Stanmore Implants Worldwide Ltd
Country not established
- Devices
- 103
- FDA recalls
- 8
What Stanmore Implants Worldwide Ltd makes
Stanmore Implants Worldwide Ltd manufactures modular orthopedic implants for the femur and knee, including extendable and patient-specific distal femoral implants, modular proximal and total femur systems, and a modular total knee replacement. Their portfolio also includes robotic guidance arms designed to assist in surgical planning and precision during joint replacement procedures.
These devices are classified as Class II under FDA regulations and are listed via the 510(k) pathway. The company operates from the United States, where its products are authorised for use in orthopedic surgeries.
Catalogue
Search and filter all 103 devicesSculptor Robotic Guidance Arm (Rga) Ref K121765
Regimes: 510(k) Registry: 510(k)Mets Modular Distal Femur Ref K121029
Regimes: 510(k) Registry: 510(k)Mets Smiles Total Knee Replacement Ref K120992
Regimes: 510(k) Registry: 510(k)
Showing 101-103 of 103
Company details
- DUNS number
- 494989502
FDA recalls
FDA recall records list 8 recalls initiated by Stanmore Implants Worldwide Ltd, the earliest on 2016-04-12 and the most recent on 2017-07-04. By recorded status: 8 terminated. The reasons filed fall into these groups: packaging and manufacturing or quality process. A recall is often a precaution started by the manufacturer, and a count of records is not a verdict on a company.
Recalls initiated per year
- 2017 1
- 2018 0
- 2019 0
- 2020 0
- 2021 0
- 2022 0
- 2023 0
- 2024 0
- 2025 0
- 2026 0
2017
- 4 Jul 2017 Z-3104-2017 Class II Terminated Three (3) complaints were identified in which the potential for utilization of the JTS Grower Drive in the incorrect orientation in relation to the limb entry point of the device, together with the potential to select incorrect settings for the extension procedure. As a result, the prostheses in such cases failed to extend as intended.
2016
- 7 Dec 2016 Z-0820-2017 Class II Terminated The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- 7 Dec 2016 Z-0821-2017 Class II Terminated The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- 7 Dec 2016 Z-0822-2017 Class II Terminated The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- 7 Dec 2016 Z-0823-2017 Class II Terminated The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- 7 Dec 2016 Z-0824-2017 Class II Terminated The recalling firm has received 5 customer inquiries reporting that at least one pouch of the double pouch packaging configuration has been punctured by the tip of the device's stem.
- 31 May 2016 Z-2553-2016 Class II Terminated The manufacturer has identified that the stated Femoral Plateau Plates may not fit into the intended Femoral Knee mating component and thus sit flush upon the Femoral Knee component prior to implantation.
- 12 Apr 2016 Z-1623-2016 Class II Terminated Incorrect component used to manufacture distal femoral replacement.