Orthoscan, Ltd.

United Statesfrom the FDA registration address FDA FEI 3005254598

Devices
6
Registrations
FDA
FDA recalls
117 ongoing

At a glance

Counted by deviCERTS from registry records

Market reach

United States
FDA FEI, FDA recalls

Catalogue6

Newest first, rows 1-6 of 6

  • Orthoscan Tau Mini C-Arm Ref K213113

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Orthoscan Mobile Di Mini C-Arm Ref K113708

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Orthoscan Fd Mini C-Arm Ref K133174

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • OrthoScan TAU Mini C-Arm Ref K183220

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Orthoscan, Orthoscan Hd Ref K051754

    Active Regimes: 510(k) Registry: 510(k) Class:II
  • Ortelius 800 Ref K011827

    Active Regimes: 510(k) Registry: 510(k) Class:I

Company details

FDA FEI number
3005254598
Address
14555 N. 82nd Street, SCOTTSDALE, AZ, 85260

FDA recalls11

Recalls are matched to the firm by name. FDA recall data as of 22 Oct 2026.

Recalls initiated per year

  • 2017 4
  • 2018 0
  • 2019 0
  • 2020 0
  • 2021 0
  • 2022 0
  • 2023 0
  • 2024 7
  • 2025 0
  • 2026 0

Showing the 10 most recent of 11 recalls.

  • 2024

    • 8 Aug 2024 Z-0573-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K092438
    • 8 Aug 2024 Z-0574-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K234109
    • 8 Aug 2024 Z-0575-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K073346
    • 8 Aug 2024 Z-0576-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K240020
    • 8 Aug 2024 Z-0577-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K061203
    • 8 Aug 2024 Z-0578-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K240099
    • 8 Aug 2024 Z-0579-2025 Class II Ongoing Ziehm Imaging, Inc., distributor, have been informed by manufacturer, Ziehm Imaging GmbH that installed devices may not fully meet the requirements of IEC 60601-2-54 and or 21 CFR 1020.32(c) with respect to wired hand switch may fail when being exposed to mechanical stress such as shocks (e.g. dropping) or impacts can irreversibly damage the switching element in the hand switch. A defective switching element may then potentially initiate unintended radiation of the connected mobile X-ray unit. Distribution: U.S. (including Puerto Rico). Submission number: K231701
  • 2017

    • 31 Aug 2017 Z-0258-2018 Class II Terminated OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user Distribution: Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria,…
      Full distribution text Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.
      Submission number: K051754
    • 31 Aug 2017 Z-0259-2018 Class II Terminated OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user Distribution: Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria,…
      Full distribution text Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.
      Submission number: K051754
    • 31 Aug 2017 Z-0260-2018 Class II Terminated OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user Distribution: Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria,…
      Full distribution text Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.
      Submission number: K113708
Show all 11 recalls
  • 2017

    • 31 Aug 2017 Z-0261-2018 Class II Terminated OrthoScan Inc. discovered during investigation of a non-standard work flow a non-conformity of the devices. Specifically, the system software allows a user to activate the Digital Zoom feature in the course of taking a live image exposing a larger x-ray field than can be viewed by the user Distribution: Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria,…
      Full distribution text Worldwide Distribution - USA (nationwide) including Puerto Rico and Hawaii and to the countries of : Australia, Austria, Belgium, Canada, Dominican Republic, Estonia , Finland, France, Germany, Hanover, Hong Kong, Israel, Italy, Japan, Jonesboro, Korea, Malaysia. Netherland, New Zealand, Norway, Poland, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Taiwan, UAE, and United Kingdom.
      Submission number: K133174

Frequently asked questions

How many devices does Orthoscan, Ltd. have in the registry?

The registry currently lists 6 devices for Orthoscan, Ltd.. Devices recorded under merged duplicate company entries are counted once.

Which registries list Orthoscan, Ltd.'s devices?

Among its 6 most recently registered devices, Orthoscan, Ltd. has devices listed in FDA 510(k) (6 devices).

What risk classes do Orthoscan, Ltd.'s devices fall under?

Among its 6 most recently registered devices, the classes recorded for Orthoscan, Ltd.'s devices are Class I and Class II. Each device page shows its own class and authorisation.

Are there FDA recall records for Orthoscan, Ltd.?

FDA recall records list 11 recalls initiated by Orthoscan, Ltd., the most recent initiated on 8 Aug 2024. 7 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.

What are the newest devices from Orthoscan, Ltd. in the registry?

The most recently registered devices listed for Orthoscan, Ltd. include Orthoscan Tau Mini C-Arm, ORTHOSCAN MOBILE DI MINI C-ARM, ORTHOSCAN FD MINI C-ARM and ORTHOSCAN, ORTHOSCAN HD.

Generated automatically from registry data and not verified by a human reviewer.