Radiometer Medical ApS
Country not established
Radiometer Medical ApS is a medical device manufacturer, with 355 devices listed in FDA GUDID, EUDAMED and FDA 510(k) (FDA Class II and I; EU Class A and I) and 26 FDA recalls as of 25 Nov 2026.
- Devices
- 355
- FDA recalls
- 269 ongoing since 2021 (10 open in total)
Registry facts
- Registries
- FDA GUDID 299
- EUDAMED 81
- FDA 510(k) 35
- Device classes
- FDA Class II 266
- FDA Class I 68
- EU Class A 11
- EU Class I 7
- EU Class IIa 2
- EU Class IIb 1
- Top EMDN categories
- FDA recalls
- Sites and related companies
- Radiometer Medical Aps Denmark
- Identifiers
- DUNS 308667539
At a glance
Market reach
- United States
- FDA recalls
What Radiometer Medical ApS makes
Radiometer Medical ApS produces in vitro diagnostics (IVDs) for blood gas analysis, haemoximetry, and electrolyte measurement, including portable and benchtop analysers, control materials, calibrators, and consumables like sampling kits and capillary tubes. Their portfolio also includes transcutaneous blood gas monitoring sensors, electrical connection cables for medical devices, and clinical chemistry analysers for metabolic profiling. Additional products encompass cleaning solutions for medical devices and calibration gases for analytical instruments.
The company’s devices are classified under FDA categories as Class I and Class II, with regulatory submissions documented via 510(k) premarket notifications and Unique Device Identifiers (UDIs). These products are listed in the FDA’s regulatory databases, ensuring compliance with US medical device regulations.
Catalogue
Search and filter all 355 devicesShowing 1-50 of 355
Company details
- DUNS number
- 308667539
- Phone
- +459999999999
FDA recalls
FDA recall records list 26 recalls initiated by Radiometer Medical ApS, the earliest on 2018-02-27 and the most recent on 2026-08-18. By recorded status: 10 ongoing and 16 terminated. The reasons filed fall into these groups: labeling or instructions for use, software or firmware, sterility or contamination, electrical or power and performance or accuracy. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.
Recalls initiated per year
- 2017 0
- 2018 1
- 2019 3
- 2020 12
- 2021 4
- 2022 2
- 2023 1
- 2024 1
- 2025 1
- 2026 1
2026
- 18 Aug 2026 Z-0190-2027 Class II Ongoing In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.
Full distribution text
- 18 Aug 2026 Z-0190-2027 Class II Ongoing In vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer suitable for testing FHbF and ctBil, 2) Reportable ranges for some parameters, 3) Analyzer configuration, and 4) limitations and warnings.
2025
- 21 Mar 2025 Z-1659-2025 Class II Ongoing ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
Full distribution text
- 21 Mar 2025 Z-1659-2025 Class II Ongoing ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
2024
- 31 May 2024 Z-2426-2024 Class II Ongoing Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.
Full distribution text
- 31 May 2024 Z-2426-2024 Class II Ongoing Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.
2023
- 8 Mar 2023 Z-1394-2023 Class II Ongoing Due to potential software issue that may result in patient mix-up information.
Full distribution text
- 8 Mar 2023 Z-1394-2023 Class II Ongoing Due to potential software issue that may result in patient mix-up information.
2022
- 18 May 2022 Z-1321-2022 Class II Ongoing There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.
Full distribution text
- 8 Apr 2022 Z-1115-2022 Class II Ongoing Arterial Blood Sampler sterility cannot be guaranteed throughout the shelf life of the product.
Full distribution text
- 18 May 2022 Z-1321-2022 Class II Ongoing There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.
2021
- 14 Apr 2021 Z-1856-2021 Class II Ongoing Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Full distribution text
- 14 Apr 2021 Z-1857-2021 Class II Ongoing Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Full distribution text
- 14 Apr 2021 Z-1858-2021 Class II Ongoing Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Full distribution text
- 26 Mar 2021 Z-1683-2021 Class II Terminated Radiometer has become aware of an issue with PICO70 Arterial Blood Sampler. The issue relates to the sterile barrier system for the product. During internal stress testing on PICO70 Arterial Blood Samplers a breach in the sterile barrier system was observed in one sample. Risk for the patient: The described error is considered to have a remote possibility of resulting in bacterial bloodstream infection. The bloodstream infection may be asymptomatic but may also progress to sepsis or life-threatening septic shock. Immunosuppressed patients are at particular risk. Due to issues (breaches) with the sterile barrier system for the Arterial Blood Sampler.
Full distribution text
- 14 Apr 2021 Z-1856-2021 Class II Ongoing Due to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Show the 25 most recent of 26 recalls
2020
- 26 May 2020 Z-2493-2020 Class II Terminated Potential risk of patient mix-up on analyzers due to software issues.
Full distribution text
- 26 May 2020 Z-2494-2020 Class II Terminated Potential risk of patient mix-up on analyzers due to software issues.
Full distribution text
- 11 May 2020 Z-2467-2020 Class II Terminated Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.
Full distribution text
- 11 May 2020 Z-2468-2020 Class II Terminated Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.
Full distribution text
- 17 Mar 2020 Z-2247-2020 Class II Terminated The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost samples.
Full distribution text
- 17 Mar 2020 Z-2248-2020 Class II Terminated The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost samples.
Full distribution text
- 17 Mar 2020 Z-2249-2020 Class II Terminated The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost samples.
Full distribution text
- 10 Mar 2020 Z-1684-2020 Class II Terminated Analyzer's barcode reader misinterprets the contents of barcode label used for entering patient ID or accession number. The issue is related to barcode types not using a check digit. This could result in patient mixup or loss of sample resulting in delayed medical treatment
Full distribution text
- 6 Mar 2020 Z-1836-2020 Class II Open since 2020 Through internal investigation is has been determined that an incorrect component was used within the power supply. The incorrect component may cause a short circuit within the power supply resulting in a delay in treatment. The delay could result in a range of risk from cyanosis and low blood pressure to cardiac arrhythmias and severe impairment of cardiac function.
- 20 Feb 2020 Z-1755-2020 Class II Terminated Reports have been received about the transcutaneous monitoring system display flickering or monitor becoming non responsive and beeping. This may result in delayed patient monitoring as well as potentially compromise the product performance.
Full distribution text
- 3 Feb 2020 Z-1760-2020 Class II Terminated The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
Full distribution text
- 3 Feb 2020 Z-1761-2020 Class II Terminated The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
Full distribution text
- 26 May 2020 Z-2493-2020 Class II Terminated Potential risk of patient mix-up on analyzers due to software issues.
2019
- 21 Nov 2019 Z-2138-2020 Class II Terminated The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
- 21 Nov 2019 Z-2139-2020 Class II Terminated The analyzers fail on S7950 (Range +) QC measurements due to the hemolyzer cuvettes, a subcomponent of the optical system which measures ctHb and derivates. In a cuvette with light shunt, some of the received light has not been transmitted through the sample. This will lead to a perturbed and non-linear absorbance spectrum which results in measurement errors.
- 8 Aug 2019 Z-0864-2020 Class II Terminated The firm has become aware that there is a potential problem relating to the blood gas and immunoassay analyzer Systems that may result in patient mix-up when connected to some third-party devices. The error may lead to serious adverse health consequences for the patient caused by patient data mix-up.
Full distribution text
Sites
- Radiometer Medical Aps DKFEI 3002807968 21 devices
Frequently asked questions
How many devices does Radiometer Medical ApS have in the registry?
The registry currently lists 355 devices for Radiometer Medical ApS. Devices recorded under merged duplicate company entries are counted once.
Which registries list Radiometer Medical ApS's devices?
Among its 50 most recently registered devices, Radiometer Medical ApS has devices listed in EUDAMED (25 devices) and FDA GUDID (29 devices). A device can be listed in more than one registry, so these counts overlap.
What risk classes do Radiometer Medical ApS's devices fall under?
Among its 50 most recently registered devices, the classes recorded for Radiometer Medical ApS's devices are Class I, Class IIa, Class IIb, Class II and IVDR Class A. Each device page shows its own class and authorisation.
Are there FDA recall records for Radiometer Medical ApS?
FDA recall records list 26 recalls initiated by Radiometer Medical ApS, the most recent initiated on 18 Aug 2026. 10 of them are recorded as ongoing. A recall is often a precaution started by the manufacturer; the records say nothing about the company's overall record.
What are the newest devices from Radiometer Medical ApS in the registry?
The most recently registered devices listed for Radiometer Medical ApS include S7620 QUALICHECK 7+, level 0 box of 10 ampoules for ABL90 FLEX, ABL90 944-519, S7630 QUALICHECK 7+, level 1 box of 10 ampoules for ABL90 FLEX, ABL90 944-520, S7640 QUALICHECK 7+, level 2 box of 10 ampoules for ABL90 FLEX, ABL90 944-521 and S7660 QUALICHECK 7+, level 4 box of 10 ampoules for ABL90 FLEX, ABL90 944-523.