated These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Full distribution text
- 21 Aug 2018 Z-0182-2019 Class II Terminated These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Full distribution text
- 21 Aug 2018 Z-0183-2019 Class II Terminated These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Full distribution text
- 21 Aug 2018 Z-0184-2019 Class II Terminated These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Full distribution text
- 21 Aug 2018 Z-0102-2019 Class II Terminated The catheter failed the endotoxin testing.
Show all 20 recalls
2017
- 2 Aug 2017 Z-0021-2018 Class I Terminated Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO
Catheter has had a separation of the clear extension tube from the hub connection. These
adverse events resulted in patient injury (blood loss).
Full distribution text
- 17 Jan 2017 Z-1227-2019 Class I Terminated Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Full distribution text
- 17 Jan 2017 Z-1228-2019 Class I Terminated Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Full distribution text
- 17 Jan 2017 Z-1229-2019 Class I Terminated Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Full distribution text
- 17 Jan 2017 Z-1230-2019 Class I Terminated Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Full distribution text
- 17 Jan 2017 Z-1231-2019 Class I Terminated Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Full distribution text
- 17 Jan 2017 Z-1232-2019 Class I Terminated Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.
Full distribution text
- 2 Aug 2017 Z-0021-2018 Class I Terminated Two users have reported to OriGen that a VV28F OriGen Reinforced Dual Lumen ECMO
Catheter has had a separation of the clear extension tube from the hub connection. These
adverse events resulted in patient injury (blood loss).
2015
- 30 Mar 2015 Z-0269-2019 Class I Terminated Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage
Full distribution text
- 30 Mar 2015 Z-1456-2015 Class I Terminated Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage.
- 30 Mar 2015 Z-0269-2019 Class I Terminated Potential for clear extension tube to separate from the hub, which could result in blood loss and may require intervention to prevent permanent impairment/damage
2002
- 16 Dec 2002 Z-0055-06 Unclassified (pre-2012) Terminated Leak between two lumens causes reduced performance.
Full distribution text
- 16 Dec 2002 Z-0055-06 Unclassified (pre-2012) Terminated Leak between two lumens causes reduced performance.
FDA establishments
- OriGen Biomedical, Inc. Austin, TX · US FEI 1000160256
From the company website
OriGen Biomedical manufactures cryo bags for cryopreservation of blood components, cell culture bags, dimethylsulfoxide, and respiratory products.
Frequently asked questions
How many devices does OriGen Biomedical Inc. have in the registry?
The registry currently lists 296 devices for OriGen Biomedical Inc.. Devices recorded under merged duplicate company entries are counted once.
Are there FDA recall records for OriGen Biomedical Inc.?
FDA recall records list 20 recalls initiated by OriGen Biomedical Inc., the most recent initiated on 19 Dec 2023. 1 of them is recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
Which FDA establishments are linked to OriGen Biomedical Inc.?
1 FDA establishment is linked to OriGen Biomedical Inc. in the registry; they are listed on the company page.