Is this your brand? Sign in to claim ownership.

Sign in to claim this brand

Terumo Cardiovascular Systems Corporation

US

Last updated September 17, 2026

Information

Country
US
Address
6200 Jackson Road, ANN ARBOR, MI, 48103
Website
www.terumo-cvs.com

This website did not respond when we last checked — the domain may no longer be active. Last checked 2026-09-18.

LinkedIn
—
Facebook
—
Email
—
Phone
—
PRRC Contact
—
EUDAMED SRN
—
FDA FEI Number
1828100
DUNS Number
—

Catalogue (5925)

Page 1 of 119
DeviceModel / ReferenceRegistriesClassStatus
Cardiovascular Procedure Kit CX-K31
FDA UDI
N/AActive
Cardiovascular Procedure Kit 76787
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-IS055X
FDA UDI
N/AActive
Cardiovascular Procedure Kit 73151-02
FDA UDI
N/AActive
Cardiovascular Procedure Kit B71964-04
FDA UDI
N/AActive
Cardiovascular Procedure Kit 75071
FDA UDI
N/AActive
Cardiovascular Procedure Kit 66203
FDA UDI
N/AActive
Cardiovascular Procedure Kit 66039
FDA UDI
N/AActive
Cardiovascular Procedure Kit 66527
FDA UDI
N/AActive
Cardiovascular Procedure Kit 75999
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-NL155X
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-GE242
FDA UDI
N/AActive
Cardiovascular Procedure Kit 77294
FDA UDI
N/AActive
Cardiovascular Procedure Kit 74372-02
FDA UDI
N/AActive
Cardiovascular Procedure Kit 77383-01
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-GG046
FDA UDI
N/AActive
Terumo® Advanced Perfusion System 1 164267
FDA UDI
Class IIActive
Cardiovascular Procedure Kit CX-UK618
FDA UDI
N/AActive
Cardiovascular Procedure Kit 70893-09
FDA UDI
Class IIActive
Cardiovascular Procedure Kit CX-IT254X
FDA UDI
N/AActive
Cardiovascular Procedure Kit 77369
FDA UDI
N/AActive
Cardiovascular Procedure Kit 76705
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-SP184X
FDA UDI
N/AActive
Cardiovascular Procedure Kit 66742
FDA UDI
N/AActive
Cardiovascular Procedure Kit 66089
FDA UDI
N/AActive
Sarns™ Modular Perfusion System 8000 7278
FDA UDI
Class IIActive
Cardiovascular Procedure Kit 77375
FDA UDI
N/AActive
Terumo® Advanced Perfusion System 1 802110
FDA UDI
Class IIActive
Cardiovascular Procedure Kit 75364
FDA UDI
N/AActive
Cardiovascular Procedure Kit 70365-03
FDA UDI
N/AActive
Cardiovascular Procedure Kit 76138
FDA UDI
N/AActive
Cardiovascular Procedure Kit 76390-02
FDA UDI
N/AActive
Cardiovascular Procedure Kit 74321-01
FDA UDI
N/AActive
Cardiovascular Procedure Kit 62780
FDA UDI
N/AActive
Cardiovascular Procedure Kit 71270-02
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-RU100XC
FDA UDI
N/AActive
Cardiovascular Procedure Kit 65449-02
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-UK550
FDA UDI
N/AActive
Cardiovascular Procedure Kit 77435-01
FDA UDI
N/AActive
Cardiovascular Procedure Kit 72891
FDA UDI
N/AActive
Cardiovascular Procedure Kit 70658-09
FDA UDI
N/AUnknown
Cardiovascular Procedure Kit CX-RU100XA
FDA UDI
N/AActive
Cardiovascular Procedure Kit 76202-02
FDA UDI
Class IIActive
Cardiovascular Procedure Kit CX-DB469
FDA UDI
N/AActive
Cardiovascular Procedure Kit 76999
FDA UDI
N/AActive
Cardiovascular Procedure Kit 71975
FDA UDI
N/AActive
Cardiovascular Procedure Kit 78418
FDA UDI
N/AActive
Cardiovascular Procedure Kit CX-TR027X
FDA UDI
N/AActive
Cardiovascular Procedure Kit 72624-03
FDA UDI
N/AActive
Cardiovascular Procedure Kit 75492
FDA UDI
N/AActive

Related Companies

Authorised Representatives

No authorised representatives on record.

Importers

Not available yet.

Notified Bodies

Not available yet.

Authorities

Not available yet.

FDA Recalls (100)

Recalls reported by this firm to the FDA. Listed at company level — FDA recall records don't identify a specific catalogue device.

DateRecall No.ClassStatusReason
May 28, 2025Z-2018-2025—open, classified

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

Mar 3, 2025Z-1492-2025—open, classified

During some clinical cases, it has been observed that the SO2 value may periodically appear as dashes on the Touchscreen Display indicating unavailable data. In these cases, a yellow low-limit alarm will also be triggered to the user, due to the dashed-out value.

Feb 20, 2024Z-1401-2024—open, classified

Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention

Mar 1, 2023Z-1346-2023—open, classified

Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.

Mar 1, 2023Z-1347-2023—open, classified

Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.

Mar 1, 2023Z-1345-2023—open, classified

Some Capiox NX, RX and FX Oxygenators may contain an arterial thermistor component that is improperly bonded to the oxygenator port. This may result in a leak at the point where the arterial thermistor is improperly bonded to the oxygenator port.

Feb 3, 2022Z-0949-2022—terminated

Potential for Cardiovascular Procedure Kit packaging damage that occurred during shipping.

Jun 8, 2021Z-2350-2021—terminated

intermittent failure causes a blender initialization fault upon start-up, or during an EPGS calibration attempt following the 15-minute warm-up period. The control sliders on the Central Control Monitor (CCM) are disabled and the gas system can only be controlled by the local control knobs

Apr 16, 2021Z-1790-2021—terminated

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Apr 16, 2021Z-1791-2021—terminated

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Apr 16, 2021Z-1789-2021—terminated

Terumo CVS has been unable to validate a cleaning protocol to satisfy current regulatory concerns and expectations. As a result, an updated cleaning protocol will not be developed by Terumo CVS and it has been determined that the best course of action is for users to discontinue use of and dispose of HX2, TCM I and TCM II devices.

Mar 2, 2021Z-1367-2021—terminated

There is a potential for the stainless steel locking plates on the device's sternal retractor locking mechanism to fracture.

Nov 11, 2020Z-0565-2021—terminated

Capiox FX25 Advance oxygenator may be the incorrect configuration (East vs. West)

Aug 25, 2020Z-2960-2020—terminated

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Aug 25, 2020Z-2959-2020—terminated

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Aug 25, 2020Z-2961-2020—terminated

Internal testing at the supplier, revealed that a crimped wire was placed into an air bubble detector (ABD) sensor connector backwards, preventing the crimped wire from locking into the mating wire connector. This intermittent connection may result in false air bubble alarms.

Jan 10, 2020Z-1109-2020—terminated

These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

Jan 10, 2020Z-1110-2020—terminated

These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

Jan 10, 2020Z-1108-2020—terminated

These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

Jan 10, 2020Z-1111-2020—terminated

These roller pumps were assembled with pump shafts that may fail at the location of the thread where a locknut is installed. The failure is associated with a noise that is atypical of roller pump operation. The noise is caused when an unsecured shaft end contacts the rotating pulley of the roller pump. Prolonged use of a roller pump in this condition could lead to an eventual pump stop, resulting in the inability to create pump flow due to pulley loosening and drive belt damage.

Jun 14, 2019Z-2102-2019—terminated

Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component

Jun 14, 2019Z-2104-2019—terminated

Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component

Jun 14, 2019Z-2103-2019—terminated

Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component

Jun 14, 2019Z-2101-2019—terminated

Capiox NX19 Oxygenator may contain a non-functioning arterial thermistor component

Apr 18, 2019Z-1337-2019—terminated

The gas flow rate output of the EPGS may be inaccurate due to a defect in the internal flowmeter.

Dec 20, 2018Z-0966-2019—terminated

Presence of natural rubber latex is not declared in the label

Oct 17, 2018Z-1769-2019—terminated

An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.

Sep 13, 2018Z-0176-2019—terminated

The service manual was not updated at the time of the release of software version 1.30.

Aug 31, 2018Z-0227-2019—terminated

The system's IFU contains a list of generators that the have been deemed compatible. Recent reports indicate that one energy platform is not compatible with the system. Therefore, it has been removed from the list of compatible generators within the IFU.

Aug 28, 2018Z-3242-2018—terminated

Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.

Aug 28, 2018Z-3243-2018—terminated

Screws on a Terumo Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system

Jun 26, 2018Z-2751-2018—terminated

H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.

Jun 26, 2018Z-2750-2018—terminated

H/Sat Hybrid Intensity Shift has been seen in a subset of the H/Sat population, which has resulted in an increase of Hybrids failing the H/Sat Color Chip Test which is an automated self-test when the device is powered on. This will not result in device inaccuracy, only Color Chip Test Failure which disables the H/Sat module.

Mar 16, 2018Z-1418-2018—terminated

Update to cooler-heater cleaning instructions.

Mar 16, 2018Z-1416-2018—terminated

Update to cooler-heater cleaning instructions.

Mar 16, 2018Z-1417-2018—terminated

Update to cooler-heater cleaning instructions.

Jan 19, 2018Z-1478-2018—terminated

Inaccurrate flow readings. Depending on the degree of inaccuracy, this issue may not be easy for the user to detect during setup or use (for example, following a Flow Probe relocation or manipulation).

Dec 18, 2017Z-1314-2018—terminated

Possible blood leaks through the hydrophobic portion of the Pall LG6NS LeukoGuard Leukocyte Reduction Arterial Blood Filters.

Jun 27, 2017Z-0256-2018—terminated

Terumo Cardiovascular Systems (Terumo CVS) has received reports of no flow through the over-pressure safety (OPS) Valve.

Jun 27, 2017Z-0257-2018—terminated

Terumo Cardiovascular Systems (Terumo CVS) has received reports of no flow through the over-pressure safety (OPS) Valve.

May 12, 2017Z-2550-2018—terminated

Kit was labeled with the incorrect Expiration Date of April 30, 2019.

Feb 20, 2017Z-1458-2017—terminated

Terumo CVS initiated a voluntary recall for the Level Sensor II Pads and Level Sensor Gel Pads due to non-compliant labeling because the product expiration date is displayed in a format that may not be recognizable to all users.

Feb 20, 2017Z-1459-2017—terminated

Terumo CVS initiated a voluntary recall for the Level Sensor II Pads and Level Sensor Gel Pads due to non-compliant labeling because the product expiration date is displayed in a format that may not be recognizable to all users.

Jul 25, 2016Z-2706-2016—terminated

Terumo Custom Cardiovascular Procedure Kits are labeled as Xcoating surface coated tubing, but tubing is non-coated

Jun 8, 2016Z-2210-2016—terminated

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Inter chlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Jun 8, 2016Z-2207-2016—terminated

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Jun 8, 2016Z-2209-2016—terminated

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Jun 8, 2016Z-2208-2016—terminated

Terumo Cardiovascular Systems (Terumo CVS) is conducting a voluntary recall of its heater-cooler devices due to the distributor no longer selling the Interchlor sanitizer (Product Code 205) referenced in its heater-cooler cleaning instructions.

Mar 14, 2016Z-1250-2016—terminated

Specific CDI Blood Parameter Monitoring System 500 devices are being voluntarily recalled because the BPM Sensor Head Assemblys Thermistor, which provides the blood temperature value that results in accurate display values on the monitor, does not meet specification. This may cause inaccurate temperature measurement and inaccurate analyte display values on the CDI System 500 monitor. Terumo Cardiovascular Systems (Terumo CVS) received complaints of inaccurate temperature measurements for specific devices distributed since November 5, 2015. These include both new production devices and recently serviced devices. POTENTIAL HAZARD There have been no reported illnesses or injuries as a result of this issue. A user who is not aware that the CDI System 500 is displaying inaccurate temperature values may not manage patient temperature appropriately. Using inaccurate information to manage warming and cooling strategies for a procedure could result in prolonged time on bypass. It could also lead to unnecessarily aggressive temperature management, resulting in excessive hyper or hypothermia, with potential neurologic and organ dysfunction, or increased blood component damage. Inaccurate temperature measurement could also cause inaccurate measurements of other BPM values including potassium (K+), pO2, pCO2, and pH. The greater the temperature inaccuracy, the greater the degree of inaccuracy of these other BPM values due to the dependence of their algorithms on temperature for the calculations. Inaccurate measurement of these values could result in inappropriate patient management strategies being employed to address them with the potential to result in moderate patient injury.

Dec 11, 2015Z-0001-2017—terminated

Terumo Cardiovascular Systems (Terumo CVS) has received complaints of the CDI System 500 monitor displaying the H/S DISCONNECT AT CUVETTE error message when the CDI H/S Cuvette does not make a proper connection to the CDI H/S Probe. In this situation, blood parameter values for HCT, Hgb, and SO2 do not display.

Aug 24, 2015Z-2788-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 24, 2015Z-2786-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 24, 2015Z-2789-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 24, 2015Z-2787-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 24, 2015Z-2790-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 24, 2015Z-2791-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 24, 2015Z-2792-2015—terminated

Custom Cardiovascular Procedure Kits contain recalled component Medivators HPH700 High Performance Hemoconcentrators

Aug 7, 2015Z-2742-2015—terminated

Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69.

May 27, 2015Z-1235-2017—terminated

Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures. In each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks. Terumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.

May 6, 2015Z-1828-2015—terminated

The Overpressure Safety Valve is not cleared with an indication for use on the arterial side of the extracoporeal bypass circuit. It is intended for use only on the venous side. There is a remote possibility of minimal blood loss, fluid weeping, or dripping on the user.

May 1, 2015Z-1688-2015—terminated

Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to preven

May 1, 2015Z-1687-2015—terminated

Terumo CVS is implementing field correction activities to address the identified causes of the Terumo System 1 experiencing loss of system power. These activities will be implemented in a phased approach beginning late 2015: - Implementing design changes and upgrades to improve the reliability of power switching and battery backup. - Adding a switch protector to the On/Off power switch to preven

Feb 6, 2015Z-1116-2015—terminated

Malfunctioning mixing valve and loss of cooling and/or heating capability. System goes to standby mode and shows error messages EOd (invalid mix valve measured range of motion) or EOE (valve unable to move or complete a motion). Mixing valves need to be replaced at regular intervals.

Jan 5, 2015Z-1232-2017—terminated

Terumo Cardiovascular Systems (Terumo CVS) is voluntarily recalling certain lots of Titan" Stabilizers and Titan" 360 Stabilizers because the Suction Tubing Clip may break when it is being attached to a Terumo CVS Hercules Stabilizing Arm. Although the clip is for convenience purposes and is not required for device operation, there is a potential for broken Suction Tubing Clip fragments to fall into the surgical site.

Aug 20, 2014Z-1201-2017—terminated

Terumo Cardiovascular Systems (Terumo CVS) is recalling certain lots of Hercules TM 360 Universal Stabilizing Arms because the two pins that hold the sleeve attached to the distal end of the device may come loose, causing the pins and sleeve to separate from the device. While the arm can still function without the pins and sleeve in place, there is potential for the loose components to fall into the surgical site.

Jun 17, 2014Z-2019-2014—terminated

An evaluation of the APS1 Operator Manual found that existing instructions lack clarity and accuracy related to specific items.

Jun 17, 2014Z-2018-2014—terminated

An evaluation of the APS1 Operator Manual found that existing instructions lack clarity and accuracy related to specific items.

May 23, 2014Z-2097-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2104-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2095-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2100-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2098-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2099-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2096-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2103-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2106-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2105-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2094-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2101-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 23, 2014Z-2102-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications

May 9, 2014Z-1634-2014—terminated

There is a potential for inaccuracies in blood gas readings/measurements. There is a potential risk of respiratory alkalosis or cerebral vasoconstriction with low paCO2. If the Hemoglobin and Hematocrit measurements are not accurate, there is a potential risk of unnecesary blood transfusion with resulting risk of infection or inflammatory reaction. BPM shipped without meeting manufacturing accep

Apr 14, 2014Z-2028-2014—terminated

During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.

Apr 14, 2014Z-2030-2014—terminated

During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.

Apr 14, 2014Z-2029-2014—terminated

During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.

Mar 28, 2014Z-1465-2014—terminated

Incorrect customer notification label applied to the Tyvek lid of this specific lot of Terumo Cardiovascular Procedure Kits

Mar 3, 2014Z-1224-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Mar 3, 2014Z-1221-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Mar 3, 2014Z-1218-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Mar 3, 2014Z-1223-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Mar 3, 2014Z-1219-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Mar 3, 2014Z-1222-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Mar 3, 2014Z-1220-2014—terminated

During in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified exposed basket wire on certain product lots of Sarns Malleable Dual-stage Venous Return Cannulae and Sarns Dual-stage Venous Return Cannulae.

Dec 12, 2013Z-1069-2014—terminated

Sterility of medical devices intended for use in surgical procedures may be compromised.

Oct 23, 2013Z-0193-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow Aortic Cannulae and Sarns" Venous Return Cannulae.

Oct 23, 2013Z-0336-2014—terminated

Terumo Cardiovascular Systems (Terumo CVS) became aware of a customer complaint that a 15mm x 8mm Gelsoft graft was received which the customer believed was a 15mm x 6mm Gelsoft graft unit. The manufacturer confirmed a batch of product code 631508 was incorrectly labeled.

Oct 23, 2013Z-0196-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow Aortic Cannulae and Sarns" Venous Return Cannulae.

Oct 23, 2013Z-0194-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow Aortic Cannulae and Sarns" Venous Return Cannulae.

Oct 23, 2013Z-0195-2014—terminated

During an in-process inspection, Terumo Cardiovascular Systems (Terumo CVS) identified the presence of loose fiber particulate that exceeded finished product specifications on certain product lots of Sarns" Soft-Flow Aortic Cannulae and Sarns" Venous Return Cannulae.

Jul 26, 2013Z-1969-2013—terminated

The operators manual does not fully describe the sequence of expected events that will occur when a specific operating condition exists. Specifically, if the user is adjusting the System 1 centrifugal pump speed via the Central Control Monitor at the same time an alarm condition occurs with configuration for a pump coast response, the labeling in the operators manual does not fully describe how the pump will react.

Jul 26, 2013Z-1974-2013—terminated

Terumo Cardiovascular Systems (TCVS) has become aware that an Advanced Perfusion System 1 (APS1) Centrifugal Pump 'Service Pump' message may occur due to user induced motor movement. In this instance, the APS1 will identify the movement (as it is designed) and display a Service Pump message error. The user might interpret this as a pump which requires service rather than a message displayed due to pump movement while not activated.